| Literature DB >> 32752014 |
Bin Liu1,2, Chengwei Kang2, Fengzhou Fang2,3.
Abstract
Biometric measurement of the anterior segment is of great importance for the ophthalmology,Entities:
Keywords: anterior segment; corneal topography; geometric measurement; tomography
Mesh:
Year: 2020 PMID: 32752014 PMCID: PMC7435894 DOI: 10.3390/s20154285
Source DB: PubMed Journal: Sensors (Basel) ISSN: 1424-8220 Impact factor: 3.576
Figure 1Human eye anatomy [3].
Figure 2Development of the anterior segment measurement technologies.
Figure 3Placido disc and representative patterns of corneal shapes [36].
Figure 4Triangulation principle for corneal surface measurement [48].
Figure 5Two-wavelength holographic interferometer for contour evaluation of human corneas [53].
Figure 6Schematic diagram of the projection Moiré profilometry [61].
Figure 7Schematic diagram of Kawara’s design [63].
Figure 8Optical schematic of the Maastricht Topographer [65].
Figure 9Setup for measuring the corneal surface based on the Moiré method [67].
Figure 10Twyman-Green interferometer for testing the corneal surface [43].
Figure 11Eye surface profiler using two symmetrical projectors [80].
Figure 12Overlapping slits to map the cornea using the Scanning-slit technology [36].
Figure 13Principle of Scheimpflug imaging [90].
Figure 14The basic principle of ultrasound biomicroscopy (UBM). The transducer generates an ultrasound pulse. The pulse encounters the anterior segment tissues. The backscattered echoes would be detected by the same transducer, resulting in the A-scan signal to indicate the distances between the tissues. The B-scan image of the anterior segment could be obtained by scanning.
Figure 15Schematic diagram of a time domain-optical coherence tomography (TD-OCT) system [85]. PD: Photodetector; BS: Beam splitter.
Figure 16Schematic diagram of a spectral domain OCT (SD-OCT) system [85]. BS: Beam splitter.
Figure 17Schematic diagram of a SS-OCT system [85]. BS: Beam splitter; BPD: Balanced photodetector.
Figure 18Schematic diagram of a full field OCT (FF-OCT) system [172].
The pros and cons of the different categories of OCT.
| OCT Categories | Pros. | Cons. |
|---|---|---|
| TD-OCT | simple construction and signal processing method | low scanning speed and low detection sensitivity |
| SD-OCT | significant detection efficiency | strong SNR roll-off in depth and low detection resolution of the spectral distribution |
| SS-OCT | reduced fringe wash-out effects, lower sensitivity attenuation, higher detection efficiency, improved imaging speed and higher sensitivity | the intrinsic instability of the light sources and still adopt the point scanning scheme |
| FF-OCT | en-face (transverse) tomographic images directly without point-by-point raster scanning, simplicity and a larger field of view | difficult to align the optical pathway, inherent sensitivity to motion, the crucial degradation of the contrast and image resolution with the imaging depth |
Figure 19The trend of the development of OCT systems [147].
The comparison of the mentioned measuring approaches.
| Measuring Strategy | Technologies | Speed | Resolution | Covering Area | Penetration Depth | Major Limitations |
|---|---|---|---|---|---|---|
| Pattern projection | Placido disk | tens of ms/image | 10 μm | 14 mm in diameter | only the anterior corneal surface | Surface reconstruction of the irregular cornea fails as a result of that the rings or raster patterns could merge or cross in this case. |
| Rasterstereography [ | tens of ms/image | 4 μm | over 12 mm in diameter | |||
| Interferometry | Holographic technique [ | 1 ms | 0.1 μm | entire corneal surface | The hologram is particularly easy to be affected by the vibrations and air turbulence. | |
| Moiré technique [ | 1 s | 2.6 μm | 7 mm in square | The fine structures or those with a larger gradient on the cornea would be erased due to the inevitable application of the low-pass filter in the Moiré technique, TGI, and FTP. | ||
| TGI [ | tens of ms/image | 6 μm | 6 mm in square | |||
| FTP [ | <1 s | <10 μm | 20 mm in diameter | |||
| Parallel line scanning | Scanning-slit [ | 1.5 s | >10 μm | 11 mm in diameter | anterior segment | Utilizing parallel line scanning causes difficulty in the image registration due to the lack of shared points during scanning. The depth of focus is so limited that the imaging quality of the lens is poor. |
| Rotational line scanning | Scheimpflug imaging [ | 1~2 s | <10 μm | 14~16 mm in diameter | anterior segment | Visualization of the entire lens and anterior chamber is inaccessible. |
| Point scanning | UBM [ | 50~100 ms/B-Scan | 20~50 μm, | over 20 mm in diameter | from the anterior segment to the retina | As an immersion technique, the contact and time-consuming nature limit its application. |
| TD-OCT [ | 2000 A-Scans/s | 18 μm | 16 mm in width | Quantitative measurement can only be retrieved accurately upon the correction of the fan and optical distortions. OCT is unable to see through the opaque tissues. | ||
| SD-OCT [ | >300,000 A-Scans/s | 5 μm | 13 mm in width | |||
| SS-OCT [ | >2,000,000 A-Scans / s | 8 μm | 12 mm in width | |||
| En-face | FF-OCT [ | tens of ms/image | 1 μm | 1~2 cm2 |
The specifications of ATLAS 9000 system (Carl Zeiss Meditec) using the Placido disk are quoted; The imaging process is a single shot by the CCD or CMOS cameras. The speed is dependent on the frame rate of the camera; The resolution using 10 MHz ultrasound is 150 µm, while the one using 50~100 MHz ultrasound is 20~50 μm; UBM is capable of imaging the retina only when the 10 MHz ultrasound is applied. The 50~100 MHz ultrasound is suitable for imaging the anterior segment. Analogous to UBM, OCT uses 850 nm light to investigate the retina and 1310 nm light to depict the anterior segment typically.
Figure 20Single and multiple back scattered light [119].