Comparison and repeatability of keratometric and corneal power measurements obtained by Orbscan II, Pentacam, and Galilei corneal tomography systems.

Alexandra Z Crawford1, Dipika V Patel, Charles N J McGhee.   

Abstract

PURPOSE: To assess the repeatability and comparability of corneal power and central corneal thickness (CCT) measurements obtained using Orbscan II (Bausch & Lomb), Pentacam (Oculus), and Galilei (Ziemer) tomographers.
DESIGN: Prospective, comparative study.
METHODS: setting: Departments of Ophthalmology, University of Auckland and Auckland District Health Board, Auckland, New Zealand. study population: Thirty eyes of 30 healthy participants. observations. CCT and corneal power measured using Orbscan II, Pentacam, and Galilei tomography. main outcome measures: Degree of agreement in and repeatability of CCT and corneal power measures.
RESULTS: Orbscan II measured significantly lower CCT compared with Pentacam (20 μm; P < .0005) and Galilei (18 μm; P < .0005). The Orbscan II had wide limits of agreement when compared with both the Galilei (-11 to 47 μm) and Pentacam (-88 to 47 μm). For each device, the intraclass correlation coefficient for CCT was higher than 0.9. The coefficient of variation ranged from 0.33% to 0.93%. There was no significant difference in mean steep keratometry or mean flat keratometry between instrument pairs. However, there was poor agreement in flat keratometry and steep keratometry obtained by Orbscan II compared with those obtained by the Galilei and Pentacam.
CONCLUSIONS: The keratometry and pachymetry measurements obtained by Orbscan II, Pentcam, and Galilei tomographers were sufficiently disparate that the 3 devices could not be considered equivalent. All 3 devices demonstrated a high level of repeatability, although the Galilei exhibited the best repeatability.
Copyright © 2013 Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2013        PMID: 23540708     DOI: 10.1016/j.ajo.2013.01.029

Source DB:  PubMed          Journal:  Am J Ophthalmol        ISSN: 0002-9394            Impact factor:   5.258


  38 in total

1.  Difference in normal corneal thickness and curvature between Mongolian and Han nationalities.

Authors:  Hai-Xia Zhao; Li Zhang; Wen-Ying Guan
Journal:  Int J Ophthalmol       Date:  2015-04-18       Impact factor: 1.779

2.  [Difference and distance between the central and thinnest points of the cornea: impact of refractive state, age and ocular side].

Authors:  J Steinberg; C Kohl; T Katz; G Richard; S J Linke
Journal:  Ophthalmologe       Date:  2014-04       Impact factor: 1.059

3.  Assessment of total corneal power after myopic corneal refractive surgery in Chinese eyes.

Authors:  Yanjun Hua; Chao Pan; Qinmei Wang
Journal:  Int Ophthalmol       Date:  2019-03-01       Impact factor: 2.031

4.  Clinical Evaluation and Validation of the Dutch Crosslinking for Keratoconus Score.

Authors:  Robert P L Wisse; Rob W P Simons; Martijn J B van der Vossen; Marc B Muijzer; Nienke Soeters; Rudy M M A Nuijts; Daniel A Godefrooij
Journal:  JAMA Ophthalmol       Date:  2019-06-01       Impact factor: 7.389

5.  A novel Hartman Shack-based topography system: repeatability and agreement for corneal power with Scheimpflug+Placido topographer and rotating prism auto-keratorefractor.

Authors:  Gaurav Prakash; Dhruv Srivastava; Sounak Choudhuri
Journal:  Int Ophthalmol       Date:  2015-03-28       Impact factor: 2.031

6.  Scanning-slit topography in patients with keratoconus.

Authors:  László Módis; Gábor Németh; Eszter Szalai; Zsuzsa Flaskó; Berthold Seitz
Journal:  Int J Ophthalmol       Date:  2017-11-18       Impact factor: 1.779

7.  Accommodation response and spherical aberration during orthokeratology.

Authors:  L Batres; S Peruzzo; M Serramito; G Carracedo
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2019-11-12       Impact factor: 3.117

8.  Comparison of objective and subjective refractive surgery screening parameters between regular and high-resolution Scheimpflug imaging devices.

Authors:  J Bradley Randleman; Jihan Akhtar; Michael J Lynn; Renato Ambrósio; William J Dupps; Ronald R Krueger; Stephen D Klyce
Journal:  J Cataract Refract Surg       Date:  2014-12-20       Impact factor: 3.351

9.  Agreement between clinical history method, Orbscan IIz, and Pentacam in estimating corneal power after myopic excimer laser surgery.

Authors:  Kaevalin Lekhanont; Manachai Nonpassopon; Khemruetai Wannarosapark; Varintorn Chuckpaiwong
Journal:  PLoS One       Date:  2015-04-08       Impact factor: 3.240

10.  The effect of corneal irregularity on astigmatism measurement by automated versus ray tracing keratometry.

Authors:  Hyun Cheol Roh; Roy S Chuck; Jimmy K Lee; Choul Yong Park
Journal:  Medicine (Baltimore)       Date:  2015-04       Impact factor: 1.889

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.