Repeatability of automatic measurements performed by a dual Scheimpflug analyzer in unoperated and post-refractive surgery eyes.

Giacomo Savini1, Michele Carbonelli, Piero Barboni, Kenneth J Hoffer.   

Abstract

PURPOSE: To assess the repeatability of the anterior segment measurements performed with a dual Scheimpflug analyzer (Galilei) in unoperated and post-refractive surgery eyes.
SETTING: Private practice.
DESIGN: Evaluation of diagnostic test.
METHODS: Three consecutive scans were acquired in unoperated eyes and in eyes that had excimer laser surgery for myopia. Unoperated eyes were enrolled from 3 subgroups: younger than 50 years, aged between 50 and 70 years, and older than 70 years. The following parameters were evaluated: simulated keratometry, posterior corneal power, total corneal power, anterior and posterior best-fit sphere radius, mean and thinnest central corneal thickness, anterior chamber depth and volume, horizontal and vertical corneal diameter, iridocorneal angle in the 4 quadrants, and corneal spherical aberration. Repeatability was assessed using analysis of variance (ANOVA), the coefficient of variation (COV), intraclass correlation (ICC), and test-retest variability.
RESULTS: The study evaluated 45 unoperated eyes (n = 45) and 15 post-refractive surgery eyes (n = 15). Each age subgroup in the unoperated group comprised 15 eyes. The ANOVA did not detect significant differences between the 3 measurements for any parameter. The COV was less than 0.5% for corneal power measurements and less than 3.5% for all remaining parameters except spherical aberration (16.68%). The ICC was more than 0.99 for corneal power measurements and more than 0.94 for all remaining parameters.
CONCLUSIONS: The anterior segment measurements provided by the dual Scheimpflug analyzer were highly repeatable. Repeatability did not change with age or after myopic photorefractive keratectomy or laser in situ keratomileusis.
Copyright © 2011 ASCRS and ESCRS. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2011        PMID: 21241913     DOI: 10.1016/j.jcrs.2010.07.039

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  26 in total

1.  Placido disk-based topography versus high-resolution rotating Scheimpflug camera for corneal power measurements in keratoconic and post-LASIK eyes: reliability and agreement.

Authors:  Rachele R Penna; Ugo de Sanctis; Martina Catalano; Luca Brusasco; Federico M Grignolo
Journal:  Int J Ophthalmol       Date:  2017-03-18       Impact factor: 1.779

2.  Comparative analysis of the relationship between anterior and posterior corneal shape analyzed by Scheimpflug photography in normal and keratoconus eyes.

Authors:  Raúl Montalbán; Jorge L Alio; Jaime Javaloy; David P Piñero
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2013-01-20       Impact factor: 3.117

3.  Repeatability and comparability of anterior segment biometry obtained by the Sirius and the Pentacam analyzers.

Authors:  Paola De la Parra-Colín; Manuel Garza-León; Tonatiuh Barrientos-Gutierrez
Journal:  Int Ophthalmol       Date:  2013-04-25       Impact factor: 2.031

4.  Human ocular biometry.

Authors:  Robert C Augusteyn; Derek Nankivil; Ashik Mohamed; Bianca Maceo; Faradia Pierre; Jean-Marie Parel
Journal:  Exp Eye Res       Date:  2012-07-20       Impact factor: 3.467

5.  Comparison of Anterior Chamber Depth Measurements from the Galilei Dual Scheimpflug Analyzer with IOLMaster.

Authors:  Roma P Patel; Rahul T Pandit
Journal:  J Ophthalmol       Date:  2012-01-19       Impact factor: 1.909

6.  Precision of corneal thickness measurements obtained using the scheimpflug-placido imaging and agreement with ultrasound pachymetry.

Authors:  Jinhai Huang; Giacomo Savini; Chengfang Wang; Weicong Lu; Rongrong Gao; Yuanguang Li; Qinmei Wang; Yune Zhao
Journal:  J Ophthalmol       Date:  2015-02-24       Impact factor: 1.909

7.  Agreement between clinical history method, Orbscan IIz, and Pentacam in estimating corneal power after myopic excimer laser surgery.

Authors:  Kaevalin Lekhanont; Manachai Nonpassopon; Khemruetai Wannarosapark; Varintorn Chuckpaiwong
Journal:  PLoS One       Date:  2015-04-08       Impact factor: 3.240

8.  No change in anterior chamber dimensions after femtosecond LASIK for hyperopia.

Authors:  Xiaodong Zhou; Tao Li; Zhi Chen; Lingling Niu; Xingtao Zhou; Zimei Zhou
Journal:  Eye Contact Lens       Date:  2015-05       Impact factor: 2.018

9.  Correlation between Corneal Topographic Indices and Higher-Order Aberrations in Keratoconus.

Authors:  Sepehr Feizi; Bahram Einollahi; Alireza Raminkhoo; Shahram Salehirad
Journal:  J Ophthalmic Vis Res       Date:  2013-04

10.  Comparison of high-resolution Scheimpflug and high-frequency ultrasound biomicroscopy to anterior-segment OCT corneal thickness measurements.

Authors:  Anastasios John Kanellopoulos; George Asimellis
Journal:  Clin Ophthalmol       Date:  2013-11-20
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.