Literature DB >> 16163498

Evaluation of biometric methods for measuring the anterior chamber depth in the non-contact mode.

B Meinhardt1, O Stachs, J Stave, R Beck, R Guthoff.   

Abstract

PURPOSE: High-resolution biometry of the anterior ocular segment is now becoming more and more important against a background of refractive surgery and the evaluation of potentially accommodative lens replacement materials. The aim of this study was a systematic investigation of the currently available non-contact methods for measuring the anterior chamber depth (ACD).
METHODS: The ACDs of 50 phakic eyes of 27 patients aged between 19 and 59 years were measured with the IOL-Master (Zeiss), the AC-Master (Zeiss), the Pentacam (Oculus) and slit-lamp pachymetry by Jaeger (Haag-Streit).
RESULTS: The median anterior chamber depth in the investigated eyes was 3.63 mm for the IOL-Master (minimum 2.88 mm, maximum 4.22 mm), 3.802 mm for the AC-Master (2.816 mm-4.373 mm), 3.915 mm for the Pentacam (minimum 2.994 mm, maximum 4.614 mm) and 3.75 mm for Jaeger (2.887 mm-4.29 mm). With a probability of error of alpha=0.05 there were no significant differences concerning the ACD between the methods of Jaeger and AC-Master, Jaeger and IOL-Master, or Pentacam and AC-Master (Wilcoxon and Wilcox). The intra-individual variability was +/-5.4 microm for AC-Master, +/-12.7 microm for Pentacam, +/-24.5 microm for IOL-Master and +/-41.2 microm for Jaeger. The maximum method-dependent difference in ACD determination was 285 microm.
CONCLUSIONS: All the methods allow non-contact biometry, but the results might differ due to measuring principles inherent to the system, experience of the examiner and compliance of the patient. Partial coherence interferometry with the AC-Master offers the advantage of measurement exactly along the optical axis with the highest reproducibility and patient compliance.

Entities:  

Mesh:

Year:  2005        PMID: 16163498     DOI: 10.1007/s00417-005-0103-7

Source DB:  PubMed          Journal:  Graefes Arch Clin Exp Ophthalmol        ISSN: 0721-832X            Impact factor:   3.117


  17 in total

Review 1.  Measuring agreement in method comparison studies.

Authors:  J M Bland; D G Altman
Journal:  Stat Methods Med Res       Date:  1999-06       Impact factor: 3.021

2.  The repeatability and accuracy of axial length and anterior chamber depth measurements from the IOLMaster .

Authors:  A K Lam; R Chan; P C Pang
Journal:  Ophthalmic Physiol Opt       Date:  2001-11       Impact factor: 3.117

3.  Measurement of accommodation after implantation of an accommodating posterior chamber intraocular lens.

Authors:  Achim Langenbucher; Stefan Huber; Nhung X Nguyen; Berthold Seitz; Gabriele C Gusek-Schneider; Michael Küchle
Journal:  J Cataract Refract Surg       Date:  2003-04       Impact factor: 3.351

4.  Eye-length measurement by interferometry with partially coherent light.

Authors:  A F Fercher; K Mengedoht; W Werner
Journal:  Opt Lett       Date:  1988-03-01       Impact factor: 3.776

5.  Biometric investigation of changes in the anterior eye segment during accommodation.

Authors:  W Drexler; A Baumgartner; O Findl; C K Hitzenberger; A F Fercher
Journal:  Vision Res       Date:  1997-10       Impact factor: 1.886

6.  Practice styles and preferences of ASCRS members--2000 survey. American Society of Cataract and Refractive Surgery.

Authors:  D V Leaming
Journal:  J Cataract Refract Surg       Date:  2001-06       Impact factor: 3.351

7.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

8.  Reproducibility of optical biometry using partial coherence interferometry : intraobserver and interobserver reliability.

Authors:  A Vogel; H B Dick; F Krummenauer
Journal:  J Cataract Refract Surg       Date:  2001-12       Impact factor: 3.351

9.  [Validity of anterior chamber depth measurements for the evaluation of accommodation after implantation of an accommodative Humanoptics 1CU intraocular lens].

Authors:  G U Auffarth; M Martin; H A Fuchs; T M Rabsilber; K A Becker; I Schmack
Journal:  Ophthalmologe       Date:  2002-11       Impact factor: 1.059

10.  Anterior chamber depth measured by two methods in myopic and hyperopic phakic IOL implant.

Authors:  M Vetrugno; N Cardascia; L Cardia
Journal:  Br J Ophthalmol       Date:  2000-10       Impact factor: 4.638

View more
  10 in total

Review 1.  Recent advances in ophthalmic anterior segment imaging: a new era for ophthalmic diagnosis?

Authors:  Aristides Konstantopoulos; Parwez Hossain; David F Anderson
Journal:  Br J Ophthalmol       Date:  2007-04       Impact factor: 4.638

2.  Comparison of anterior segment measurements using rotating Scheimpflug imaging and partial coherence interferometry.

Authors:  Mohammad Soleimani; Hassan Hashemi; Shiva Mehravaran; Mehdi Khabazkhoob; Mohammad Hassan Emamian; Mohammad Shariati; Akbar Fotouhi
Journal:  Int J Ophthalmol       Date:  2013-08-18       Impact factor: 1.779

3.  [Biometry of the anterior eye segment for implantation of phakic anterior chamber lenses. A comparison of current measurement devices].

Authors:  L Kiraly; G Duncker
Journal:  Ophthalmologe       Date:  2012-03       Impact factor: 1.059

4.  Usefulness of anterior chamber depth measurement for efficacy assessment of steroid pulse therapy in patients with Vogt-Koyanagi-Harada disease.

Authors:  Tomohiro Otsuki; Kimiya Shimizu; Akihito Igarashi; Kazutaka Kamiya
Journal:  Jpn J Ophthalmol       Date:  2010-11-05       Impact factor: 2.447

5.  [Precision of a new device for biometric measurements in pseudophakic eyes].

Authors:  A L Hildebrandt; G U Auffarth; M P Holzer
Journal:  Ophthalmologe       Date:  2011-08       Impact factor: 1.059

6.  Comparison of Ocular Morphological Parameters Related to Lens Position by Anterior Segment Optical Coherence Tomography and Ultrasound Biomicroscopy.

Authors:  Zhiying Yu; Fenglei Wang; Fang Dong; Na Li; Dabo Wang; Ling Wang
Journal:  Int J Clin Pract       Date:  2022-03-20       Impact factor: 3.149

7.  Effect of anterior chamber depth on the choice of intraocular lens calculation formula in patients with normal axial length.

Authors:  Mohammad Miraftab; Hassan Hashemi; Akbar Fotouhi; Mehdi Khabazkhoob; Farhad Rezvan; Soheila Asgari
Journal:  Middle East Afr J Ophthalmol       Date:  2014 Oct-Dec

8.  Comparison of Anterior Segment Measurements Obtained by Aladdin Optical Biometer and Sirius Corneal Topography.

Authors:  Onur Polat; Zeki Baysal; Serkan Özcan; Sibel İnan; Ümit Übeyt İnan
Journal:  Turk J Ophthalmol       Date:  2016-12-01

Review 9.  Biometric Measurement of Anterior Segment: A Review.

Authors:  Bin Liu; Chengwei Kang; Fengzhou Fang
Journal:  Sensors (Basel)       Date:  2020-07-31       Impact factor: 3.576

10.  Development and Verification of an Adjustment Factor for Determining the Axial Length Using Optical Biometry in Silicone Oil-Filled Eyes.

Authors:  Gerd U Auffarth; Tadas Naujokaitis; Louise Blöck; Anna Daghbashyan; Jan Meis; Victor A Augustin; Ramin Khoramnia; Timur M Yildirim
Journal:  Diagnostics (Basel)       Date:  2022-01-11
  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.