Pachymetric measurements with a new Scheimpflug photography-based system: intraobserver repeatability and agreement with optical coherence tomography pachymetry.

Myriam Milla1, David P Piñero, Francisco Amparo, Jorge L Alió.   

Abstract

PURPOSE: To evaluate the intraobserver repeatability of pachymetric measurements obtained with a new Scheimpflug photography-based system and the agreement of the measurements with those obtained by optical coherence tomography (OCT).
SETTING: Vissum Corporation, Alicante, Spain.
DESIGN: Evaluation of diagnostic technology.
METHODS: Corneal thickness was measured at different locations (center; 2.5 mm and 4.0 mm nasal, temporal, superior, and inferior) in 18 eyes of 18 patients with no ocular pathology or previous surgery with a Scheimpflug photography-based system (Sirius). Three consecutive measurements at each location were obtained with the system to evaluate intraobserver repeatability. Afterward, pachymetric measurements were obtained at the same locations with an OCT system (Visante) to evaluate the agreement between techniques using the Bland-Altman method.
RESULTS: The intraclass correlation coefficient of repeated measures ranged from 0.990 to 0.997. The coefficient of variation was lower than 1% and the standard deviations of the repeated measurements (S(w)) were below 6 μm at all corneal locations. Statistically significant differences were found between the coefficient of variation and S(w) values of the repeated measurements corresponding to the 2.5 mm and 4.0 mm locations (P ≤ .037). The agreement with OCT pachymetric measurements was poor, with ranges of agreement larger than 20 μm at all corneal locations. Statistically significant differences were found between standard deviations of the differences between the 2 devices at the 2.5 mm and 4.0 mm locations (P ≤ .017).
CONCLUSION: Scheimpflug photography-based technology provided repeatable pachymetric measurements that were not equivalent to those obtained with OCT.
Copyright © 2011 ASCRS and ESCRS. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2011        PMID: 21241914     DOI: 10.1016/j.jcrs.2010.08.038

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  24 in total

1.  Intra-session repeatability of iridocorneal angle measurements provided by a Scheimpflug photography-based system in healthy eyes.

Authors:  Clara Ruiz-Belda; David P Piñero; Pedro Ruiz-Fortes; Roberto Soto-Negro; Myriam Moya; Rafael J Pérez-Cambrodí; Alberto Artola
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2015-07-15       Impact factor: 3.117

2.  Repeatability and agreement of central corneal thickness measurement with non-contact methods: a comparative study.

Authors:  Tarannum Mansoori; Nagalla Balakrishna
Journal:  Int Ophthalmol       Date:  2017-04-22       Impact factor: 2.031

3.  Comparison of central corneal thickness measurements with three new optical devices and a standard ultrasonic pachymeter.

Authors:  Hasan Ali Bayhan; Seray Aslan Bayhan; Izzet Can
Journal:  Int J Ophthalmol       Date:  2014-04-18       Impact factor: 1.779

4.  Comparison between corneal elevation maps using different reference surfaces with Scheimpflug-Placido topographer.

Authors:  Engy Mohamed Mostafa
Journal:  Int Ophthalmol       Date:  2016-07-28       Impact factor: 2.031

5.  Could the percent tissue altered (PTA) index be considered as a unique factor in ectasia risk assessment?

Authors:  Carlos Rocha-de-Lossada; José-María Sánchez-González; Rahul Rachwani-Anil; Juan-Luis García-Madrona; Federico Alonso-Aliste; Sandra Figueroa-Ardila; Elvira Colmenero-Reina
Journal:  Int Ophthalmol       Date:  2020-07-27       Impact factor: 2.031

6.  OPD-Scan III: a repeatability and inter-device agreement study of a multifunctional device in emmetropia, ametropia, and keratoconus.

Authors:  Soheila Asgari; Hassan Hashemi; Ebrahim Jafarzadehpur; Alireza Mohamadi; Farhad Rezvan; Akbar Fotouhi
Journal:  Int Ophthalmol       Date:  2016-02-08       Impact factor: 2.031

7.  Comparison of artemis 2 ultrasound and Visante optical coherence tomography corneal thickness profiles.

Authors:  Roxana Ursea; Matthew Feng; Raksha Urs; Arindam RoyChoudhury; Ronald H Silverman
Journal:  J Refract Surg       Date:  2012-11-30       Impact factor: 3.573

8.  Corneal Morphologic Characteristics in Patients With Down Syndrome.

Authors:  Jorge L Alio; Alfredo Vega-Estrada; Pablo Sanz; Amr A Osman; Ahmed M Kamal; Amr Mamoon; Hany Soliman
Journal:  JAMA Ophthalmol       Date:  2018-09-01       Impact factor: 7.389

9.  Central corneal thickness and anterior chamber depth measurement by Sirius(®) Scheimpflug tomography and ultrasound.

Authors:  J Jorge; Jl Rosado; Ja Díaz-Rey; Jm González-Méijome
Journal:  Clin Ophthalmol       Date:  2013-02-26

10.  Outcomes of photorefractive keratectomy in patients with posterior corneal steepening.

Authors:  Yumna Busool Abu-Eta; Achia Nemet; Igor Vainer; Michael Mimouni; Tzahi Sela; Gur Munzer; Igor Kaiserman
Journal:  Eye (Lond)       Date:  2020-10-06       Impact factor: 4.456

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.