Comparison of anterior and posterior elevation map measurements between 2 Scheimpflug imaging systems.

Ramin Salouti1, Mohammad H Nowroozzadeh, Mohammed Zamani, Alireza Hafrkonandeh Fard, Sasan Niknam.   

Abstract

PURPOSE: To compare the anterior and posterior elevation measurements above the best-fit sphere (BFS) between the Galilei (dual-Scheimpflug) and Pentacam (single-Scheimpflug) systems in candidates for refractive surgery.
SETTING: Poostchi Ophthalmology Research Center, Shiraz, Iran.
METHODS: In this prospective study, both eyes of consecutive patients were examined with the 2 imaging systems. The anterior and posterior BFS and the central and maximum elevation measurements in each of 4 quadrants were compared between the dual-Scheimpflug system and the single-Scheimpflug system. Bland-Altman plots were used to evaluate the agreement in the measurements between the imaging systems.
RESULTS: The study comprised 31 patients (62 eyes). The mean radius of the anterior BFS measurement was 7.72 mm +/- 0.21 (SD) with the dual-Scheimpflug system and 7.83 +/- 0.21 mm with the single-Scheimpflug system. The mean radius of the posterior BFS was 6.37 +/- 0.18 mm and 6.47 +/- 0.19 mm, respectively. On the anterior and posterior elevation maps, the single-Scheimpflug system led to statistically significantly higher measurements than the dual-Scheimpflug system in the central, inferior, and temporal elevations (P<.001). Bland-Altman plots showed more agreement between the systems in anterior elevation measurements than in posterior elevation measurements.
CONCLUSION: In refractive surgery candidates with healthy eyes, the elevation map measurements with the single-Scheimpflug system were significantly different from those with the dual-Scheimpflug system in most aspects and therefore may lead to misinterpretation of clinical data that could alter clinical decisions if the 2 imaging systems are considered interchangeable.

Entities:  

Mesh:

Year:  2009        PMID: 19393884     DOI: 10.1016/j.jcrs.2009.01.008

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  11 in total

1.  Comparison between corneal elevation maps using different reference surfaces with Scheimpflug-Placido topographer.

Authors:  Engy Mohamed Mostafa
Journal:  Int Ophthalmol       Date:  2016-07-28       Impact factor: 2.031

Review 2.  Corneal topography in keratoconus: state of the art.

Authors:  F Cavas-Martínez; E De la Cruz Sánchez; J Nieto Martínez; F J Fernández Cañavate; D G Fernández-Pacheco
Journal:  Eye Vis (Lond)       Date:  2016-02-22

3.  Reproducibility and repeatability of central corneal thickness measurement in healthy eyes using four different optical devices.

Authors:  Remzi Karadag; Murat Unluzeybek; Ozgur Cakici; Ayse Yagmur Kanra; Huseyin Bayramlar
Journal:  Int Ophthalmol       Date:  2016-10-08       Impact factor: 2.031

4.  Comparison between Two Scheimpflug Anterior Segment Analyzers.

Authors:  Alireza Baradaran-Rafii; Tahmineh Motevasseli; Forouzan Yazdizadeh; Farid Karimian; Sahba Fekri; Amir Baradaran-Rafii
Journal:  J Ophthalmic Vis Res       Date:  2017 Jan-Mar

5.  Repeatability and Reproducibility of Anterior Segment Measurements in Normal Eyes Using Dual Scheimpflug Analyzer.

Authors:  Zeynep Altıparmak; Ramazan Yağcı; Emre Güler; Zeynel Arslanyılmaz; Metin Canbal; İbrahim F Hepşen
Journal:  Turk J Ophthalmol       Date:  2015-12-05

Review 6.  Galilei Corneal Tomography for Screening of Refractive Surgery Candidates: A Review of the Literature, Part II.

Authors:  Majid Moshirfar; Mahsaw N Motlagh; Michael S Murri; Hamed Momeni-Moghaddam; Yasmyne C Ronquillo; Phillip C Hoopes
Journal:  Med Hypothesis Discov Innov Ophthalmol       Date:  2019

7.  Repeatability and agreement of five imaging systems for measuring anterior segment parameters in healthy eyes.

Authors:  Mukesh Kumar; Rohit Shetty; Chaitra Jayadev; Harsha L Rao; Debarun Dutta
Journal:  Indian J Ophthalmol       Date:  2017-04       Impact factor: 1.848

8.  Effect of Cycloplegia on Corneal Biometrics and Refractive State.

Authors:  Abbas Bagheri; Mohadeseh Feizi; Aliakbar Shafii; Amir Faramarzi; Mehdi Tavakoli; Shahin Yazdani
Journal:  J Ophthalmic Vis Res       Date:  2018 Apr-Jun

9.  Comparison of anterior segment changes after femtosecond laser LASIK and SMILE using a dual rotating Scheimpflug analyzer.

Authors:  Bu Ki Kim; Su Joung Mun; Young Hoon Yang; Ji Sun Kim; Jun Hyung Moon; Young Taek Chung
Journal:  BMC Ophthalmol       Date:  2019-12-11       Impact factor: 2.209

Review 10.  Biometric Measurement of Anterior Segment: A Review.

Authors:  Bin Liu; Chengwei Kang; Fengzhou Fang
Journal:  Sensors (Basel)       Date:  2020-07-31       Impact factor: 3.576

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.