Anterior segment biometry with the Pentacam: comprehensive assessment of repeatability of automated measurements.

Hema Shankar1, Deepa Taranath, Chandramalar T Santhirathelagan, Konrad Pesudovs.   

Abstract

PURPOSE: To comprehensively assess the reliability of automated Pentacam (Oculus, Inc.) measurements.
SETTING: Flinders Eye Centre, Flinders Medical Centre, Bedford Park, South Australia, Australia.
METHODS: Both eyes of 35 normal volunteers were tested twice on the same day by 2 different observers. All automated values were recorded, and manual analysis of topographic maps was performed only to overrule variance in corneal thickness due to pupil decentration altering the central reference point. Repeatability was determined with Bland-Altman limits of agreement and reported as the coefficient of repeatability (COR = +/-1.96 standard deviation of differences). Relative repeatability (RR) was calculated as a percentage of the ratio of COR to the mean.
RESULTS: Overall, repeatability was good. Corneal curvature, reported in diopters, showed good repeatability anteriorly (simulated keratometry mean COR+/-0.28D; RR=0.64%) and posteriorly (COR+/-011D; RR=1.85%). Peripheral corneal curvature was more reliable when calculated by the sagittal (axial) method (RR=1.57%) than by the tangential (meridional) method (RR=2.38%). Keratometric power deviation was less reliable (RR=16.39%). Anterior chamber measurements showed good reliability (RR=3.07%-5.68%) except for anterior chamber angle (RR=14.41%). Pupil diameter showed poor reliability (RR=25.77%). Central corneal thickness was comparable at pupil center and corneal vertex, but peripheral repeatability was much better when centered on the corneal vertex (COR+/-16.00microm; RR=2.56%) than at pupil center (COR+/-26.28microm; RR=4.23%).
CONCLUSIONS: Pentacam corneal curvature and anterior chamber parameters were highly repeatable, but pupil measurements had poor repeatability. Peripheral pachymetry readings were affected by pupil decentration and required manual analysis using the corneal vertex as the point of reference to achieve good repeatability.

Entities:  

Mesh:

Year:  2008        PMID: 18165089     DOI: 10.1016/j.jcrs.2007.09.013

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  53 in total

1.  Repeatability and agreement of CCT measurement in myopia using entacam and ultrasound pachymetry.

Authors:  Yong-Heng Luo; Qian Zhong; Ping-Bo Ouyang; Xiao-Jian Guo; Xuan-Chu Duan
Journal:  Int J Ophthalmol       Date:  2012-06-18       Impact factor: 1.779

2.  Comparison of anterior segment measurements using rotating Scheimpflug imaging and partial coherence interferometry.

Authors:  Mohammad Soleimani; Hassan Hashemi; Shiva Mehravaran; Mehdi Khabazkhoob; Mohammad Hassan Emamian; Mohammad Shariati; Akbar Fotouhi
Journal:  Int J Ophthalmol       Date:  2013-08-18       Impact factor: 1.779

3.  Assessment of the anterior chamber parameters after laser iridotomy in primary angle close suspect using Pentacam and gonioscopy.

Authors:  Alireza Esmaeili; Behzad Barazandeh; Sina Ahmadi; Alireza Haghi; Seyed Mahdi Ahmadi Hosseini; Fereshteh Abolbashari
Journal:  Int J Ophthalmol       Date:  2013-10-18       Impact factor: 1.779

4.  Validation of Fourier analysis of videokeratographic data.

Authors:  Haris Sideroudi; Georgios Labiris; Fienke Ditzel; Efi Tsaragli; Kimonas Georgatzoglou; Haralampos Siganos; Vassilios Kozobolis
Journal:  Int Ophthalmol       Date:  2017-06-15       Impact factor: 2.031

5.  The effects of tolterodine on anterior segment and choroidal thickness in patients with overactive bladder syndrome.

Authors:  Hande Hüsniye Telek; Omer Gokhan Doluoglu; Ayse Burcu; Gulizar Demirok; Firdevs Ornek; Ali Ayyildiz
Journal:  Ther Adv Urol       Date:  2017-04-11

6.  Corneal thickness and volume in subclinical and clinical keratoconus.

Authors:  Seyed Mahdi Ahmadi Hosseini; Norhani Mohidin; Fereshteh Abolbashari; Bariah Mohd-Ali; Chandramalar T Santhirathelagan
Journal:  Int Ophthalmol       Date:  2012-11-09       Impact factor: 2.031

7.  [Scheimpflug photography of acute angle closure].

Authors:  J D Unterlauft; Y Yafai; P Wiedemann
Journal:  Ophthalmologe       Date:  2013-09       Impact factor: 1.059

8.  Repeatability and comparability of anterior segment biometry obtained by the Sirius and the Pentacam analyzers.

Authors:  Paola De la Parra-Colín; Manuel Garza-León; Tonatiuh Barrientos-Gutierrez
Journal:  Int Ophthalmol       Date:  2013-04-25       Impact factor: 2.031

9.  Higher order aberrations of the corneal surface after laser subepithelial keratomileusis.

Authors:  Hyun Ho Jung; Yong Sok Ji; Han Jin Oh; Kyung Chul Yoon
Journal:  Korean J Ophthalmol       Date:  2014-07-22

10.  Optical coherence tomography accurately measures corneal power change from laser refractive surgery.

Authors:  Ryan P McNabb; Sina Farsiu; Sandra S Stinnett; Joseph A Izatt; Anthony N Kuo
Journal:  Ophthalmology       Date:  2014-12-06       Impact factor: 12.079

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.