We demonstrate swept source OCT utilizing vertical-cavity surface emitting laser (VCSEL) technology for in vivo high speed retinal, anterior segment and full eye imaging. The MEMS tunable VCSEL enables long coherence length, adjustable spectral sweep range and adjustable high sweeping rate (50-580 kHz axial scan rate). These features enable integration of multiple ophthalmic applications into one instrument. The operating modes of the device include: ultrahigh speed, high resolution retinal imaging (up to 580 kHz); high speed, long depth range anterior segment imaging (100 kHz) and ultralong range full eye imaging (50 kHz). High speed imaging enables wide-field retinal scanning, while increased light penetration at 1060 nm enables visualization of choroidal vasculature. Comprehensive volumetric data sets of the anterior segment from the cornea to posterior crystalline lens surface are also shown. The adjustable VCSEL sweep range and rate make it possible to achieve an extremely long imaging depth range of ~50 mm, and to demonstrate the first in vivo 3D OCT imaging spanning the entire eye for non-contact measurement of intraocular distances including axial eye length. Swept source OCT with VCSEL technology may be attractive for next generation integrated ophthalmic OCT instruments.
We demonstrate swept source OCT utilizing vertical-cavity surface emitting laser (VCSEL) technology for in vivo high speed retinal, anterior segment and full eye imaging. The MEMS tunable VCSEL enables long coherence length, adjustable spectral sweep range and adjustable high sweeping rate (50-580 kHz axial scan rate). These features enable integration of multiple ophthalmic applications into one instrument. The operating modes of the device include: ultrahigh speed, high resolution retinal imaging (up to 580 kHz); high speed, long depth range anterior segment imaging (100 kHz) and ultralong range full eye imaging (50 kHz). High speed imaging enables wide-field retinal scanning, while increased light penetration at 1060 nm enables visualization of choroidal vasculature. Comprehensive volumetric data sets of the anterior segment from the cornea to posterior crystalline lens surface are also shown. The adjustable VCSEL sweep range and rate make it possible to achieve an extremely long imaging depth range of ~50 mm, and to demonstrate the first in vivo 3D OCT imaging spanning the entire eye for non-contact measurement of intraocular distances including axial eye length. Swept source OCT with VCSEL technology may be attractive for next generation integrated ophthalmic OCT instruments.
Optical coherence tomography (OCT) is a non-invasive and non-contact imaging modality that
enables two-dimensional cross-sectional and three-dimensional volumetric imaging of tissue
architecture [1]. OCT is analogous to B-mode ultrasound,
measuring the echo time delay and intensity of reflected or backscattered light from internal tissue
structures. Coherence gating enables micrometer axial resolution without the need for confocal
detection. OCT is well suited to image semi-transparent objects and was first applied in
ophthalmology, where it has become a clinical standard for diagnosing disease and monitoring
treatment.The largest ophthalmic application of OCT is retinal imaging and therefore clinical procedures
for retinal diagnosis are well established [2]. Commercial
retinal OCT instruments operate near 840 nm wavelength and at axial scan rates up to 52 kHz. OCT
enables imaging morphology of the retina including the fovea and optic disk for diagnosis and
monitoring of therapeutic response in major diseases such as age related macular degeneration,
glaucoma and diabetic retinopathy. Moreover, functional imaging using motion (blood flow) and
polarization properties are possible and may enhance diagnostic applications [3,4].OCT imaging of the anterior segment of the eye has received considerable attention since its
first demonstration in 1994 [5]. Commercial OCT devices
dedicated to anterior segment imaging achieve axial scan rates of 26-30 kHz and use light sources in
the 840 nm or 1310 nm wavelength range [6]. OCT of the
anterior segment is valuable for diagnosis of corneal disorders such as keratoconus, and for pre-
and post-operative assessment during surgical procedures such as keratomileusis (LASIK),
phototherapeutic keratectomy (PTK), astigmatic keratotomy and lamellar keratoplasty [7]. Contact lens fitting and intraocular lens (IOL) power
calculation can be performed using volumetric OCT data [8,9]. In addition, OCT is used in anterior chamber
angle evaluation for glaucoma diagnosis and management [10].Low-coherence interferometry is another technique which is used clinically for ocular biometry,
the measurement of intraocular distances [11-13]. This technique is non-contact and allows measurement of the
central depth profile of the eye, offering higher resolution than traditional ultrasound.
Quantitative assessment of axial eye length and anterior chamber depth is crucial for proper
intraocular lens power calculation.The future of OCT is strongly influenced by technological advances. Innovations in core
technologies are essential for advances and include: (1) development of new broadband low-coherence
and tunable light sources, (2) high speed detection and data acquisition systems and (3) methods for
processing and visualization of large volumetric data sets.OCT can be performed using different methods which detect the magnitude and time delay of light.
Early OCT systems used time domain detection with an interferometer and broadband light source. In
time domain detection, interferometric fringes are recorded in time, while the interferometer
reference arm mechanically scans the optical path delay. Limitations on reference arm scanning
speeds and detection sensitivity limit the speed of time-domain OCT (TD-OCT) [1]. A dramatic increase in speed and detection sensitivity can be achieved by
utilizing Fourier domain detection and two general implementations of Fourier domain detection are
possible: spectral/Fourier domain and swept source/Fourier domain detection.In spectral/Fourier domain OCT (SD-OCT), a broadband light source is used and light is detected
with a spectrometer and line scan camera that record the interferometric signal as a function of
wavelength or frequency in the spectral domain. The interferometric signal is then Fourier
transformed to generate an axial scan. Since the entire signal is measured simultaneously, a
significant detection efficiency advantage can be achieved over time domain detection [14-16]. In addition,
advances in high speed CCD and CMOS technology allowed speed increases by up to two orders of
magnitude faster than TD-OCT [17,18]. SD-OCT quickly became a standard technology for clinical ophthalmic OCT
instruments.In contrast to SD-OCT, swept source/Fourier domain OCT (SS-OCT) uses a frequency swept light
source and a single or dual balanced detector with a high speed A/D converter [19]. SS-OCT detects the interference signal as a function of time as the light
source is swept in frequency and achieves similar sensitivity advantages to SD-OCT [15,20]. SS-OCT avoids the
need for bulky spectrometers and line scan cameras, but requires a high speed, narrow line with
swept light source. In SD-OCT, spectrometers have limited spectral resolution from grating resolving
power, beam spot size and finite pixel dimensions of the line-scan camera [21]. This limited resolving power limits the imaging depth range, producing a
sensitivity roll-off versus depth. In contrast, in SS-OCT the spectral resolution is determined by
the instantaneous linewidth, or coherence length, of the frequency swept light source, combined with
the A/D acquisition rate. The spectral resolution in SS-OCT can be much higher than in SD-OCT,
enabling extended depth range imaging with significantly reduced sensitivity roll-off. SS-OCT also
has many other advantages including: reduced fringe wash-out effects from sample motion or rapid
transverse scanning, better light detection efficiency since there are no diffraction grating losses
and photodetectors have better quantum efficiency than cameras, ease of implementing dual balanced
detection to cancel excess light source noise and ease of implementing multichannel detection
methods used in polarization sensitive detection. Finally, SS-OCT has the advantage that the light
source frequency sweep range and repetition rate can be adjusted to tailor the resolution, imaging
range and axial scan repetition rate for the specific imaging application.The important light source parameters for SS-OCT include: rapid sweep repetition rates over a
wide frequency/wavelength range, single longitudinal mode operation (narrow instantaneous linewidth)
for long coherence length, low excess noise and adjustable laser operation parameters. Table 1
shows a summary of swept source laser technology development in OCT. SS-OCT was
demonstrated as early as 1997 using a semiconductor laser with a galvanometer tuned grating external
cavity at 10 Hz rate and 840 nm wavelength [22]. Dramatic
increases in speed were achieved using external cavity tunable lasers employing resonant scanning
mirrors, diffraction gratings, dispersion prisms, rotating polygons, and scanning filters [23-27]. Although
initial sweep rates were slow, current designs achieve up to a few hundreds of kHz [28]. However, conventional external cavity tunable lasers use bulk
optics or fiber components, which makes resonators relatively long. Hence, it is difficult to
achieve single longitudinal mode operation and the coherence length is limited. Moreover, the sweep
rate is limited because the long cavity requires time for amplified spontaneous emission to reach
gain saturation as the laser frequency is swept. This limitation was overcome using Fourier-domain
mode locking (FDML) [29]. FDML lasers have a gain medium, a
long optical fiber delay and a tunable fiber Fabry-Perot filter, such that the frequency sweep
propagates in the optical fiber delay and returns to the filter as it is tuned synchronously. FDML
lasers can achieve ultrahigh sweep rates of up to 5.2 MHz by buffering or multiplexing the sweeps
[30]. FDML works optimally at 1.3 μm and 1.5 μm
wavelengths where optical fiber dispersion and loss are negligible. However, dispersion can be
compensated using fiber Bragg gratings to improve performance at 1 μm and 1.3 μm
wavelengths [31,32].
Recently, external cavity tunable lasers have been miniaturized using microelectromechanical
systems (MEMS) technology [33]. This led to an increase in
sweep rates enabling OCT imaging up to 150 kHz axial scan rates. Commercial devices are available at
wavelengths around 840 nm, 1060 nm, 1310 nm and 1550 nm and an overview of specifications is
presented in Table 1. However, most technologies require
that the MEMS filter bandwidth be broad enough to tune multiple longitudinal modes in order to
reduce excess noise associated with mode competition. Consequently, the coherence length of
MEMS-tunable short cavity lasers can be limited. The reduction of laser cavity length to achieve
single longitudinal mode operation significantly improves SS-OCT performance. This can be achieved
using vertical-cavity surface emitting laser (VCSEL) technology. Although VCSELs were developed in
late 1970s, applications were limited to photonics [34-36]. Recently, OCT imaging using
MEMS-tunable VCSELs at 1300 nm was reported [37,38].In this paper, we demonstrate VCSEL light source technology at 1060 nm wavelengths for high speed
ophthalmic OCT imaging. We examine VCSEL specifications which make the light source suitable for OCT
imaging and enable integration of multiple ophthalmic imaging modes into a single instrument. The
operating modes of the device include: ultrahigh speed, high resolution retinal and choroidal
imaging; high speed, high resolution anterior segment imaging and long depth range full eye imaging.
The extremely long imaging depth range enables the first in vivo 3D OCT imaging
spanning the entire eye from the cornea to the retina. This full eye imaging enables measurement of
intraocular distances including anterior chamber depth and total axial eye length.
2. Methods
2.1. VCSEL and its properties
The VCSEL (Praevium Research Inc./Thorlabs Inc.) consists of an optically pumped InGaAs
multi-quantum well active layer between two highly reflective mirrors. The bottom mirror is an
oxidized GaAs/AlAs distributed Bragg reflector. Together with a dielectric top mirror suspended on
an electrostatic MEMS actuator, this forms a microcavity. Tuning is performed by electrostatic
deflection of the MEMS-tunable mirror driven by a programmable waveform generator (AFG3102,
Tektronix Inc.) and an amplifier (Fig. 1(a)
). The VCSEL is optically pumped using an 830 nm laser diode and generates light centered
around 1060 nm wavelength, which is directed to semiconductor booster amplifiers (SOA). A second
waveform generator channel is used to modulate the first SOA current (FM2.5, Messtec Power Converter
GmbH). The output of the second SOA is greater than 20 mW, which allows use of a high splitting
ratio 80/20 coupler in the OCT interferometer for increased system sensitivity.
Fig. 1
MEMS-tunable VCSEL light source and its performance. (a) Schematic of VCSEL module: WDM,
wavelength-division multiplexer; ISO, isolator; SOA, booster optical amplifiers. Electrical signals
indicated by dashed lines. (b) Signal vs. depth of 1060 nm swept light sources: VCSEL and short
external-cavity laser.
MEMS-tunable VCSEL light source and its performance. (a) Schematic of VCSEL module: WDM,
wavelength-division multiplexer; ISO, isolator; SOA, booster optical amplifiers. Electrical signals
indicated by dashed lines. (b) Signal vs. depth of 1060 nm swept light sources: VCSEL and short
external-cavity laser.The VCSEL has several features which are well suited for OCT applications. First, the short
cavity length means that only one longitudinal mode overlaps the laser gain bandwidth. Single mode
operation and lack of mode hopping makes the laser coherence length extremely long, and
significantly reduces sensitivity roll-off with imaging range. The coherence lengths of the VCSEL
operating at 50 kHz sweep rate with 45 nm tuning range versus a MEMS Fabry-Perot tunable short
cavity laser (Axsun Technologies, Inc.) were measured by acquiring interference traces from a
Mach-Zehnder interferometer using two high-speed 1.2 GHz InGaAs photodetectors and a 1 GHz bandwidth
oscilloscope (DPO7104, Tektronix Inc.). Figure 1(b) plots
signal versus depth, where total path delay is two times the depth. The coherence length is usually
defined as the full width at −3 dB amplitude (or −6 dB intensity). The coherence
length of VCSEL exceeds 200 mm (measurement limited by the oscilloscope bandwidth), whereas a MEMS
Fabry-Perot tunable short cavity laser has ~24 mm coherence length. In addition, MEMS technology
enables adjustable sweep rate. Along with adjustable and wide frequency sweep range, this enables
adjustment of imaging speed, axial resolution and depth range. Finally, the VCSEL can be driven by
custom waveforms which linearize the sweep to optimize imaging range, light exposure and A/D
bandwidth utilization.
2.2. Prototype OCT instrument
A schematic diagram of the OCT imaging system is shown in Fig.
2
. The average output power was ~800 μW directly from the VCSEL and greater than 20 mW
after the two-stage amplification for all imaging configurations with the SOA currents adjusted as
shown in Fig. 3
. Before entering the system, the light from VCSEL module was attenuated by a fiber coupler
(not shown in Fig. 2). The output was divided between the OCT
interferometer and a sweep calibration Mach-Zehnder interferometer (first 80:20 fiber coupler). The
light entering the OCT interferometer was split into sample and reference arm by another 80:20
coupler. The sample arm of the OCT interferometer was attached to a slit-lamp patient interface for
scanning the eye. The light incident on the eye was 1.9 mW, consistent with American National
Standard Institute (ANSI) standards (ANSI Z136.1-2007) [49].
The system had a telecentric anterior eye imaging configuration with an f = 150 mm, 50 mm diameter
lens. Retinal imaging required adding an adapter lens to collimate the incident beam on the eye and
relay the beam scanning pivot point to the pupil. A dichroic mirror coupled a fixation target view
into the OCT beam scanning path. Light from a single pass reference arm and sample arm was
interfered with a fiber coupler and the signal detected by a prototype high speed, dual-balanced
InGaAs photodetector receiver PDB1 (custom prototype; Thorlabs Inc.). Signal from the dual-balanced
photodetector was digitized by a high speed A/D converter. An additional trigger output from the
programmable waveform generator served as a start signal for sweep acquisition (Fig. 1(a)). Another prototype dual-balance photodiode PDB2 (custom prototype;
Thorlabs Inc.) was used to detect interference fringes from the calibration MZI. A 12-bit 500 MSPS
A/D converter (ATS9350; Alazar Technologies Inc.) was used for the retinal imaging mode. Both
anterior segment and full eye imaging required a faster sampling rate to support sufficient imaging
depth range. Accordingly, an 8-bit card 1 GSPS A/D converter (ATS9870; Alazar Technologies Inc.) was
used for anterior segment and full eye imaging.
Fig. 2
Experimental setup. Retinal imaging was performed by adding ocular lens to anterior segment
configuration and adjusting the fixation target path. SC – galvanometric scanners, FT
– fixation target, DM – dichroic mirror, DC – dispersion compensation glass, RR
– retroreflector, PDB1/PDB2 – balanced photodetectors, MZI – Mach-Zehnder
interferometer.
Fig. 3
Signals in VCSEL module (module driving signal, SOA current waveforms, sweep trigger), MZI fringe
and integrated spectrum for selected tuning frequencies: 290 kHz, 100 kHz and 50 kHz.
Experimental setup. Retinal imaging was performed by adding ocular lens to anterior segment
configuration and adjusting the fixation target path. SC – galvanometric scanners, FT
– fixation target, DM – dichroic mirror, DC – dispersion compensation glass, RR
– retroreflector, PDB1/PDB2 – balanced photodetectors, MZI – Mach-Zehnder
interferometer.Signals in VCSEL module (module driving signal, SOA current waveforms, sweep trigger), MZI fringe
and integrated spectrum for selected tuning frequencies: 290 kHz, 100 kHz and 50 kHz.
2.3. Sweep optimization
The VCSEL was driven by a periodic signal, producing forward and backward sweeps during each
drive period. Figure 3 shows various signals, including the
VCSEL MEMS drive signal, booster current modulation, sweep trigger and MZI calibration fringes. The
integrated spectra of VCSEL output were measured by an optical spectrum analyzer and had larger
amplitude at the edges of the sweeps where the sweep speed was slower. The three operating modes
were achieved using MEMS sinusoidal drive signals at 50 kHz, 100 kHz and 290 kHz. Other signals at
200 kHz are not shown in Fig. 3. The MEMS-tunable VCSEL
itself can sweep ~100 nm [38]. However, the SOA technology
used for this study did not support the full bandwidth, which limited the sweep range to about 85 nm
and compromises axial resolution. Broader bandwidth SOAs are currently being developed and it is
expected that a 100 nm amplified output can ultimately be achieved.Data was only acquired during using the forward sweep for the 50 kHz, 100 kHz and 200 kHz
operating modes. The backward sweeps were removed by modulating the first SOA stage current. This
optimized ocular exposure and sensitivity, since the power during the forward sweep could be
increased for a given average power. The sweep duty cycle was 51% at 50 kHz and increased to 60% at
100 kHz (Fig. 3, second row). The change in duty cycle
resulted from an asymmetry of the forward and backward sweeps due to nonlinear MEMS behavior when
the sweep rate was not at the MEMC resonance frequency. When the MEMS-VCSEL was driven at its
resonance frequency of 290 kHz, it had a sinusoidal sweep and nearly 100% duty cycle, so the
effective imaging speed was doubled. We have previously shown that a MEMS VCSEL can be driven with
customized waveforms that compensate sweep nonlinearity and linearize the frequency sweep and
increase high duty cycle for more effectively use of A/D bandwidth [38]. Although sweep linearization is possible, small deviations from a perfectly linear
sweep exist and it is still necessary to re-scale the sweep so that it is linear in frequency or
wavenumber before Fourier transforming.Sweep-to-sweep variations in the VCSEL frequency/wavelength scanning are noticeable and a single
calibration MZI trace could not be used, except when the VCSEL was operated at resonance. Our
previous work with VCSELs for OCT used optical clocking methods to clock the A/D converter at a
varying frequency clock signal derived from an interferometer, in order to sample the signal so that
it is linear in frequency or wavenumber [38]. Optical
clocking compensated sweep-to-sweep variation and worked up to 400 MSPS in these previous studies.
However, there is currently no A/D card that can reliably clock up to 1 GSPS. In order to perform
the extended imaging range experiments in these studies, dual channel acquisition with an MZI trace
obtained on the second channel was used to perform individual sweep-by-sweep calibration.
Limitations with the A/D card bus speeds when acquiring large data sets with dual channels prevented
acquiring both the forwards and backwards sweeps for all operating modes, except for when the VCSEL
was operated at resonance. On resonance, where sweep-to-sweep repeatability is high and a single
calibration trace acquired before imaging could be applied to all of the sweeps so that only one
channel of A/D was required.
2.4. System performance and in vivo OCT imaging
The OCT system integrated multiple imaging modes in a single instrument. Operating modes included
retinal, anterior segment and full eye imaging. Table 2
shows details of the configuration and system performance for each operating mode. The
effective axial scan rate was two times higher than MEMS VCSEL drive frequency if both forward and
backward sweeps were acquired (100% duty cycle). The VCSEL design enabled adjustment of
frequency/wavelength sweep range to modify imaging depth range and/or axial resolution.
Table 2
Configuration and performance of integrated SS-OCT system
Parameter
Retinal imaging
Anterior segment imaging
Full eye imaging/ ocular biometry
A/D acquisition card
ATS9350 (AlazarTech)
ATS9870 (AlazarTech)
ATS9870 (AlazarTech)
Detector bandwidth
330 MHz
1.5 GHz
1.5 GHz
Acquisition scheme
Forward sweep/Dual channel (OCT signal + MZI)
Dual sweep/Single channel
Forward sweep/Dual channel (OCT signal + MZI)
Forward sweep/Dual channel (OCT signal + MZI)
Clocking scheme
Internal clock
Internal clock
Internal clock
Clocking rate
500 MS/s
1 GS/s
1 GS/s
VCSEL tuning frequency
100 kHz
200 kHz
290 kHz
100 kHz
50 kHz
Effective A-scan rate
100 kHz
200 kHz
580 kHz
100 kHz
50 kHz
Acquired samples/sweep/channel
2048
896
1280
4096
7424
Wavelength tuning range
83 nm
83 nm
83 nm
85 nm
45 nm
Axial resolution*
9.0 μm
8.8 μm
8.9 μm
9.0 μm
12.4 μm
Transverse resolution
20 μm§
20 μm§
20 μm§
73 μm†
73 μm†
Imaging depth range*
6.8 mm
2.6 mm
1.9 mm
13.6 mm
37.7 mm
Sensitivity
100.5 dB
97.4 dB
96.3 dB
98.8 dB
100.0 dB
6 dB roll-off depth‡
6.8 mm
2.9 mm
2.2 mm
11.0 mm
39.0 mm
*In tissue (refractive index n = 1.35).
§Estimated based on aberration-free model eye.
†Measured using a beam profiler.
‡In air.
*In tissue (refractive index n = 1.35).§Estimated based on aberration-free model eye.†Measured using a beam profiler.‡In air.Retinal imaging was performed using internally clocked A/D acquisition at 500 MSPS. Retinal
imaging at sweep rates ranging from 100 kHz to 580 kHz were tested. Since the clock rate was fixed,
increasing the axial scan rates resulted in a reduction of the imaging range (from 6.8 mm to 1.9 mm
in tissue). In addition, faster imaging speeds resulted in lower sensitivity (decreasing from 100.5
dB at axial scan rates of 100 kHz to 96.3 dB at 580 kHz). The sensitivity was measured using 1.9 mW
of optical power incident on a test mirror with a calibrated attenuator. Figure 4
shows sensitivity roll-off for the different operating modes. Similar to imaging depth range,
the sensitivity 6 dB drop depth becomes shallower as the sweep rate increases. The axial image
resolution was ~9 μm.
Fig. 4
Sensitivity roll-offs of VCSEL-OCT system: (a) point spread functions at different depths in air
for the full eye imaging mode at 50 kHz, (b) signal roll-offs for different imaging modes. 6 dB
signal drop is indicated by the dashed line.
Sensitivity roll-offs of VCSEL-OCT system: (a) point spread functions at different depths in air
for the full eye imaging mode at 50 kHz, (b) signal roll-offs for different imaging modes. 6 dB
signal drop is indicated by the dashed line.Anterior segment imaging required modification of the patient interface to telecentrically scan
and focus the OCT beam onto the anterior chamber (Fig. 2).
The estimated transverse resolution of this configuration was ~75 μm. Since the anterior
chamber and crystalline lens span a longer axial range than the retina, the imaging range must be
extended by trading off axial resolution and/or reducing imaging speed. The anterior segment imaging
mode operated at axial scan rates of 100 kHz using an A/D data acquisition card at a fixed 1 GSPS
sampling rate with dual channel acquisition. Simultaneous acquisition of OCT signal and MZI fringes
was required to calibrate of each laser sweep and account for sweep-to-sweep variation. The anterior
segment imaging mode had a 13.6 mm imaging range in tissue and enabled visualizing ocular structures
spanning the cornea to posterior surface of the crystalline lens. The axial resolution was 9
μm and the measured sensitivity was 98.8 dB, dropping by 6 dB at a depth of 11 mm in air
(Fig. 4). Corneo-scleral imaging was performed using
parameters analogous to those used for retinal imaging at 100 kHz axial scan rates.Finally, full eye imaging was performed at axial scan rates of 50 kHz. The data acquisition
scheme was similar to that used for anterior segment imaging and the sweep range was reduced to 45
nm in order to extend the imaging range to ~50 mm in air (~38 mm in tissue). This long imaging range
enabled OCT imaging of the anterior segment and retina in a single data set. The very long VCSEL
coherence length and broad detection bandwidth supported a 6 dB signal roll-off at a 40 mm range in
air (Fig. 4). The OCT system sensitivity remained the same as
in the 100 kHz retinal imaging mode. However, the reduced sweep range caused a decrease in axial
resolution to 12 μm in tissue.Full eye imaging is an extremely challenging operating mode because it requires long imaging
range to visualize both anterior and retinal structures (minimum of 40 mm in air) as well as an
optical focusing and scanning design to collect light from both the anterior segment and retina.
Refractive errors in the eye can make it difficult to simultaneously maximize collection efficiency
from both anterior segment and retina. Imaging was performed using the anterior segment scanning
configuration with a long focal length (f = 150 mm) ocular lens and a 73 μm incident spot
size on the anterior segment of the eye. This provided sufficient depth of focus to image the
anterior segment while normal ocular refraction produced focusing onto the retina. The beam spot
size at the retina depended on the refraction status of the eye.The procedures involving human subjects were approved by the Committee on Use of Humans as
Experimental Subjects (COUHES) at MIT. Written informed consent was obtained prior to the study.
3. Results
3.1. Retinal and choroidal imaging
Volumetric OCT data sets were acquired by raster scanning at sweep rates of 100 kHz, 200 kHz and
580 kHz. Fundus views and corresponding cross-sectional images are shown in Fig. 5
. The images shown are from a 32 year old normal subject with myopia. The transverse scan
sampling density was kept constant at 12 μm per axial scan (~2x less than estimated spot size
at the retina) by scaling the transverse scan area by the number of axial scans.
Fig. 5
Retinal OCT imaging using VCSEL-tunable light source. Fundus images and selected cross-sections
from volumetric data sets acquired at 100 kHz (a), 200 kHz (b) and 580 kHz (c). Red-free fundus
photograph indicating scanned areas at different speeds (d). Transverse sampling density and
acquisition time are kept constant. Aspect ratio of all presented cross-sections is the same. High
speeds enable wide-field imaging.
Retinal OCT imaging using VCSEL-tunable light source. Fundus images and selected cross-sections
from volumetric data sets acquired at 100 kHz (a), 200 kHz (b) and 580 kHz (c). Red-free fundus
photograph indicating scanned areas at different speeds (d). Transverse sampling density and
acquisition time are kept constant. Aspect ratio of all presented cross-sections is the same. High
speeds enable wide-field imaging.A total measurement time of ~2 seconds was used for each volume, consistent with a typical
clinically acceptable acquisition time. The data sets comprised 400 × 400, 600 × 600
and 1000 × 1000 axial scans and covered a scanned areas of 5 × 5 mm2, 7
× 7 mm2 and 12 × 12 mm2 for sweep rates of 100 kHz, 200 kHz and
580 kHz, respectively. The scanned area at 100 kHz required separate acquisitions for the central
macular region versus the optical nerve head (ONH), whereas imaging at 580 kHz enabled an almost
six-fold increase in scanned field covering both macular and optical nerve head in a single data
set. All cross-sectional images in Fig. 5 were cropped to
show a depth of 1.5 mm in tissue and the aspect ratio was kept constant. The 9 μm axial
resolution enables visualization of the retinal layers. Enhanced light penetration at 1060 nm
allowed imaging of the choroid.Ultrahigh imaging speeds enabled acquisition of dense volumetric data rapidly enough to reduce
motion artifacts in vivo. In order to see examples of the effects of eye motion,
the optic disk region was imaged at different axial scan rates. In all cases, the scanning protocol
consisted of 500 × 500 axial scans and imaging was performed over a 6 × 6
mm2 area. Figure 6
shows cross-sectional images extracted from the volumes along the slow scanning axis. An OCT
fundus image was generated for each volume to confirm the B-scans were taken from the same location.
The data demonstrates that motion artifacts in the slow scan direction are reduced as the
acquisition time is reduced from 2.6 s to 0.5 s.
Fig. 6
Imaging of the optic nerve head region at different speeds: fundus image (a) and extracted
central fast and slow cross-sections showing reduced motion artifacts with increased speed (b).
Imaging of the optic nerve head region at different speeds: fundus image (a) and extracted
central fast and slow cross-sections showing reduced motion artifacts with increased speed (b).A sweep rate of 580 kHz was used to image the retina and choroid over a large field of view. A
rendering of a wide-field volumetric data set consisting of 1000 × 1000 axial scans is shown
in Fig. 7(a)
. The acquisition time was ~1.8 s, so that there is relatively little motion artifact present.
In this case, Tthe volumetric data set was averaged using 5 × 5 pixel lateral kernel,
preserving axial resolution. The volumetric data set enables the generation of arbitrary
cross-sectional images. A virtual scan that crosses the fovea and optic disk was extracted (Fig. 7(b)), and demonstrates the ability to visualize deep
choroidal layers and the choroid-scleral interface. Additionally, the high sensitivity enables
imaging of scleral arteries.
Fig. 7
Wide-field choroidal OCT imaging using VCSEL tunable light source. (a) Rendering of volumetric
wide-field data set. (b) Virtual (arbitrary) cross-sectional image showing deep light penetration
and ability to visualize choroid and sclera. Arrow indicates scleral vessel. (c) Projection OCT
image of the entire choroid. Signal below RPE was integrated. Red line indicates direction of
section in (b). OCT projection images at a depth of (d) 30 μm, (e) 80 μm and (f) 200
μm below RPE showing choroidal layers and sclera. Signal was integrated from 40 μm
thick slices below RPE.
Wide-field choroidal OCT imaging using VCSEL tunable light source. (a) Rendering of volumetric
wide-field data set. (b) Virtual (arbitrary) cross-sectional image showing deep light penetration
and ability to visualize choroid and sclera. Arrow indicates scleral vessel. (c) Projection OCT
image of the entire choroid. Signal below RPE was integrated. Red line indicates direction of
section in (b). OCT projection images at a depth of (d) 30 μm, (e) 80 μm and (f) 200
μm below RPE showing choroidal layers and sclera. Signal was integrated from 40 μm
thick slices below RPE.Projection en-face OCT images can also be generated from the volumetric data. The highly
reflective retinal pigment epithelium (RPE) was identified by using an edge-detection algorithm. The
retinal data set was flattened with respect to the RPE boundary and projection OCT images of the
choroid at different depths below RPE were generated by axial summation of the OCT image intensity
from 40 μm thick slices. Anatomically distinguishable choroidal layers such as the
choriocapillaris, Sattler’s layer and Haller’s layer are visible and characterized by
blood vessels of different size, as shown in Figs. 7(d)-7(f).Wide-field OCT data can be also used to visualize vascular networks and generate images analogous
to angiography. Figure 8
shows an OCT fundus image and retinal and choroidal angiography like images extracted from
volumetric OCT data. Since retinal vessels generate shadows, it is possible to increase contrast in
a projection image by summing the intensity from depths 25 μm above to 25 μm below the
RPE layer and filtering out the background signal. The projection in inverse grey scale Fig. 8(b) differentiates retinal vessels indicated in orange.
Additionally, a fundus projection image of the dense choroidal blood vessels was generated by
integrating the OCT signal below RPE and inverting the scale (indicated in green in Fig. 8(c)). Due to shadowing effects, retinal vessels appear also
in the choroidal vasculature en-face image in Fig. 8(c). It
is possible to distinguish both vascular systems by comparing retinal and choroidal en-face
projections. A combined ‘angiographic’ OCT image can be generated (two-color image in
Fig. 8(d)) and agrees well with a late phase indocyanine
green (ICG) angiography image (Fig. 8(f)). OCT has the
advantage that it is non-invasive, while ICG requires intravenous dyes. The field of view in Fig. 8(a) was comparable to standard fundus photography (Fig. 8(e)). However, OCT is volumetric data which can be used to
display cross sectional and en-face projections of different retinal and choroidal layers.
Fig. 8
Wide-field OCT fundus angiography. OCT fundus image (a), segmented retinal (b) and choroidal
vasculature (c), combined OCT fundus image (d). Red-free fundus photography (e). Indocyanine green
(ICG) angiography (f). Comparison of details of retinal and choroidal vascular systems (g).
Wide-field OCT fundus angiography. OCT fundus image (a), segmented retinal (b) and choroidal
vasculature (c), combined OCT fundus image (d). Red-free fundus photography (e). Indocyanine green
(ICG) angiography (f). Comparison of details of retinal and choroidal vascular systems (g).
3.2. Anterior segment imaging
Anterior segment architecture can also be imaged using the longer imaging range, intermediate
axial scan rate OCT instrument operating mode. Figure 9
shows a rendering of a volumetric data set covering 16 × 16 mm2 of the
anterior eye. The volume consists of 500 × 500 axial scans acquired in 2.6 s. The
cross-sectional image shows the cornea, iris and entire crystalline lens and spans the entire
transverse anterior chamber width, from limbus to limbus. The cross-sectional image in Fig. 9(b) is generated by averaging 5 consecutive B-scans to
enhance signal to noise. Zoomed inserts show corneal sublayers such as the epithelium,
Bowman’s membrane, stroma and endothelium (Fig.
9(c)).
Fig. 9
Anterior segment imaging with VCSEL-OCT: (a) rendering of the volume, (b) central averaged
cross-sectional image, (c) zoomed fragments of the B-scan showing details of the corneal and the
crystalline lens.
Anterior segment imaging with VCSEL-OCT: (a) rendering of the volume, (b) central averaged
cross-sectional image, (c) zoomed fragments of the B-scan showing details of the corneal and the
crystalline lens.High-resolution imaging of the anterior segment structures enables visualization of limbus and
anterior chamber angle architecture. The system was operated in high-resolution mode with reduced
depth range, similar to that used for retinal imaging at 100 kHz. Figure 10
shows corneo-scleral imaging of the eye. The volumetric data set consisted of 500 ×
500 axial scans and covered a 7 × 7 mm2 area. This dense scan of the anterior
chamber angle enabled visualization of the limbal region along with landmarks such as the
corneo-scleral junction and rich scleral vasculature. Elements of the outflow system such as
Schlemm’s canal could be also identified (Figs.
10(a)-10(b)). The corneo-scleral junction could be
distinguished since scleral tissue is more scattering than corneal tissue. The enhanced penetration
of 1060 nm light enabled visualization of deeper structures such as the ciliary body. OCT volumetric
data were also used to visualize the scleral vascular system (Fig.
10(d)). The scleral interface was segmented and a projection image from a 1 mm deep slice is
shown.
Fig. 10
Corneo-scleral imaging: (a) cross-sectional OCT image presenting details of the anterior chamber
angle (S – sclera, CB – ciliary body, C – cornea, I – iris); (b) zoomed
portion of the image showing aqueous outflow structures (SC – Schlemm’s canal, TM
– trabecular meshwork) (c) en-face view of the volumetric data set (red line indicates
extracted section presented in (a)); (d) En-face visualization of scleral vessels. Scleral interface
was segmented and 1 mm thick slice was integrated.
Corneo-scleral imaging: (a) cross-sectional OCT image presenting details of the anterior chamber
angle (S – sclera, CB – ciliary body, C – cornea, I – iris); (b) zoomed
portion of the image showing aqueous outflow structures (SC – Schlemm’s canal, TM
– trabecular meshwork) (c) en-face view of the volumetric data set (red line indicates
extracted section presented in (a)); (d) En-face visualization of scleral vessels. Scleral interface
was segmented and 1 mm thick slice was integrated.
3.3. Full eye imaging and axial eye length measurement
Intraocular distances such as axial eye length or anterior chamber depth are essential for
accurate intraocular lens (IOL) calculation. The long imaging range needed to span axial eye length
was achieved by reducing the laser sweep range, trading off axial resolution.Imaging was performed on a normal subject with myopia (−7 diopters). The left eye was
scanned over 15 × 15 mm2 area with 350 × 350 axial scans with an
acquisition time of 2.6 s. Figure 11(a)
shows a rendering of the full eye from the cornea to the retina. To our knowledge, this is
the first demonstration of 3D full eye imaging using OCT.
Fig. 11
Full eye imaging with ultralong depth range OCT: (a) 3D rendering of volumetric data set
(Media 1), (b)
central cross-sectional image, (c) central B-scan extracted from data set corrected for light
refraction, (d) central depth profile with echoes from the cornea, crystalline lens and the retina
allows for determination of intraocular distances.
Full eye imaging with ultralong depth range OCT: (a) 3D rendering of volumetric data set
(Media 1), (b)
central cross-sectional image, (c) central B-scan extracted from data set corrected for light
refraction, (d) central depth profile with echoes from the cornea, crystalline lens and the retina
allows for determination of intraocular distances.Figure 11(b) shows a selected cross-sectional image
spanning the entire eye from anterior chamber and crystalline lens to the retina. The components of
the eye introduce dispersion mismatch. Therefore, the retinal signal was dispersion corrected in
post-processing by selecting the bottom part of the image using a Heaviside function, and inverse
Fourier transforming back into frequency or wavenumber. The argument of Heaviside step function was
a pixel depth position close to the retinal signal. The dispersion correction coefficients were then
selected, the spectral signal was phase corrected and Fourier transformed. The anterior segment and
retinal signals were merged to create a final B-scan where the anterior and retinal signal had
separate dispersion compensation parameters. Because the OCT beam is refracted towards the fovea as
it propagates to the retina, the effective scanning area becomes smaller with increasing depth.
Ultimately the retinal signal in Fig. 11 came from nearly a
single point on the fundus (if there were no aberrations or motion during acquisition). A star-like
pattern in the crystalline lens was visible in the 3D reconstruction (Fig. 11(a)) and en-face view (Fig. 11b)
which comes from the sutures of the lens. The depth profile in Fig.
11(c) is an averaged axial scan from the central 100 axial scans (central 10 × 10
axial scans). This enables measurement of intraocular distances after correcting for the refractive
index of each ocular component. The axial resolution was 12 μm over almost 40 mm in tissue.
The intensity peaks in Fig. 11(c) can be identified as
reflections from the anterior and posterior surfaces of the cornea, anterior and posterior
interfaces of the crystalline lens, and retina. The standard measurement of axial eye length
requires finding the distance between corneal vertex and RPE/Bruch’s membrane. The same eye
was also measured with two clinical instruments, the gold standard of partial coherence
interferometry (PCI) (IOL Master, Carl Zeiss Meditec) and an immersion A-scan ultrasound biometer
(Axis II PR, Quantel Medical). Comparison of intraocular distances measured using our OCT prototype
instrument, partial coherence interferometry and A-scan ultrasound biometer is shown in Table 3
.
Table 3
Comparison of ocular biometry measurements using VCSEL-OCT, partial coherence interferometry
and immersion ultrasound
Biometric parameter
VCSEL-OCT
PCI (IOL Master)
Immersion ultrasound (Axis II PR)
Central corneal thickness
0.52 mm
N/A
N/A
Anterior chamber depth
3.70 mm
3.74 mm
3.80 mm
Lens thickness
3.88 mm
N/A
3.94 mm
Axial eye length
25.77 mm
25.93 mm
25.81 mm
4. Discussion
Ophthalmic swept source/Fourier domain OCT using VCSEL light source technology enables the
integration of multiple imaging modalities into a single instrument. The prototype 1060 nm SS-OCT
system demonstrated in this manuscript has operating modes that enable comprehensive retinal and
anterior segment diagnostics. Along with full eye imaging capability, this instrument can be
regarded as a 3-in-1 OCT imaging device.The VCSEL light source reported here uses dual stage SOA post amplification. Sweep bandwidth was
limited because of mismatch between amplifiers, reducing axial image resolution. Dual stages were
used because the prototype 1060 nm SOAs had lower gains due to coupling losses, compared with
expected values from 1300 nm SOAs. SOAs are currently being fabricated which has higher gains and
match the VCSEL bandwidth of ~100 nm.The ultrahigh speed axial scan rate of 580 kHz demonstrated here is ~10 to 20x faster than
clinical OCT devices. High speed imaging enables acquisition of more axial scans covering larger
areas of the retina without sacrificing the ability to visualize small focal features. The
transverse scan range of the wide-field OCT images is similar to the field of view of fundus
photography. OCT fundus images correlate well with red-free fundus photography and ICG angiography.
However, OCT provides dense, isotropically sampled 3D volumetric structural information on the
retina. Volumetric data also can be used to generate arbitrary cross-sectional images or projection
en-face images. In addition, high speed imaging minimizes motion artifacts, enabling more
reproducible measurement of markers of disease progression (e.g. retinal nerve fiber layer
thickness) or ocular structures parameter (e.g. corneal thickness and refractive power). The imaging
speed is currently limited by the MEMS resonant frequency, but may be increased by using a stiffer
MEMS design. Operation at axial scan rates of up to 1.2 MHz have been demonstrated with 1300
μm MEMS-VCSEL devices [37,38]. Higher imaging speeds have also been demonstrated using multiplexed FDML light
sources [30,43].
However, imaging speeds will ultimately be limited by signal to noise constraints.The system operates at wavelengths around 1060 nm. Most retinal OCT instruments use light sources
at 840 nm and anterior segment instruments image mostly at 1310 nm, which cannot be used for retinal
imaging because of vitreal absorption. Retinal and anterior segment imaging with the same light
source is possible at 1060 nm wavelength. Although bandwidth at 1060 nm is limited by vitreal
absorption and axial resolution is reduced compared with 840 nm, there is less scattering than
shorter wavelengths. Longer wavelengths are less attenuated by ocular opacities such as cataracts
and enable imaging the choroid or anterior angle.The single longitudinal mode operation of the VCSEL gives a narrow instantaneous linewidth and
the coherence length is therefore at least 10x longer than most other swept lasers used in OCT. This
can significantly simplify scanning procedures in the clinical setting because sensitivity roll-off
with axial range is no longer an issue in patient alignment. This would be especially helpful in
intraoperative OCT. Furthermore, the long coherence length enables new imaging applications in the
anterior segment that require long depth range. Long depth range OCT imaging of the anterior eye can
provide comprehensive topographic and keratometric information. SD-OCT systems require special
full-range techniques in order to perform long range anterior segment imaging [18,50]. In contrast, SS-OCT with VCSEL light
sources can image the entire anterior segment without complex instrument modifications. However,
light collection efficiency across a deep axial range must still be optimized using a long focal
depth (confocal parameter) beam. In addition, long depth range imaging with SS-OCT requires special
attention to system design. Parasitic reflections from optical elements or fiber interfaces can act
as etalons and produce beat frequencies which generate lines or fixed pattern noise in the OCT
images.The extremely long coherence length supports an ultralong range imaging mode for full eye
visualization and axial eye length measurement. Currently, axial eye length is measured using
low-coherence interferometry and the commercial IOL Master is the clinical standard for IOL fitting.
Previous methods to extend OCT imaging range included full-range techniques or multiple reference
mirrors, however to date these methods have not shown full eye images. Although full eye OCT imaging
in small animals has been reported, imaging the human eye is challenging because of its long length.
This paper demonstrates what we believe is the first 3D high speed in vivo full eye
imaging in humans. Measurements of intraocular distances using SS-OCT were compared with clinical
optical and ultrasound biometry, and good correspondence was found. Volumetric data sets allow
measurement of intraocular distances as well as keratometry values. These parameters are important
for calculating IOL power. These results are preliminary and it is important to note here that
precise calibration of the imaging depth range has an impact on the reproducibility of intraocular
measurements and their correlation with current clinical standard devices.Another advantage of SS-OCT using VCSEL light sources is its adjustability, making it possible to
dynamically configure the axial scan rate, resolution and imaging range to utilize available A/D
bandwidth. Depth range adjustment by changing the sweep range was previously demonstrated using an
FDML-based SS-OCT system, however the sweep repetition rate in FDML lasers must be equal to a
harmonic of the cavity round trip [51]. VCSEL technology
enables adjustment of both sweep repetition rate and sweep range. Both the forward and backward
sweeps can be used to achieve almost 100% duty cycle without buffering/multiplexing. Buffering can
also be used to obtain unidirectional sweeps or multiply the sweep repetition rate, in a manner
similar to FDML lasers. However, there are constraints on the MEMS drive amplitude which limit the
short duty cycle required for high multiplexing of the sweep repetition rate. In addition, VCSEL
stability is reduced when operating at sweep rates away from the resonance frequency. For this
reason, dual channel acquisition, with an MZI channel as a calibration trace for each sweep was
required. When the VCSEL operated at resonance frequency (290 kHz), it was possible to use a single
calibration trace, acquired before OCT data acquisition, to calibrate all sweeps in the OCT data
set. In the future, it should be possible to use direct optical clocking of the A/D to acquire OCT
signals which are calibrated in frequency or wavenumber at high 1 GSPS speeds. This avoids the
computationally expensive step of MZI recalibration and reduces data rate and memory requirements.
The long coherence length of the VCSEL should also provide a stable optical clock signal. Selected
data in preliminary studies was acquired using optical clocking with a 400 MSPS card, but currently
available GHz bandwidth A/D cards do not support optically clocked acquisition. Therefore, the data
reported here was acquired using dual channel acquisition on both cards for consistency.
5. Summary and conclusions
To summarize, this manuscript demonstrates swept source OCT utilizing VCSEL technology for
ophthalmic imaging at 1060 nm. The MEMS tunable VCSEL has a long coherence length with adjustable
sweep repetition rates (50 kHz-580 kHz axial scan rate) and adjustable frequency/wavelength sweep
range. This enables adjustment of the imaging speed, resolution and depth range and integration of
multiple ophthalmic applications into a single instrument. The instrument can perform ultrahigh
speed retinal and choroidal imaging, long depth range anterior segment imaging and ultralong depth
range full eye imaging. Wide field retinal and choroidal imaging was demonstrated. Comprehensive
volumetric data of the anterior segment from the cornea to the posterior crystalline lens surface
were presented. Full eye 3D-OCT imaging from the cornea to the retina was also demonstrated. Swept
source OCT using VCSEL technology promises to enable the development of integrated OCT instruments
which reduce patient chair/visit time and save clinical space. The high performance of SS-OCT using
VCSEL technology promises to enable new applications for OCT imaging of retinal, optic nerve head
and choroidal pathologies as well as ocular biometry.
Authors: Wolfgang Wieser; Benjamin R Biedermann; Thomas Klein; Christoph M Eigenwillig; Robert Huber Journal: Opt Express Date: 2010-07-05 Impact factor: 3.894
Authors: Bartłomiej J Kałuzny; Jakub J Kaluzny; Anna Szkulmowska; Iwona Gorczyńska; Maciej Szkulmowski; Tomasz Bajraszewski; Piotr Targowski; Andrzej Kowalczyk Journal: Ophthalmic Physiol Opt Date: 2006-03 Impact factor: 3.117
Authors: Thomas Klein; Wolfgang Wieser; Christoph M Eigenwillig; Benjamin R Biedermann; Robert Huber Journal: Opt Express Date: 2011-02-14 Impact factor: 3.894
Authors: WooJhon Choi; Eric M Moult; Nadia K Waheed; Mehreen Adhi; ByungKun Lee; Chen D Lu; Talisa E de Carlo; Vijaysekhar Jayaraman; Philip J Rosenfeld; Jay S Duker; James G Fujimoto Journal: Ophthalmology Date: 2015-10-17 Impact factor: 12.079
Authors: WooJhon Choi; Benjamin Potsaid; Vijaysekhar Jayaraman; Bernhard Baumann; Ireneusz Grulkowski; Jonathan J Liu; Chen D Lu; Alex E Cable; David Huang; Jay S Duker; James G Fujimoto Journal: Opt Lett Date: 2013-02-01 Impact factor: 3.776
Authors: Ireneusz Grulkowski; Jonathan J Liu; Benjamin Potsaid; Vijaysekhar Jayaraman; James Jiang; James G Fujimoto; Alex E Cable Journal: Opt Lett Date: 2013-03-01 Impact factor: 3.776