Validity and repeatability of anterior chamber depth measurements with Pentacam and Orbscan.

Birgit Lackner1, Gerald Schmidinger, Christian Skorpik.   

Abstract

PURPOSE: The purpose of this study was to determine the validity and repeatability of anterior chamber depth (ACD) measurements obtained with a novel rotating Scheimpflug camera (Pentacam; Oculus Optikgeräte GmbH, Wetzlar, Germany) to scanning slit topography (Orbscan; Bausch & Lomb, Rochester, NY).
METHODS: ACD in 60 healthy eyes was measured twice by two independent observers with each modality in random order. A total of eight measurements was performed on each eye. The mean differences between measurements, between observers, and between modalities were calculated, and 95% limits of agreement (LoA) were given as mean +/- 1.96 * standard deviation (SD) of the mean.
RESULTS: The mean ACD values as determined with the different modalities (+/- SD) were 3.18 +/- 0.38 mm for Pentacam imaging and 3.23 +/- 0.40 for Orbscan. The mean difference between the two modalities was 0.047 mm (LoA: 0.176 to -0.081 mm). Within each modality, the limits of agreement of the differences between individual measurements and between different observers were within 0.073 mm or 4% of total ACD. The total variance in the observed data was 0.145. The variance resulting from interindividual differences in ACD was 85%, as a result of the two different modalities 15%, as a result of different observers 0.007%, and as a result of different measurements 0.007%.
CONCLUSIONS: In the assessment of normal eyes, the differences of ACD values measured with Orbscan and Pentacam were within clinically acceptable levels, and inter- and intraobserver variability was considerably below clinically significant levels. Thus, these two modalities can be regarded as interchangeable.

Mesh:

Year:  2005        PMID: 16189497     DOI: 10.1097/01.opx.0000177804.53192.15

Source DB:  PubMed          Journal:  Optom Vis Sci        ISSN: 1040-5488            Impact factor:   1.973


  36 in total

1.  Measurements of anterior segment parameters using three different non-contact optical devices in keratoconus patients.

Authors:  Ahmet Taylan Yazıcı; Gökhan Pekel; Ercüment Bozkurt; Yusuf Yıldırım; Evre Pekel; Ahmet Demirok; Omer Faruk Yılmaz
Journal:  Int J Ophthalmol       Date:  2013-08-18       Impact factor: 1.779

2.  Comparison of anterior segment measurements using rotating Scheimpflug imaging and partial coherence interferometry.

Authors:  Mohammad Soleimani; Hassan Hashemi; Shiva Mehravaran; Mehdi Khabazkhoob; Mohammad Hassan Emamian; Mohammad Shariati; Akbar Fotouhi
Journal:  Int J Ophthalmol       Date:  2013-08-18       Impact factor: 1.779

3.  Anterior chamber depth in normal subjects by rotating scheimpflug imaging.

Authors:  Matthew T Feng; Michael W Belin; Renato Ambrósio; Satinder P S Grewal; Wang Yan; Mohamed S Shaheen; Charles McGhee; Naoyuki Maeda; Tobias H Neuhann; H Burkhard Dick; Saleh A Alageel; Andreas Steinmueller
Journal:  Saudi J Ophthalmol       Date:  2011-04-30

4.  Assessment of the anterior chamber parameters after laser iridotomy in primary angle close suspect using Pentacam and gonioscopy.

Authors:  Alireza Esmaeili; Behzad Barazandeh; Sina Ahmadi; Alireza Haghi; Seyed Mahdi Ahmadi Hosseini; Fereshteh Abolbashari
Journal:  Int J Ophthalmol       Date:  2013-10-18       Impact factor: 1.779

5.  [Scheimpflug and topography systems in ophthalmologic diagnostics].

Authors:  G U Auffarth; A F M Borkenstein; A Ehmer; A Mannsfeld; T M Rabsilber; M P Holzer
Journal:  Ophthalmologe       Date:  2008-09       Impact factor: 1.059

6.  Comparison of corneal measurements in keratoconic eyes using rotating Scheimpflug camera and scanning-slit topography.

Authors:  Mohammad Naderan; Saeed Shoar; Morteza Naderan; Mohammad Amin Kamaleddin; Mohammad Taher Rajabi
Journal:  Int J Ophthalmol       Date:  2015-04-18       Impact factor: 1.779

7.  Repeatability and comparability of anterior segment biometry obtained by the Sirius and the Pentacam analyzers.

Authors:  Paola De la Parra-Colín; Manuel Garza-León; Tonatiuh Barrientos-Gutierrez
Journal:  Int Ophthalmol       Date:  2013-04-25       Impact factor: 2.031

8.  Detection of occludable angles with the Pentacam and the anterior segment optical coherence tomography.

Authors:  Samin Hong; Jeong-Ho Yi; Sung Yong Kang; Gong Je Seong; Chan Yun Kim
Journal:  Yonsei Med J       Date:  2009-08-19       Impact factor: 2.759

9.  Comparison of central corneal thickness and anterior chamber depth measurements using three imaging technologies in normal eyes and after phakic intraocular lens implantation.

Authors:  Muriël Doors; Lars P J Cruysberg; Tos T J M Berendschot; John de Brabander; Frenne Verbakel; Carroll A B Webers; Rudy M M A Nuijts
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2009-04-30       Impact factor: 3.117

10.  Repeatability of corneal parameters with Pentacam after laser in situ keratomileusis.

Authors:  Rajeev Jain; Grewal Dilraj; Satinder Pal Singh Grewal
Journal:  Indian J Ophthalmol       Date:  2007 Sep-Oct       Impact factor: 1.848

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.