Scheimpflug imaging to determine intraocular lens power in vivo.

Stephen J Turner1, Edward J K Lee, Victor Hu, Emma J Hollick.   

Abstract

PURPOSE: To assess the feasibility of determining intraocular lens (IOL) power by measurement of the central optic thickness using clinically available Scheimpflug imaging (Pentacam, Oculus).
SETTING: King's College Hospital Ophthalmology Department, London, United Kingdom.
METHODS: Sixty-seven eyes were assessed 1 month after uneventful phacoemulsification with in-the-bag implantation of AcrySof MA60AC IOLs (Alcon). The correlation between IOL thickness measurement and IOL power was calculated. Repeatability of central optic thickness measurement was determined from 10 successive scans of 4 patients.
RESULTS: Within-subject standard deviation increased with the subject mean. The coefficient of variability was 1.4%. Measured lens thickness was highly correlated with lens power (R(2) = 0.94, P<.001). Over the measured range, 95% confidence intervals varied between +/-0.83 diopters (D) and +/-0.92 D.
CONCLUSIONS: Central IOL thickness measurements with the Pentacam Scheimpflug camera were highly repeatable and closely correlated with the known IOL power. The IOL power, calculated from a regression equation, is likely to be less than +/-1.00 D away from the actual power. Approximate in vivo IOL power determination is feasible with clinically available Scheimpflug imaging. This could be applied clinically in cases of unexplained postoperative refractive error.

Entities:  

Mesh:

Year:  2007        PMID: 17531700     DOI: 10.1016/j.jcrs.2007.02.023

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  4 in total

Review 1.  "Numbers sometimes lie"- refractive surprise following IOL mislabeling by the manufacturer.

Authors:  Marija Antičić; Navid Ardjomand; Stephanie Sarny; Jakob Schweighofer; Yosuf El-Shabrawi
Journal:  Eye (Lond)       Date:  2019-02-13       Impact factor: 3.775

2.  Anterior segment study with the Pentacam Scheimpflug camera in refractive surgery candidates.

Authors:  Masih Hashemi; Khalil Ghasemi Falavarjani; Gholam Hosseyn Aghai; Kaveh Abri Aghdam; Arzhang Gordiz
Journal:  Middle East Afr J Ophthalmol       Date:  2013 Jul-Sep

3.  In Vivo Intraocular Lens Thickness Measurement and Power Estimation Using Optical Coherence Tomography.

Authors:  Ehsan Barzanouni; Diba Idani; Farideh Sharifipour
Journal:  J Ophthalmic Vis Res       Date:  2022-08-15

Review 4.  Biometric Measurement of Anterior Segment: A Review.

Authors:  Bin Liu; Chengwei Kang; Fengzhou Fang
Journal:  Sensors (Basel)       Date:  2020-07-31       Impact factor: 3.576

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.