Evaluation of corneal thickness using a Scheimpflug-Placido disk corneal analyzer and comparison with ultrasound pachymetry in eyes after laser in situ keratomileusis.

Jinhai Huang1, Weicong Lu, Giacomo Savini, Liang Hu, Chao Pan, Jing Wang, Weina Tan, Jia Chen, Qinmei Wang.   

Abstract

PURPOSE: To evaluate the repeatability and reproducibility of corneal thickness measurements in post-laser in situ keratomileusis (LASIK) eyes using a rotating Scheimpflug camera combined with a Placido disk corneal topographer (Sirius) and compare the results with those of ultrasound (US) pachymetry.
SETTING: Eye Hospital of Wenzhou Medical College, Wenzhou, China.
DESIGN: Comparative evaluation of a diagnostic test or technology.
METHODS: Patients were examined 3 times with the Scheimpflug-Placido topographer by 2 examiners. The central pupil corneal thickness (CTpupil), apical corneal thickness (CTapex), and thinnest corneal thickness (CTthinnest) were recorded. After noncontact examinations, US pachymetry was used to obtain the central corneal thickness (CCT).
RESULTS: The Scheimpflug-Placido topographer showed high intraoperator repeatability as indicated by a test-retest repeatability of less than 8.5 μm for CTpupil, CTapex, and CTthinnest, The coefficients of variation (CoV) were less than 0.7%, and the intraclass correlation coefficient was higher than 0.99. Excellent results were also obtained for interoperator reproducibility. All CoVs were less than 0.5%. The 95% limits of agreement between the Scheimpflug-Placido measurement and the US pachymetry measurements were narrow (-16.62 to 12.44 μm for CTpupil versus US pachymetry CCT; -17.49 to 12.16 μm for CTapex versus US pachymetry CCT; -18.59 to 10.90 μm for CTthinnest versus US pachymetry CCT).
CONCLUSIONS: The Scheimpflug-Placido topographer showed excellent intraoperator repeatability and interoperator reproducibility of CTpupil, CTapex, and CTthinnest measurements in post-LASIK eyes. The CCT measurements obtained using the device were in high agreement with those obtained by US pachymetry, suggesting that the 2 devices are interchangeable. FINANCIAL DISCLOSURE: No author has a financial or proprietary interest in any material or method mentioned. Crown
Copyright © 2013. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2013        PMID: 23680632     DOI: 10.1016/j.jcrs.2013.01.038

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  10 in total

1.  Difference in normal corneal thickness and curvature between Mongolian and Han nationalities.

Authors:  Hai-Xia Zhao; Li Zhang; Wen-Ying Guan
Journal:  Int J Ophthalmol       Date:  2015-04-18       Impact factor: 1.779

2.  Topographic indices and pachymetry in healthy adolescents obtained with Sirius topographer.

Authors:  Emilia Cantera; Magdalena Cortes; Roberto Sacco; Gabriele Vestri; Alessandra Micera
Journal:  Int Ophthalmol       Date:  2017-10-30       Impact factor: 2.031

3.  Agreement Between Predicted and Actual Measured Ablation Depth After FS-LASIK Using Different Rotating Scheimpflug Cameras and OCT.

Authors:  Hao Chen; Ziqian Wang; Kunke Li; Yiran Wang; Xin Li; Lan Du; Meimin Lin; Giacomo Savini; Qinmei Wang; Ayong Yu; Sisi Chen
Journal:  Front Med (Lausanne)       Date:  2022-05-19

4.  Effects of posterior corneal astigmatism on the accuracy of AcrySof toric intraocular lens astigmatism correction.

Authors:  Bin Zhang; Jing-Xue Ma; Dan-Yan Liu; Cong-Rong Guo; Ying-Hua Du; Xiu-Jin Guo; Yue-Xian Cui
Journal:  Int J Ophthalmol       Date:  2016-09-18       Impact factor: 1.779

5.  Comparison of two different scheimpflug devices in the detection of keratoconus, regular astigmatism, and healthy corneas.

Authors:  David Finis; Bernhard Ralla; Maria Karbe; Maria Borrelli; Stefan Schrader; Gerd Geerling
Journal:  J Ophthalmol       Date:  2015-04-14       Impact factor: 1.909

6.  Comparison of Anterior Segment Measurements with Scheimpflug/Placido Photography-Based Topography System and IOLMaster Partial Coherence Interferometry in Patients with Cataracts.

Authors:  Jinhai Huang; Na Liao; Giacomo Savini; Fangjun Bao; Ye Yu; Weicong Lu; Qingjie Hu; Qinmei Wang
Journal:  J Ophthalmol       Date:  2014-10-23       Impact factor: 1.909

7.  Sirius Scheimpflug-Placido versus ultrasound pachymetry for central corneal thickness: meta-analysis.

Authors:  Yili Jin; Colm McAlinden; Yong Sun; Daizong Wen; Yiran Wang; Jinjin Yu; Ke Feng; Benhao Song; Qinmei Wang; Shihao Chen; Jinhai Huang
Journal:  Eye Vis (Lond)       Date:  2021-02-18

8.  Central corneal thickness of Iraqi population in relation to age, gender, refractive errors, and corneal curvature: a hospital-based cross-sectional study.

Authors:  Yasir Jawad Kadhim; Qasim K Farhood
Journal:  Clin Ophthalmol       Date:  2016-11-25

9.  Comparison of anterior segment measurements using Sirius Topographer® and Nidek Axial Length-Scan® with assessing repeatability in patients with cataracts.

Authors:  Resat Duman; Ersan Çetinkaya; Rahmi Duman; Mustafa Dogan; Mehmet Cem Sabaner
Journal:  Indian J Ophthalmol       Date:  2018-03       Impact factor: 1.848

Review 10.  Biometric Measurement of Anterior Segment: A Review.

Authors:  Bin Liu; Chengwei Kang; Fengzhou Fang
Journal:  Sensors (Basel)       Date:  2020-07-31       Impact factor: 3.576

  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.