Literature DB >> 22541828

Intrasubject repeatability of corneal morphology measurements obtained with a new Scheimpflug photography-based system.

Raúl Montalbán1, David P Piñero, Jaime Javaloy, Jorge L Alió.   

Abstract

PURPOSE: To evaluate in normal healthy eyes the intrasubject repeatability of anterior and posterior corneal curvature measurements and other anatomic anterior segment measurements obtained with a new topography system combining Scheimpflug photography and Placido-disk technology.
SETTING: Vissum Corp., Alicante, Spain.
DESIGN: Evaluation of technology.
METHODS: All eyes received a comprehensive ophthalmologic examination including anterior segment analysis with the Sirius system. Three consecutive measurements were performed with the device to assess the intrasubject repeatability of the following parameters: anterior and posterior corneal curvature and shape factor, white-to-white (WTW) corneal diameter, central and minimum corneal thickness, and anterior chamber depth (ACD). The within-subject standard deviation (S(w)) and intraclass correlation coefficient (ICC) were calculated.
RESULTS: This study included 117 eyes of 117 subjects (mean age 42 years; range 7 to 80 years). For anterior and posterior corneal curvatures, the S(w) was 0.04 mm or lower and the ICC was higher than 0.990. For shape-factor measurements, the S(w) was below 0.08 in all cases and ICC values ranged between 0.909 and 0.994. Significantly larger S(w) values were found for the anterior and posterior shape factor calculated for 8.0 mm compared with 4.5 mm (P<.01). An S(w) value below 3 μm was observed for the central and minimum thickness, with ICC values close to 1. The mean S(w) for ACD and WTW was below 0.1 mm.
CONCLUSION: In healthy eyes, the new topography system provided repeatable measurements of several anterior segment parameters, including anterior and posterior curvature and pachymetry.
Copyright © 2012 ASCRS and ESCRS. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2012        PMID: 22541828     DOI: 10.1016/j.jcrs.2011.12.029

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  19 in total

1.  Comparative analysis of the relationship between anterior and posterior corneal shape analyzed by Scheimpflug photography in normal and keratoconus eyes.

Authors:  Raúl Montalbán; Jorge L Alio; Jaime Javaloy; David P Piñero
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2013-01-20       Impact factor: 3.117

2.  Anatomical characterization of central, apical and minimal corneal thickness.

Authors:  Federico Saenz-Frances; Martha Cecilia Bermúdez-Vallecilla; Lara Borrego-Sanz; Luis Jañez; José María Martinez-de-la-Casa; Laura Morales-Fernandez; Enrique Santos-Bueso; Julián Garcia-Sanchez; Julián Garcia-Feijoo
Journal:  Int J Ophthalmol       Date:  2014-08-18       Impact factor: 1.779

3.  Corneal topographic alterations after selective laser trabeculoplasty.

Authors:  Kursat Atalay; Ahmet Kirgiz; Kubra Serefoglu Cabuk; Havva Erdogan Kaldirim
Journal:  Int Ophthalmol       Date:  2016-09-15       Impact factor: 2.031

4.  Validation of corneal topographic and aberrometric measurements obtained by color light-emitting diode reflection topography in healthy eyes.

Authors:  David P Piñero; Ainhoa Molina-Martín; Vicent J Camps; Dolores de Fez; María Teresa Caballero
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2019-09-03       Impact factor: 3.117

5.  Corneal Morphologic Characteristics in Patients With Down Syndrome.

Authors:  Jorge L Alio; Alfredo Vega-Estrada; Pablo Sanz; Amr A Osman; Ahmed M Kamal; Amr Mamoon; Hany Soliman
Journal:  JAMA Ophthalmol       Date:  2018-09-01       Impact factor: 7.389

6.  Comparison of Anterior Segment Measurements with Scheimpflug/Placido Photography-Based Topography System and IOLMaster Partial Coherence Interferometry in Patients with Cataracts.

Authors:  Jinhai Huang; Na Liao; Giacomo Savini; Fangjun Bao; Ye Yu; Weicong Lu; Qingjie Hu; Qinmei Wang
Journal:  J Ophthalmol       Date:  2014-10-23       Impact factor: 1.909

7.  Geometrical custom modeling of human cornea in vivo and its use for the diagnosis of corneal ectasia.

Authors:  Francisco Cavas-Martínez; Daniel G Fernández-Pacheco; Ernesto De la Cruz-Sánchez; José Nieto Martínez; Francisco J Fernández Cañavate; Alfredo Vega-Estrada; Ana B Plaza-Puche; Jorge L Alió
Journal:  PLoS One       Date:  2014-10-17       Impact factor: 3.240

8.  Polar Value Analysis of Low to Moderate Astigmatism with Wavefront-Guided Sub-Bowman Keratomileusis.

Authors:  Pisong Yan; Zhiyu Du; Yu Zhang
Journal:  J Ophthalmol       Date:  2017-08-02       Impact factor: 1.909

9.  Single session, intrauser repeatability of anterior chamber biometric and corneal pachy-volumetric parameters using a new Scheimpflug+Placido device.

Authors:  Gaurav Prakash; Dhruv Srivastava
Journal:  J Optom       Date:  2015-09-01

10.  Comparison of the Optical Quality between Small Incision Lenticule Extraction and Femtosecond Laser LASIK.

Authors:  Ying Jin; Yan Wang; Lulu Xu; Tong Zuo; Hua Li; Rui Dou; Jiamei Zhang
Journal:  J Ophthalmol       Date:  2016-11-10       Impact factor: 1.909

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.