Accuracy of Orbscan optical pachymetry in corneas with haze.

Francesco Boscia1, Maria Gabriella La Tegola, Giovanni Alessio, Carlo Sborgia.   

Abstract

PURPOSE: To compare corneal thickness measurements with the Orbscan instrument (Orbtek, Inc.) and an ultrasonic pachymeter (Alcon Surgical) in corneas with haze.
SETTING: Department of Ophthalmology and Otolaryngology, University of Bari, Bari, Italy.
METHODS: Six eyes of 6 patients having phototherapeutic keratectomy for treatment of grade 2+ haze were assessed preoperatively and postoperatively by Orbscan optical pachymetry and ultrasonic pachymetry. Contralateral transparent untreated corneas of the same patients were used as controls. Two Orbscan and 5 ultrasound pachymetry measurements were performed at each visit and the mean values recorded. The accuracy and repeatability of the measurements were evaluated.
RESULTS: Preoperatively, the median corneal thickness in the 6 eyes with haze was 512.7 microm by ultrasonic pachymetry and 375.0 microm by optical pachymetry. The mean difference of 143.1 microm was statistically significant (Wilcoxon signed rank test, P= .0277). Postoperatively, the median pachymetry was 439.0 microm and 433.0 microm, respectively; the mean difference was 10.7 microm (P= .0277). In the control group, the median ultrasonic pachymetry was 530.9 microm and the median optical pachymetry, 529.0 microm; the mean difference was 4.7 microm (P= .1730). Precision was good in all groups; the standard deviation was less than 6.0 microm.
CONCLUSIONS: The Orbscan system showed a decreased accuracy in measuring corneal thickness when clinically significant haze was present. The repeatability of the measures was good, suggesting that the cause of error lies in the optical acquisition process.

Entities:  

Mesh:

Year:  2002        PMID: 11821206     DOI: 10.1016/s0886-3350(01)01162-2

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  19 in total

1.  Comparison of corneal thickness measurements using Orbscan II, non-contact specular microscopy, and ultrasonic pachymetry in eyes after laser in situ keratomileusis.

Authors:  K Kawana; T Tokunaga; K Miyata; F Okamoto; T Kiuchi; T Oshika
Journal:  Br J Ophthalmol       Date:  2004-04       Impact factor: 4.638

2.  Comparison of ultrasonic pachymetry with Orbscan in corneal haze.

Authors:  Rana Altan-Yaycioglu; Aysel Pelit; Yonca Aydin Akova
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2007-06-22       Impact factor: 3.117

3.  Comparison between central corneal thickness measurements by oculus pentacam and ultrasonic pachymetry.

Authors:  Hani S Al-Mezaine; Saleh A Al-Amro; Dustan Kangave; Abdulkareem Sadaawy; Taher A Wehaib; Saleh Al-Obeidan
Journal:  Int Ophthalmol       Date:  2007-09-26       Impact factor: 2.031

4.  Intraocular lens power calculation after myopic and hyperopic laser vision correction using optical coherence tomography.

Authors:  Maolong Tang; Li Wang; Douglas D Koch; Yan Li; David Huang
Journal:  Saudi J Ophthalmol       Date:  2012-01

5.  Comparison of central corneal thickness measurements by Orbscan II and Pentacam after corneal refractive surgery.

Authors:  Jumpei Matsuda; Osamu Hieda; Shigeru Kinoshita
Journal:  Jpn J Ophthalmol       Date:  2008-09-05       Impact factor: 2.447

6.  Intraocular lens power calculation after previous myopic laser vision correction based on corneal power measured by Fourier-domain optical coherence tomography.

Authors:  Maolong Tang; Li Wang; Douglas D Koch; Yan Li; David Huang
Journal:  J Cataract Refract Surg       Date:  2012-04       Impact factor: 3.351

7.  Effect of 2% fluorescein on Scheimpflug central corneal thickness measurements.

Authors:  Stella Briggs; Marwa Bin Moammar
Journal:  Int J Ophthalmol       Date:  2016-02-18       Impact factor: 1.779

8.  Implantable collamer lens in a case of corneal scar with anisometropic amblyopia in an adult: an expanded indication.

Authors:  Gaurav Prakash; Kavitha Avadhani; Jay Kalliath; Dhruv Srivastava
Journal:  BMJ Case Rep       Date:  2015-03-05

9.  Corneal pachymetry mapping with high-speed optical coherence tomography.

Authors:  Yan Li; Raj Shekhar; David Huang
Journal:  Ophthalmology       Date:  2006-05       Impact factor: 12.079

10.  Repeatability of corneal parameters with Pentacam after laser in situ keratomileusis.

Authors:  Rajeev Jain; Grewal Dilraj; Satinder Pal Singh Grewal
Journal:  Indian J Ophthalmol       Date:  2007 Sep-Oct       Impact factor: 1.848

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.