To assess the repeatability and reproducibility of AL-Scan in agreement with those by the IOLMaster in healthy children, two skilled operators measured ocular parameters in 58 children. The parameters included keratometry (K) values, anterior chamber depth (ACD), axial length (AL), central corneal thickness (CCT), pupil diameter (PD), and corneal diameter (CD). The cohort comprised of 32 boys and 26 girls. The AL-Scan measurements showed high repeatability, as the test-retest repeatability (TRT) values of AL, CCT, ACD, Kf, Ks, Km, CD, and PD were 0.09 mm, 5.1 μm, 0.04 mm, 0.28 D, 0.24 D, 0.21 D, 0.39 mm, and 0.22 mm, respectively. The within-subject coefficient of variation (CoV) was low (<0.35%) and the intraclass correlation coefficients (ICC) of all parameters were >0.85. The interobserver reproducibility was excellent with low values of TRT and ICC > 0.95. The CoV of AL, CCT, ACD, and K was <0.22%. The 95% limits of agreement between the AL-Scan and the IOLMaster were narrow for all parameters except for CD. The repeatability and reproducibility of the new biometer, Al-Scan, was excellent for all parameters and can be routinely used in children to measure the biometric values.Entities:
Mesh:
Year: 2018 PMID: 29358637 PMCID: PMC5778016 DOI: 10.1038/s41598-018-19605-6
Source DB: PubMed Journal: Sci Rep ISSN: 2045-2322 Impact factor: 4.379
Intraobserver repeatability outcomes for biometric measurements obtained using AL-Scan partial coherence interferometry in children.
| Parameter | observer | Mean ± SD | Sw | TRT | CoV (%) | ICC |
|---|---|---|---|---|---|---|
| AL (mm) | 1st | 23.83 ± 1.22 | 0.03 | 0.09 | 0.14 | 0.999 |
| 2nd | 23.82 ± 1.22 | 0.03 | 0.09 | 0.14 | 0.999 | |
| CCT (μm) | 1st | 547.40 ± 31.97 | 1.81 | 5.01 | 0.33 | 0.997 |
| 2nd | 547.42 ± 31.98 | 1.79 | 4.95 | 0.33 | 0.997 | |
| ACD (mm) | 1st | 3.81 ± 0.20 | 0.01 | 0.04 | 0.34 | 0.995 |
| 2nd | 3.82 ± 0.20 | 0.01 | 0.03 | 0.30 | 0.997 | |
| Kf (D) | 1st | 43.76 ± 1.54 | 0.10 | 0.27 | 0.23 | 0.996 |
| 2nd | 43.77 ± 1.54 | 0.10 | 0.28 | 0.23 | 0.997 | |
| Ks (D) | 1st | 42.47 ± 1.43 | 0.09 | 0.24 | 0.20 | 0.996 |
| 2nd | 42.50 ± 1.44 | 0.08 | 0.23 | 0.19 | 0.996 | |
| Km (D) | 1st | 43.11 ± 1.44 | 0.08 | 0.21 | 0.17 | 0.997 |
| 2nd | 43.13 ± 1.44 | 0.08 | 0.21 | 0.18 | 0.997 | |
| Astigmatism | 1st | 1.28 ± 0.80 | 0.11 | 0.30 | 8.54 | 0.982 |
| 2nd | 1.28 ± 0.83 | 0.11 | 0.29 | 8.28 | 0.984 | |
| J0 (D) | 1st | −0.61 ± 0.40 | 0.05 | 0.15 | — | 0.981 |
| 2nd | −0.61 ± 0.41 | 0.06 | 0.15 | — | 0.982 | |
| J45 (D) | 1st | 0.06 ± 0.21 | 0.06 | 0.16 | — | 0.928 |
| 2nd | 0.06 ± 0.20 | 0.05 | 0.13 | — | 0.946 | |
| CD (mm) | 1st | 12.08 ± 0.35 | 0.14 | 0.38 | 1.12 | 0.862 |
| 2nd | 12.05 ± 0.35 | 0.14 | 0.39 | 1.15 | 0.859 | |
| PD | 1st | 7.94 ± 0.59 | 0.08 | 0.22 | 1.00 | 0.982 |
| 2nd | 7.96 ± 0.57 | 0.07 | 0.20 | 0.90 | 0.984 |
AL = axial length, CCT = central corneal thickness, ACD = anterior chamber depth, Kf = flattest keratometry, Ks = steepest keratometry, Km = mean keratometry, CD = corneal diameter, SD = standard deviation, Sw = within-subject standard deviation, TRT = test-retest repeatability (2.77 Sw), CoV = within-subject coefficient of variation, ICC = intraclass correlation coefficient.
Interobserver reproducibility outcomes for biometric measurements obtained using AL-Scan partial coherence interferometry in children.
| Parameter | Mean difference ± SD | Sw | TRT | CoV (%) | ICC |
|---|---|---|---|---|---|
| AL (mm) | 0.01 ± 0.03 | 0.02 | 0.06 | 0.09 | 1.000 |
| CCT (μm) | −0.02 ± 1.51 | 1.06 | 2.93 | 0.19 | 0.999 |
| ACD (mm) | 0.00 ± 0.01 | 0.01 | 0.02 | 0.22 | 0.998 |
| Kf (D) | −0.02 ± 0.09 | 0.08 | 0.21 | 0.17 | 0.998 |
| Ks (D) | −0.02 ± 0.11 | 0.06 | 0.18 | 0.15 | 0.998 |
| Km (D) | −0.02 ± 0.08 | 0.06 | 0.16 | 0.13 | 0.998 |
| Astigmatism | 0.01 ± 0.12 | 0.09 | 0.24 | 6.68 | 0.989 |
| J0 (D) | 0.00 ± 0.06 | 0.04 | 0.11 | 0.990 | |
| J45 (D) | 0.00 ± 0.05 | 0.03 | 0.09 | 0.976 | |
| CD (mm) | 0.03 ± 0.10 | 0.07 | 0.21 | 0.62 | 0.955 |
| PD | −0.01 ± 0.11 | 0.08 | 0.22 | 1.01 | 0.981 |
AL = axial length, CCT = central corneal thickness, ACD = anterior chamber depth, Kf = flattest keratometry, Ks = steepest keratometry, Km = mean keratometry, CD = corneal diameter, SD = standard deviation, Sw = within-subject standard deviation, TRT = test-retest repeatability (2.77 Sw), CoV = within-subject coefficient of variation, ICC = intraclass correlation coefficient.
The mean difference, paired T-test, and 95% limits of agreement (LoA) for differences between the AL-Scan and the IOLMaster partial coherence interferometry in children.
| Device Pairings | Mean Difference ± SD | 95% LoA | |
|---|---|---|---|
| AL (mm) | 0.00 ± 0.03 | 0.346 | −0.05 to 0.05 |
| ACD (mm) | 0.17 ± 0.06 | 0.000 | 0.05 to 0.29 |
| Kf (D) | −0.08 ± 0.14 | 0.000 | −0.36 to 0.20 |
| Ks (D) | −0.19 ± 0.17 | 0.000 | −0.52 to 0.14 |
| Km (D) | −0.13 ± 0.12 | 0.000 | −0.37 to 0.10 |
| Astigmatism | −0.11 ± 0.20 | 0.000 | −0.50 to 0.27 |
| J0 (D) | 0.02 ± 0.19 | 0.537 | −0.37 to 0.40 |
| J45 (D) | −0.01 ± 0.12 | 0.608 | −0.25 to 0.24 |
| CD (mm) | −0.42 ± 0.35 | 0.000 | −1.10 to 0.26 |
AL = axial length, ACD = anterior chamber depth, Kf = flattest keratometry, Ks = steepest keratometry, Km = mean keratometry, CD = corneal diameter, SD = standard deviation.
Figure 1Bland–Altman graphs for pairwise comparisons between the AL-Scan and IOLMaster measuring the axial length in children. The mean difference and 95% limits of agreement are indicated as solid and dashed lines, respectively.
Figure 6Bland–Altman graphs for pairwise comparisons between the AL-Scan and IOLMaster measuring the white-to-white in children. The mean difference and 95% limits of agreement are indicated as solid and dashed lines, respectively.