Axial length measurement in eyes with diabetic macular edema: a-scan ultrasound versus IOLMaster.

Liat Attas-Fox1, David Zadok, Yariv Gerber, Yair Morad, Eva Eting, Nathanael Benamou, Eran Pras, Ori Segal, Isaac Avni, Yaniv Barkana.   

Abstract

OBJECTIVE: To compare axial length (AL) measurements obtained with ultrasound (US) and IOLMaster (software version 3.01.0294; Carl Zeiss Meditec, Jena, Germany) in eyes with diabetic clinically significant macular edema (CSME) and to examine the correlation between measurement difference and foveal thickness.
DESIGN: Prospective nonrandomized clinical study. PARTICIPANTS: Twenty-three healthy eyes of 23 participants and 21 eyes with CSME of 21 diabetic outpatients.
METHODS: The authors prospectively measured AL using US and IOLMaster in eyes of both groups. Foveal thickness was measured with optical coherence tomography. The difference between both methods was assessed using the t test, and its correlation to foveal thickness was evaluated by Pearson coefficient. MAIN OUTCOME MEASURE: Difference in AL measurements between IOLMaster and US.
RESULTS: Mean+/-standard deviation foveal thickness in healthy and CSME eyes was 201+/-26 microm (range, 156-240) and 475+/-106 microm (range, 331-758), respectively. Mean AL by US and IOLMaster in healthy eyes was 23.17+/-1.07 mm (range, 21.17-25.18) and 23.25+/-1.10 mm (range, 21.17-25.26), respectively (P = 0.51); in CSME eyes, it was 23.02+/-0.97 mm (range, 21.53-24.81) and 23.27+/-0.924 mm (range, 21.9-25.13), respectively (P<0.001). Correlation between AL difference and foveal thickness was poor in both healthy (r = 0.18, P = 0.40) and CSME (r = 0.02, P = 0.93) eyes.
CONCLUSIONS: Axial length measurements using applanation A-scan US and IOLMaster in eyes with CSME differ statistically and clinically significantly. This is likely a result of the fundamentally different methodology of the 2 methods in measuring eyes with a pathologically thickened retina. A conversion formula based on foveal thickness can not be provided at present.

Entities:  

Mesh:

Year:  2007        PMID: 17331581     DOI: 10.1016/j.ophtha.2006.10.053

Source DB:  PubMed          Journal:  Ophthalmology        ISSN: 0161-6420            Impact factor:   12.079


  8 in total

1.  Misleading axial length measurements with the IOLMaster due to a dense posterior vitreous surface membrane and a macular edema in a diabetic patient.

Authors:  Christoph Fidel Mayer; Navid Ardjomand; Werner Wackernagel; Michaela Velikay-Parel
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2011-11-03       Impact factor: 3.117

2.  Combined 23-gauge microincisonal vitrectomy surgery and phacoemulsification with AcrySof toric intraocular lens implantation: a comparative study.

Authors:  D H Park; J P Shin; S Y Kim
Journal:  Eye (Lond)       Date:  2011-07-15       Impact factor: 3.775

3.  Optical Coherence Tomography: Critical Tool to Manage Expectations after Cataract Extraction.

Authors:  Raquel Goldhardt; Bradley Simon Rosen
Journal:  Curr Ophthalmol Rep       Date:  2020-06-06

Review 4.  Cataract Surgery Considerations for Diabetic Patients.

Authors:  Jonathan A Go; Christina A Mamalis; Sumitra S Khandelwal
Journal:  Curr Diab Rep       Date:  2021-12-30       Impact factor: 4.810

5.  Comparison of the biometric values obtained by two different A-mode ultrasound devices (Eye Cubed vs. PalmScan): a transversal, descriptive, and comparative study.

Authors:  Raul Velez-Montoya; Eugene Mark Shusterman; Miriam Jessica López-Miranda; Mariana Mayorquin-Ruiz; Guillermo Salcedo-Villanueva; Hugo Quiroz-Mercado; Virgilio Morales-Cantón
Journal:  BMC Ophthalmol       Date:  2010-03-24       Impact factor: 2.209

6.  Association between retinal vein occlusion, axial length and vitreous chamber depth measured by optical low coherence reflectometry.

Authors:  Andrea Szigeti; Miklós Schneider; Mónika Ecsedy; Zoltán Zsolt Nagy; Zsuzsanna Récsán
Journal:  BMC Ophthalmol       Date:  2015-04-30       Impact factor: 2.209

7.  Precision of a new ocular biometer in children and comparison with IOLMaster.

Authors:  Xinxin Yu; Hao Chen; Giacomo Savini; Qianqian Zheng; Benhao Song; Ruixue Tu; Jinhai Huang; Qinmei Wang
Journal:  Sci Rep       Date:  2018-01-22       Impact factor: 4.379

8.  Axial length estimation error caused by hidden double-peak on partial coherence interferometry in an eye with epiretinal membrane: a case report.

Authors:  Yoshiyuki Kitaguchi; Shinsaku Yano; Fumi Gomi
Journal:  Clin Ophthalmol       Date:  2014-03-13
  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.