Shihao Chen1, Qiaoyue Zhang2, Giacomo Savini3, Shuangzhe Zhang1, Xiaomin Huang1, Jinjin Yu1, Yirang Wang1, Rui Ning1, Jinhai Huang1,4,5, Ruixue Tu1.
Abstract
PURPOSE: To evaluate measurement precision and to compare the Pentacam AXL (Oculus Optikgeräte, Wetzlar, German), a new optical biometer based on Scheimpflug imaging and partial coherence interferometry (PCI) with that of the OA-2000 biometer (Tomey, Nagoya, Japan), which combines swept-source optical coherence tomography (SS-OCT) and Placido-disk topography.Entities:
Keywords: Scheimpflug imaging; agreement; repeatability; reproducibility; swept source optical coherence tomography
Year: 2022 PMID: 35223885 PMCID: PMC8866319 DOI: 10.3389/fmed.2021.814519
Source DB: PubMed Journal: Front Med (Lausanne) ISSN: 2296-858X
Intraobserver repeatability outcomes for biometric measurements obtained using the new Scheimpflug imager in combination with partial coherence interferometry biometer.
|
|
|
|
|
|
|
|
|---|---|---|---|---|---|---|
| AL (mm) | 1st | 24.37 ± 1.49 | 0.04 | 0.11 | 0.16 | 0.999 (0.999–0.999) |
| 2nd | 24.38 ± 1.49 | 0.04 | 0.10 | 0.15 | 0.999 (0.999–1.000) | |
| CCT (μm) | 1st | 535.35 ± 29.68 | 3.73 | 10.33 | 0.70 | 0.984 (0.979–0.988) |
| 2nd | 535.04 ± 29.56 | 3.49 | 9.67 | 0.65 | 0.986 (0.982–0.990) | |
| AQD (mm) | 1st | 2.87 ± 0.45 | 0.02 | 0.05 | 0.68 | 0.998 (0.997–0.999) |
| 2nd | 2.86 ± 0.45 | 0.02 | 0.05 | 0.61 | 0.999 (0.998–0.999) | |
| ACD (mm) | 1st | 3.40 ± 0.45 | 0.02 | 0.06 | 0.62 | 0.998 (0.997–0.998) |
| 2nd | 3.40 ± 0.45 | 0.02 | 0.05 | 0.56 | 0.998 (0.998–0.999) | |
| Km (D) | 1st | 43.70 ± 1.55 | 0.10 | 0.28 | 0.23 | 0.996 (0.994–0.997) |
| 2nd | 43.70 ± 1.56 | 0.10 | 0.27 | 0.22 | 0.996 (0.995–0.997) | |
| J0 (D) | 1st | −0.26 ± 0.42 | 0.06 | 0.17 | −23.16 | 0.979 (0.973–0.985) |
| 2nd | −0.25 ± 0.42 | 0.06 | 0.17 | −23.97 | 0.979 (0.972–0.985) | |
| J45 (D) | 1st | 0.00 ± 0.21 | 0.07 | 0.19 | −1,417.77 | 0.896 (0.864–0.922) |
| 2nd | 0.00 ± 0.20 | 0.07 | 0.19 | 1,490.90 | 0.894 (0.862–0.921) | |
| CD (mm) | 1st | 11.53 ± 0.39 | 0.06 | 0.18 | 0.56 | 0.974 (0.965–0.980) |
| 2nd | 11.54 ± 0.38 | 0.06 | 0.17 | 0.55 | 0.974 (0.965–0.980) |
AL, axial length; CCT, central corneal thickness; AQD, aqueous depth; ACD, anterior chamber depth; Km, mean keratometry; CD, corneal diameter; SD, standard deviation; S.
Interobserver reproducibility outcomes for biometric measurements obtained using the new Scheimpflug imager in combination with partial coherence interferometry biometer.
|
|
|
|
|
|
|
|---|---|---|---|---|---|
| AL (mm) | 24.37 ± 1.49 | 0.02 | 0.06 | 0.09 | 1.000 (1.000–1.000) |
| CCT (μm) | 535.19 ± 29.56 | 2.14 | 5.92 | 0.40 | 0.995 (0.993–0.996) |
| AQD (mm) | 2.86 ± 0.45 | 0.01 | 0.04 | 0.51 | 0.999 (0.998–0.999) |
| ACD (mm) | 3.40 ± 0.45 | 0.02 | 0.04 | 0.46 | 0.999 (0.998–0.999) |
| Km (D) | 43.70 ± 1.55 | 0.06 | 0.17 | 0.14 | 0.999 (0.998–0.999) |
| J0 (D) | −0.26 ± 0.42 | 0.04 | 0.12 | - | 0.989 (0.985–0.992) |
| J45 (D) | 0.00 ± 0.21 | 0.04 | 0.12 | - | 0.957 (0.940–0.969) |
| CD (mm) | 11.54 ± 0.39 | 0.04 | 0.12 | 0.38 | 0.987 (0.982–0.991) |
AL, axial length; CCT, central corneal thickness; AQD, aqueous depth; ACD, anterior chamber depth; Km, mean keratometry; CD, corneal diameter; SD, standard deviation; S.
The mean difference, paired T-test, and 95% limits of agreement (LoA) for differences between the new Scheimpflug imager in combination with partial coherence interferometry biometer and the swept-source optical coherence tomography-based biometer.
|
|
|
|
|
|---|---|---|---|
| AL (mm) | −0.03 ± 0.03 | <0.001 | −0.09 to 0.02 |
| CCT (μm) | 10.80 ± 5.41 | <0.001 | 0.18 to 21.41 |
| AQD (mm) | 0.01 ± 0.04 | <0.001 | −0.07 to 0.10 |
| ACD (mm) | 0.03 ± 0.04 | <0.001 | −0.06 to 0.11 |
| Km (D) | −0.19 ± 0.14 | <0.001 | −0.48 to 0.09 |
| J0 (D) | 0.01 ± 0.10 | 0.242 | −0.18 to 0.20 |
| J45 (D) | −0.01 ± 0.08 | 0.302 | −0.17 to 0.15 |
| CD (mm) | −0.20 ± 0.15 | <0.001 | −0.50 to 0.09 |
AL, axial length; CCT, central corneal thickness; AQD, aqueous depth; ACD, anterior chamber depth; Km, mean keratometry; CD, corneal diameter; SD, standard deviation.
Figure 1Bland-Altman plots showing agreement between the new Scheimpflug imager in combination with partial coherence interferometry biometer and the swept-source optical coherence tomography-based biometer for measuring axial length. Solid lines represent the bias between the two devices and dotted lines represent the 95% confidence interval for the difference.
Figure 5Bland-Altman plots showing the agreement between the new Scheimpflug imager in combination with partial coherence interferometry biometer and the swept-source optical coherence tomography-based biometer for measuring mean keratometry at 3.0 mm. Solid lines represent the bias between the two devices and dotted lines represent the 95% confidence interval for the difference.