Literature DB >> 20560892

Measurement of central corneal thickness by high-resolution Scheimpflug imaging, Fourier-domain optical coherence tomography and ultrasound pachymetry.

Shihao Chen1, Jinhai Huang, Daizong Wen, Wuhe Chen, David Huang, Qinmei Wang.   

Abstract

PURPOSE: To compare the repeatability and reproducibility of central corneal thickness (CCT) measurements by high-resolution (HR) rotating Scheimpflug imaging and Fourier-domain optical coherence tomography (FD-OCT). CCT measurements were compared to those determined by ultrasound pachymetry (UP).
METHODS: In 35 healthy eyes, intra-observer repeatability for HR Scheimpflug (Pentacam) and FD-OCT (RTVue) systems was determined in consecutive images taken by an observer in the shortest time possible. Imaging was repeated again by a second observer to evaluate inter-observer reproducibility. The CCT measurements were compared among Scheimpflug, FD-OCT and UP images.
RESULTS: Mean coefficients of repeatability were 0.48% for Scheimpflug and 0.26% for FD-OCT. For Scheimpflug, the coefficient of inter-operator reproducibility was 0.87%. For FD-OCT, the coefficient of inter-operator reproducibility was 0.45%. The CCT measurements by Scheimpflug, OCT and UP images were (mean ± standard deviation) 521.7 ± 27.6 μm, 510.8 ± 28.6 μm and 516.5 ± 27.6 μm, respectively. The differences between instruments were statistically significant. The 95% limits of agreement in CCT were -0.7 to 22.5 μm for Pentacam-OCT, -13.4 to 24.0 μm for Pentacam-UP and -26.7 to 15.4 μm for OCT-UP. There was a high degree of correlation between CCT measured by all 3 methods.
CONCLUSION: Noncontact measurements of CCT with HR Scheimpflug and FD-OCT systems yielded excellent repeatability and reproducibility and can be used interchangeably. Although both devices were comparable with UP; in clinical practice, the measurements acquired by optical modalities are not directly interchangeable with UP measurements.
© 2010 The Authors. Journal compilation © 2010 Acta Ophthalmol.

Entities:  

Mesh:

Year:  2010        PMID: 20560892     DOI: 10.1111/j.1755-3768.2010.01947.x

Source DB:  PubMed          Journal:  Acta Ophthalmol        ISSN: 1755-375X            Impact factor:   3.761


  39 in total

1.  Meta-analysis of Pentacam vs. ultrasound pachymetry in central corneal thickness measurement in normal, post-LASIK or PRK, and keratoconic or keratoconus-suspect eyes.

Authors:  Wenjing Wu; Yan Wang; Lulu Xu
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2013-11-12       Impact factor: 3.117

2.  Central corneal thickness determination in corneal edema using ultrasound pachymetry, a Scheimpflug camera, and anterior segment OCT.

Authors:  David Kuerten; Niklas Plange; Eva Charlotte Koch; Antonis Koutsonas; Peter Walter; Matthias Fuest
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2015-04-21       Impact factor: 3.117

3.  Author reply: central corneal thickness determination in corneal edema.

Authors:  D Kuerten; N Plange; P Walter; M Fuest
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2017-03-22       Impact factor: 3.117

4.  Repeatability and comparability of corneal thickness measurements obtained from Dual Scheimpflug Analyzer and from ultrasonic pachymetry.

Authors:  Zaina N Al-Mohtaseb; Li Wang; Mitchell P Weikert
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2013-02-17       Impact factor: 3.117

5.  Anatomical characterization of central, apical and minimal corneal thickness.

Authors:  Federico Saenz-Frances; Martha Cecilia Bermúdez-Vallecilla; Lara Borrego-Sanz; Luis Jañez; José María Martinez-de-la-Casa; Laura Morales-Fernandez; Enrique Santos-Bueso; Julián Garcia-Sanchez; Julián Garcia-Feijoo
Journal:  Int J Ophthalmol       Date:  2014-08-18       Impact factor: 1.779

6.  Effect of 2% fluorescein on Scheimpflug central corneal thickness measurements.

Authors:  Stella Briggs; Marwa Bin Moammar
Journal:  Int J Ophthalmol       Date:  2016-02-18       Impact factor: 1.779

7.  Factors associated with changes in posterior corneal surface following photorefractive keratectomy.

Authors:  Achia Nemet; Michael Mimouni; Igor Vainer; Tzahi Sela; Igor Kaiserman
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2021-06-07       Impact factor: 3.117

8.  Multimodal Assessment of Corneal Thinning Using Optical Coherence Tomography, Scheimpflug Imaging, Pachymetry, and Slit-Lamp Examination.

Authors:  Julius T Oatts; Jeremy D Keenan; Tova Mannis; Tom M Lietman; Jennifer Rose-Nussbaumer
Journal:  Cornea       Date:  2017-04       Impact factor: 2.651

9.  Comparison of central corneal thickness with four different optical devices.

Authors:  Kuddusi Teberik; Mehmet Tahir Eski; Murat Kaya; Handan Ankaralı
Journal:  Int Ophthalmol       Date:  2017-10-11       Impact factor: 2.031

10.  [Pachymetry and intraocular pressure measurement by corneal visualization Scheimpflug technology (Corvis ST): A clinical comparison to the gold standard].

Authors:  J Steinberg; J Mehlan; A Frings; V Druchkiv; G Richard; T Katz; S J Linke
Journal:  Ophthalmologe       Date:  2015-09       Impact factor: 1.059

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.