| Literature DB >> 23049823 |
Qinmei Wang1, Giacomo Savini, Kenneth J Hoffer, Zhen Xu, Yifan Feng, Daizong Wen, Yanjun Hua, Feng Yang, Chao Pan, Jinhai Huang.
Abstract
PURPOSE: To comprehensively assess the precision and agreement of anterior corneal power measurements using 8 different devices.Entities:
Mesh:
Year: 2012 PMID: 23049823 PMCID: PMC3458095 DOI: 10.1371/journal.pone.0045607
Source DB: PubMed Journal: PLoS One ISSN: 1932-6203 Impact factor: 3.240
Intrasession Repeatability of 8 Different Devices in Measuring Flat Keratometry (N = 35).
| Device | Session | Mean (D) ± SD | Sw (D) | 2.77 Sw (D) | COV (%) | ICC |
| Tomey RC | 1st | 42.85±1.26 | 0.04 | 0.10 | 0.08 | 0.999 |
| 2nd | 42.85±1.27 | 0.03 | 0.09 | 0.08 | 0.999 | |
| Topcon KR | 1st | 42.90±1.30 | 0.05 | 0.15 | 0.13 | 0.998 |
| 2nd | 42.91±1.27 | 0.06 | 0.17 | 0.14 | 0.998 | |
| IOLMaster | 1st | 42.97±1.27 | 0.03 | 0.09 | 0.07 | 0.999 |
| 2nd | 42.99±1.27 | 0.04 | 0.10 | 0.08 | 0.999 | |
| EyeSys Vista | 1st | 42.65±1.21 | 0.13 | 0.36 | 0.30 | 0.989 |
| 2nd | 42.64±1.19 | 0.10 | 0.29 | 0.24 | 0.992 | |
| Medmont | 1st | 43.05±1.29 | 0.07 | 0.20 | 0.17 | 0.997 |
| 2nd | 43.06±1.27 | 0.07 | 0.20 | 0.17 | 0.997 | |
| Topolyzer | 1st | 42.90±1.28 | 0.10 | 0.27 | 0.23 | 0.994 |
| 2nd | 42.91±1.25 | 0.10 | 0.27 | 0.23 | 0.994 | |
| Pentacam | 1st | 42.90±1.26 | 0.08 | 0.21 | 0.18 | 0.996 |
| 2nd | 42.90±1.26 | 0.07 | 0.20 | 0.17 | 0.997 | |
| Sirius | 1st | 42.97±1.26 | 0.10 | 0.27 | 0.23 | 0.994 |
| 2nd | 43.01±1.28 | 0.08 | 0.22 | 0.18 | 0.996 |
D = diopter, SD = standard deviation, Sw = within-subject standard deviation, COV = within-subject coefficient of variation, ICC = intraclass correlation coefficient.
Intrasession Repeatability of 8 Different Devices in Measuring Steep Keratometry (N = 35).
| Device | Session | Mean (D) ± SD | Sw (D) | 2.77 Sw (D) | COV (%) | ICC |
| Tomey RC | 1st | 43.87±1.44 | 0.05 | 0.14 | 0.12 | 0.999 |
| 2nd | 43.86±1.46 | 0.04 | 0.12 | 0.10 | 0.999 | |
| Topcon KR | 1st | 43.74±1.43 | 0.08 | 0.21 | 0.17 | 0.997 |
| 2nd | 43.76±1.44 | 0.08 | 0.23 | 0.19 | 0.997 | |
| IOLMaster | 1st | 43.97±1.44 | 0.04 | 0.11 | 0.09 | 0.999 |
| 2nd | 43.96±1.43 | 0.05 | 0.15 | 0.12 | 0.999 | |
| EyeSysVista | 1st | 43.53±1.38 | 0.12 | 0.34 | 0.28 | 0.992 |
| 2nd | 43.54±1.39 | 0.13 | 0.35 | 0.29 | 0.992 | |
| Medmont | 1st | 44.08±1.44 | 0.07 | 0.20 | 0.16 | 0.998 |
| 2nd | 44.15±1.46 | 0.08 | 0.23 | 0.18 | 0.997 | |
| Topolyzer | 1st | 43.84±1.46 | 0.13 | 0.35 | 0.29 | 0.993 |
| 2nd | 43.84±1.45 | 0.12 | 0.32 | 0.27 | 0.994 | |
| Pentacam | 1st | 43.85±1.43 | 0.09 | 0.24 | 0.19 | 0.996 |
| 2nd | 43.86±1.45 | 0.08 | 0.23 | 0.19 | 0.997 | |
| Sirius | 1st | 43.90±1.46 | 0.10 | 0.29 | 0.24 | 0.995 |
| 2nd | 43.96±1.46 | 0.10 | 0.27 | 0.22 | 0.996 |
D = diopter, SD = standard deviation, Sw = within-subject standard deviation, COV = within-subject coefficient of variation, ICC = intraclass correlation coefficient.
Intrasession Repeatability of 8 Different Devices in Measuring Mean Keratometry (N = 35).
| Device | Session | Mean (D) ± SD | Sw (D) | 2.77 Sw (D) | COV (%) | ICC |
| Tomey RC | 1st | 43.36±1.31 | 0.03 | 0.08 | 0.06 | 1.000 |
| 2nd | 43.36±1.33 | 0.03 | 0.07 | 0.06 | 1.000 | |
| Topcon KR | 1st | 43.32±1.34 | 0.06 | 0.16 | 0.13 | 0.998 |
| 2nd | 43.33±1.33 | 0.06 | 0.17 | 0.14 | 0.998 | |
| IOLMaster | 1st | 43.47±1.32 | 0.02 | 0.06 | 0.05 | 1.000 |
| 2nd | 43.48±1.32 | 0.03 | 0.07 | 0.06 | 1.000 | |
| EyeSysVista | 1st | 43.09±1.26 | 0.11 | 0.32 | 0.26 | 0.992 |
| 2nd | 43.09±1.26 | 0.10 | 0.29 | 0.24 | 0.993 | |
| Medmont | 1st | 43.57±1.33 | 0.07 | 0.18 | 0.15 | 0.999 |
| 2nd | 43.60±1.33 | 0.06 | 0.17 | 0.14 | 0.998 | |
| Topolyzer | 1st | 43.37±1.34 | 0.11 | 0.29 | 0.24 | 0.994 |
| 2nd | 43.37±1.32 | 0.10 | 0.28 | 0.23 | 0.994 | |
| Pentacam | 1st | 43.38±1.31 | 0.07 | 0.18 | 0.15 | 0.998 |
| 2nd | 43.38±1.32 | 0.06 | 0.17 | 0.14 | 0.998 | |
| Sirius | 1st | 43.44±1.33 | 0.09 | 0.25 | 0.21 | 0.996 |
| 2nd | 43.48±1.34 | 0.08 | 0.21 | 0.18 | 0.997 |
D = diopter, SD = standard deviation, Sw = within-subject standard deviation, COV = within-subject coefficient of variation, ICC = intraclass correlation coefficient.
Comparison of the Flat Keratometry between 8 Different Devices.
| Devices | Mean Difference(D) ± SD |
| 95% LoA (D) |
| Tomey-Topcon | −0.05±0.09 | <.01 | −0.226 to 0.134 |
| Tomey-IOLMaster | −0.12±0.08 | <.01 | −0.278 to 0.044 |
| Tomey-EyeSys | 0.20±0.11 | <.01 | −0.010 to 0.41 |
| Tomey-Medmont | −0.20±0.15 | <.01 | −0.480 to 0.090 |
| Tomey-Topolyzer | −0.05±0.09 | <.01 | −0.231 to 0.133 |
| Tomey-Pentacam | −0.05±0.09 | <.01 | −0.236 to 0.132 |
| Tomey-Sirius | −0.12±0.11 | <.01 | −0.330 to 0.09 |
| Topcon-IOLMaster | −0.07±0.09 | <.01 | −0.246 to 0.103 |
| Topcon-EyeSys | 0.24±0.14 | <.01 | −0.02 to 0.51 |
| Topcon-Medmont | −0.15±0.14 | <.01 | −0.42 to 0.12 |
| Topcon-Topolyzer | 0.00±0.10 | .851 | −0.196 to 0.19 |
| Topcon-Pentacam | −0.01±0.11 | .735 | −0.22 to0.21 |
| Topcon-Sirius | −0.08±0.12 | <.01 | −0.31 to0.15 |
| IOLMaster-EyeSys | 0.32±0.12 | <.01 | 0.07 to 0.56 |
| IOLMaster-Medmont | −0.08±0.15 | <.01 | −0.37 to 0.21 |
| IOLMaster-Topolyzer | 0.07±0.08 | <.01 | −0.082 to 0.219 |
| IOLMaster-Pentacam | 0.07±0.11 | <.01 | −0.16 to 0.29 |
| IOLMaster-Sirius | 0.00±0.11 | .859 | −0.22 to 0.22 |
| EyeSys-Medmont | −0.40±0.17 | <.01 | −0.73 to −0.06 |
| EyeSys-Topolyzer | −0.25±0.13 | <.01 | −0.5 to 0 |
| EyeSys-Pentacam | −0.25±0.11 | <.01 | −0.47 to 0.03 |
| EyeSys-Sirius | −0.32±0.12 | <.01 | −0.56 to −0.08 |
| Medmont-Topolyzer | 0.15±0.15 | <.01 | −0.14 to 0.44 |
| Medmont-Pentacam | 0.14±0.14 | <.01 | −0.14 to 0.43 |
| Medmont-Sirius | 0.08±0.17 | .014 | −0.26 to 0.42 |
| Topolyzer-Pentacam | 0.00±0.11 | .872 | −0.23 to 0.22 |
| Topolyzer-Sirius | −0.07±0.12 | <.01 | −0.31 to0.17 |
| Pentacam-Sirius | −0.07±0.10 | <.01 | −0.263 to 0.126 |
D = diopter, SD = standard deviation, LoA = limits of agreement.
Comparison of the Steep Keratometry between 8 Different Devices.
| Devices | Mean Difference(D) ± SD |
| 95% LoA (D) |
| Tomey-Topcon | 0.13±0.12 | <.01 | −0.100 to 0.360 |
| Tomey-IOLMaster | −0.10±0.13 | <.01 | −0.360 to 0.160 |
| Tomey-EyeSys | 0.34±0.14 | <.01 | 0.060 to 0.620 |
| Tomey-Medmont | −0.21±0.18 | <.01 | −0.560 to 0.140 |
| Tomey-Topolyzer | 0.03±0.18 | .346 | −0.330 to 0.390 |
| Tomey-Pentacam | 0.02±0.14 | .357 | −0.260 to 0.310 |
| Tomey-Sirius | −0.03±0.14 | .220 | −0.310 to 0.25 |
| Topcon-IOLMaster | −0.23±0.13 | <.01 | −0.48 to 0.02 |
| Topcon-EyeSys | 0.21±0.17 | <.01 | −0.12 to 0.53 |
| Topcon-Medmont | −0.34±0.18 | <.01 | −0.69 to 0 |
| Topcon-Topolyzer | −0.07±0.06 | <.01 | −0.41 to 0.21 |
| Topcon-Pentacam | −0.11±0.16 | <.01 | −0.42 to 0.2 |
| Topcon-Sirius | −0.16±0.14 | <.01 | −0.43 to 0.11 |
| IOLMaster-EyeSys | 0.44±0.17 | <.01 | 0.11 to 0.77 |
| IOLMaster-Medmont | −0.11±0.22 | <.01 | −0.54 to 0.32 |
| IOLMaster-Topolyzer | 0.13±0.21 | <.01 | −0.28 to 0.54 |
| IOLMaster-Pentacam | 0.12±0.13 | <.01 | −0.13 to 0.37 |
| IOLMaster-Sirius | 0.07±0.17 | .021 | −0.27 to 0.41 |
| EyeSys-Medmont | −0.55±0.17 | <.01 | −0.88 to −0.22 |
| EyeSys-Topolyzer | −0.31±0.19 | <.01 | −0.68 to 0.06 |
| EyeSys-Pentacam | −0.32±0.14 | <.01 | −0.60 to −0.03 |
| EyeSys-Sirius | −0.37±0.16 | <.01 | −0.68 to −0.06 |
| Medmont-Topolyzer | 0.24±0.17 | <.01 | −0.09 to 0.57 |
| Medmont-Pentacam | 0.23±0.18 | <.01 | −0.12 to 0.59 |
| Medmont-Sirius | 0.18±0.18 | <.01 | −0.18 to 0.54 |
| Topolyzer-Pentacam | −0.01±0.19 | .838 | −0.38 to 0.36 |
| Topolyzer-Sirius | −0.06±0.15 | .025 | −0.36 to 0.24 |
| Pentacam-Sirius | −0.05±0.16 | .053 | −0.36 to 0.25 |
D = diopter, SD = standard deviation, LoA = limits of agreement.
Comparison of Mean Keratometry between 8 Different Devices.
| Devices | Mean Difference(D) ± SD |
| 95% LoA (D) |
| Tomey-Topcon | 0.04±0.08 | <.01 | −0.11 to 0.20 |
| Tomey-IOLMaster | −0.11±0.08 | <.01 | −0.268 to 0.05 |
| Tomey-EyeSys | 0.27±0.11 | <.01 | 0.06 to 0.48 |
| Tomey-Medmont | −0.21±0.12 | <.01 | −0.45 to 0.04 |
| Tomey-Topolyzer | −0.01±0.11 | .694 | −0.22 to 0.21 |
| Tomey-Pentacam | −0.01±0.10 | .439 | −0.22 to 0.19 |
| Tomey-Sirius | −0.08±0.10 | <.01 | −0.26 to 0.11 |
| Topcon-IOLMaster | −0.15±0.08 | <.01 | −0.317 to 0.11 |
| Topcon-EyeSys | 0.23±0.14 | <.01 | −0.04 to 0.5 |
| Topcon-Medmont | −0.25±0.13 | <.01 | −0.51to 0.01 |
| Topcon-Topolyzer | −0.10±0.16 | <.01 | −0.25 to 0.15 |
| Topcon-Pentacam | −0.06±0.12 | <.01 | −0.29 to 0.18 |
| Topcon-Sirius | −0.12±0.11 | <.01 | −0.33 to 0.09 |
| IOLMaster-EyeSys | 0.38±0.12 | <.01 | 0.14 to 0.62 |
| IOLMaster-Medmont | −0.10±0.15 | <.01 | −0.39 to 0.20 |
| IOLMaster-Topolyzer | 0.10±0.11 | <.01 | −0.12 to 0.32 |
| IOLMaster-Pentacam | 0.10±0.10 | <.01 | −0.091 to 0.283 |
| IOLMaster-Sirius | 0.03±0.11 | .074 | −0.18 to 0.25 |
| EyeSys-Medmont | −0.48±0.15 | <.01 | −0.77 to −0.18 |
| EyeSys-Topolyzer | −0.28±0.14 | <.01 | −0.5 5to 0 |
| EyeSys-Pentacam | −0.28±0.11 | <.01 | −0.49 to −0.07 |
| EyeSys-Sirius | −0.35±0.13 | <.01 | −0.6 to −0.09 |
| Medmont-Topolyzer | 0.19±0.14 | <.01 | −0.09 to 0.47 |
| Medmont-Pentacam | 0.20±0.14 | <.01 | −0.08 to 0.47 |
| Medmont-Sirius | 0.13±0.16 | <.01 | −0.18 to 0.45 |
| Topolyzer-Pentacam | −0.01±0.14 | .793 | −0.28 to 0.27 |
| Topolyzer-Sirius | −0.07±0.12 | <.01 | −0.31 to 0.18 |
| Pentacam-Sirius | −0.06±0.11 | <.01 | −0.28 to 0.16 |
D = diopter, SD = standard deviation, LoA = limits of agreement.