Literature DB >> 24656166

Comparison of ocular biometry and intraocular lens power using a new biometer and a standard biometer.

Sabong Srivannaboon1, Chareenun Chirapapaisan2, Pratuangsri Chonpimai1, Sunisa Koodkaew1.   

Abstract

PURPOSE: To compare the repeatability and reproducibility of ocular biometry and intraocular lens (IOL) power obtained with a new optical biometer (AL-Scan) and a standard optical biometer (IOLMaster 500).
SETTING: Siriraj Hospital, Mahidol University, Bangkok, Thailand.
DESIGN: Prospective comparative study.
METHODS: Two independent operators measured eyes with cataract using both biometers. The keratometry values, axial length, anterior chamber depth, white-to-white (WTW) corneal diameter, and IOL power calculated using the Holladay 1 formula obtained with each device were recorded. Intraoperator repeatability and interoperator reproducibility of both devices were analyzed using the intraclass correlation coefficient (ICC). The agreement in ocular biometry and IOL power between the 2 devices was evaluated by the Bland-Altman method.
RESULTS: The study recruited 137 eyes of 81 patients. The repeatability and reproducibility of both devices were high for all ocular biometry measurements (ICC, 0.87-1.00). Except for the WTW corneal diameter (ICC, 0.44), the agreement between the biometers was also high (ICC, 0.98-0.99). The IOL powers calculated by the Holladay 1 formula were similar between the 2 biometers.
CONCLUSION: The new optical biometer provided excellent repeatability and reproducibility for all ocular biometry. Agreement with the standard optical biometer was good except for the WTW corneal diameter.
Copyright © 2014 ASCRS and ESCRS. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2014        PMID: 24656166     DOI: 10.1016/j.jcrs.2013.09.020

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  17 in total

1.  Comparison of the measurements of a novel optical biometry: Nidek AL-Scan with Sirius and a ultrasound biometry.

Authors:  Çağatay Çağlar; Sücattin İlker Kocamış; Emre Demir; Mustafa Durmuş
Journal:  Int Ophthalmol       Date:  2016-07-08       Impact factor: 2.031

2.  Repeatability of ocular biometry with IOLMaster 700 in subjects with clear lens.

Authors:  Manuel Garza-Leon; Hugo A Fuentes-de la Fuente; Ana V García-Treviño
Journal:  Int Ophthalmol       Date:  2016-10-21       Impact factor: 2.031

3.  Comparison of ocular parameters of two biometric measurement devices in highly myopic eyes.

Authors:  Xiao-Xiao Guo; Ran You; Shan-Shan Li; Xiu-Fen Yang; Lu Zhao; Fan Zhang; Yan-Ling Wang; Xi Chen
Journal:  Int J Ophthalmol       Date:  2019-10-18       Impact factor: 1.779

4.  Evaluation of the Repeatability and the Reproducibility of AL-Scan Measurements Obtained by Residents.

Authors:  Mehmet Kola; Hikmet Duran; Adem Turk; Suleyman Mollamehmetoglu; Ahmet Kalkisim; Hidayet Erdol
Journal:  J Ophthalmol       Date:  2014-07-22       Impact factor: 1.909

5.  Comparison between a New Optical Biometry Device and an Anterior Segment Optical Coherence Tomographer for Measuring Central Corneal Thickness and Anterior Chamber Depth.

Authors:  Jinhai Huang; Weicong Lu; Giacomo Savini; Hao Chen; Chengfang Wang; Xinxin Yu; Fangjun Bao; Qinmei Wang
Journal:  J Ophthalmol       Date:  2016-06-14       Impact factor: 1.909

6.  Comparison of a new optical biometer and a standard biometer in cataract patients.

Authors:  Pipat Kongsap
Journal:  Eye Vis (Lond)       Date:  2016-10-17

7.  Precision of a new ocular biometer in children and comparison with IOLMaster.

Authors:  Xinxin Yu; Hao Chen; Giacomo Savini; Qianqian Zheng; Benhao Song; Ruixue Tu; Jinhai Huang; Qinmei Wang
Journal:  Sci Rep       Date:  2018-01-22       Impact factor: 4.379

8.  Precision of a new ocular biometer in eyes with cataract using swept source optical coherence tomography combined with Placido-disk corneal topography.

Authors:  Wenwen Wang; Yaxin Miao; Giacomo Savini; Colm McAlinden; Hao Chen; Qingjie Hu; Qinmei Wang; Jinhai Huang
Journal:  Sci Rep       Date:  2017-10-23       Impact factor: 4.379

9.  Repeatability and Reproducibility of Peripapillary Choroidal Thickness Using a Medical Image-Processing Software.

Authors:  Pedro Simoes; Pedro Silva; Miguel Cordeiro; Joao Costa
Journal:  Med Hypothesis Discov Innov Ophthalmol       Date:  2018

10.  Refractive Results Using a New Optical Biometry Device: Comparison With Ultrasound Biometry Data.

Authors:  Serdar Aktas; Hatice Aktas; Mehmet Tetikoglu; Hac Murat Sagdk; Fatih Özcura
Journal:  Medicine (Baltimore)       Date:  2015-12       Impact factor: 1.817

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.