Comparison of Pentacam Scheimpflug camera with ultrasound pachymetry and noncontact specular microscopy in measuring central corneal thickness.

Miyuki Fujioka1, Makoto Nakamura, Yasuko Tatsumi, Azusa Kusuhara, Hidetaka Maeda, Akira Negi.   

Abstract

PURPOSE: Although central corneal thickness (CCT) can be measured by several methods, interchangeability of different modalities has not been fully investigated. CCT is known to correlate with intraocular pressure (IOP). The aim of this study was to evaluate the agreement of Pentacam Scheimpflug system with noncontact specular microscopy (NCSM) and ultrasound (US) pachymetry in measuring CCT and the relation between IOP taken with Goldmann applanation tonometer (GAT) and the CCT measured with these three methods.
METHODS: The right eyes of 135 enrolled persons without antiglaucoma drug use (100 females and 35 males), who comprised 32 patients with primary open-angle glaucoma, 14 with ocular hypertension, 45 with primary angle-closure glaucoma, and 44 controls, were studied. Intermethod comparison of CCT was made by the 95% limits of agreement analysis according to Bland and Altman. Linear regression analysis was used to assess the relationship between IOP and CCT taken with each modality.
RESULTS: The mean CCT (+/-SD) taken with Scheimpflug, US, and NCSM was 559.49 +/- 38.44 microm, 553.01 +/- 39.33 microm, and 552.04 +/- 42.95 microm, respectively. The average values of CCT taken with the three instruments were not significantly different (one-factor ANOVA; p = 0.26), although the marginal mean difference between Scheimpflug and US or NCSM was statistically significant (paired t test; p = 0.0009 and 0.005, respectively). The 95% limits of agreement were 6.47 +/- 43.21 microm between Scheimpflug and US, 7.45 +/- 58.86 microm between Scheimpflug and NCSM, and 0.98 +/- 51.69 microm between US and NCSM. There was a positive association between IOP and CCT measured with US or NCSM, whereas there was no correlation between IOP and CCT measured with Scheimpflug.
CONCLUSIONS: Although CCT values measured with Scheimpflug, US, and NCSM are closely similar, clinicians should keep in mind that these methods are not simply interchangeable.

Entities:  

Mesh:

Year:  2007        PMID: 17364741     DOI: 10.1080/02713680601115010

Source DB:  PubMed          Journal:  Curr Eye Res        ISSN: 0271-3683            Impact factor:   2.424


  18 in total

1.  Repeatability and agreement of CCT measurement in myopia using entacam and ultrasound pachymetry.

Authors:  Yong-Heng Luo; Qian Zhong; Ping-Bo Ouyang; Xiao-Jian Guo; Xuan-Chu Duan
Journal:  Int J Ophthalmol       Date:  2012-06-18       Impact factor: 1.779

2.  Comparison between central corneal thickness measurements by oculus pentacam and ultrasonic pachymetry.

Authors:  Hani S Al-Mezaine; Saleh A Al-Amro; Dustan Kangave; Abdulkareem Sadaawy; Taher A Wehaib; Saleh Al-Obeidan
Journal:  Int Ophthalmol       Date:  2007-09-26       Impact factor: 2.031

3.  Measurements of anterior segment parameters using three different non-contact optical devices in keratoconus patients.

Authors:  Ahmet Taylan Yazıcı; Gökhan Pekel; Ercüment Bozkurt; Yusuf Yıldırım; Evre Pekel; Ahmet Demirok; Omer Faruk Yılmaz
Journal:  Int J Ophthalmol       Date:  2013-08-18       Impact factor: 1.779

4.  Comparison of central corneal thickness measurements by Orbscan II and Pentacam after corneal refractive surgery.

Authors:  Jumpei Matsuda; Osamu Hieda; Shigeru Kinoshita
Journal:  Jpn J Ophthalmol       Date:  2008-09-05       Impact factor: 2.447

5.  Anatomical characterization of central, apical and minimal corneal thickness.

Authors:  Federico Saenz-Frances; Martha Cecilia Bermúdez-Vallecilla; Lara Borrego-Sanz; Luis Jañez; José María Martinez-de-la-Casa; Laura Morales-Fernandez; Enrique Santos-Bueso; Julián Garcia-Sanchez; Julián Garcia-Feijoo
Journal:  Int J Ophthalmol       Date:  2014-08-18       Impact factor: 1.779

6.  A multimodal ophthalmic analysis in patients with systemic sclerosis using ocular response analyzer, corneal topography and specular microscopy.

Authors:  Huseyin Mayali; Muhammed Altinisik; Secil Sencan; Timur Pirildar; Emin Kurt
Journal:  Int Ophthalmol       Date:  2019-09-28       Impact factor: 2.031

7.  Effect of 2% fluorescein on Scheimpflug central corneal thickness measurements.

Authors:  Stella Briggs; Marwa Bin Moammar
Journal:  Int J Ophthalmol       Date:  2016-02-18       Impact factor: 1.779

8.  Comparison of central corneal thickness measurements by Pentacam, noncontact specular microscope, and ultrasound pachymetry in normal and post-LASIK eyes.

Authors:  Saleh Al-Ageel; Abdulrahman M Al-Muammar
Journal:  Saudi J Ophthalmol       Date:  2009-10-24

9.  Comparison of central corneal thickness with four different optical devices.

Authors:  Kuddusi Teberik; Mehmet Tahir Eski; Murat Kaya; Handan Ankaralı
Journal:  Int Ophthalmol       Date:  2017-10-11       Impact factor: 2.031

10.  Comparison of central corneal thickness and anterior chamber depth measurements using three imaging technologies in normal eyes and after phakic intraocular lens implantation.

Authors:  Muriël Doors; Lars P J Cruysberg; Tos T J M Berendschot; John de Brabander; Frenne Verbakel; Carroll A B Webers; Rudy M M A Nuijts
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2009-04-30       Impact factor: 3.117

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.