Sang Woo Moon1, Sung Hyup Lim1, Ho Young Lee2.
Abstract
PURPOSE: To compare the refractive results of cataract surgery measured by applanation ultrasound and the new partial coherence interferometer, AL-scan.Entities:
Keywords: Biometry; Interferometery; Phacoemulsification; Ultrasonography
Mesh:
Year: 2014 PMID: 25435746 PMCID: PMC4239462 DOI: 10.3341/kjo.2014.28.6.444
Source DB: PubMed Journal: Korean J Ophthalmol ISSN: 1011-8942
Fig. 1Scatter plots to compare means for axial length measured with ultrasound and AL-scan. Axial length measurement by two methods showed a statistically significant relationship. USAXL = axial length measured with applanation ultrasound; ALAXL = axial length measured with AL-scan. r = 0.976, p < 0.01.
Fig. 2Bland-Altman plots for assessing agreement of pairs of two methods. y = 0.012 X - 0.317; mean = -0.0334 ; mean +2 SD = 0.32508 ; mean -2 SD = -0.39188. The solid line represents the average mean difference (-0.0334) and the dotted line represents the 95 percentile confidence interval. The slope of regression line is 0.012 and represents low error between two methods and high reliability. In the scatter plot, most values are on the dotted line (±1.96 SD, -0.3919 to 0.3251), suggesting good comparability. SD = standard deviation; USAXL = axial length measured with applanation ultrasound; ALAXL = axial length measured with AL-scan.
Refractive results: comparison of the ultrasound group and the AL-scan group (all eyes, n = 98)
Values are presented as mean ± standard deviation.
D = diopter; K = corneal refractive power (keratometric diopter).
Percentage of cases predicted to within ±0.50 D, ±1.00 D, and ±1.50 D of each group
Values are presented as %.
D = diopter; US = ultrasound; USAXL = axial length measured with applanation ultrasound; ALAXL = axial length measured with AL-scan.
Refractive results: comparison of the ultrasound group and the AL-scan group (axial length ≤24.4 mm, n = 86)
Values are presented as mean ± standard deviation.
D = diopter; K = corneal refractive power (keratometric diopter).
Refractive results: comparison of the ultrasound group and the AL-scan group (axial length >24.4 mm, n = 12)
Values are presented as mean ± standard deviation.
D = diopter; K = corneal refractive power (keratometric diopter).
Refractive results: comparison of the ultrasound group and the AL-scan group (ultrasound axial length > AL-scan axial length, n = 26)
Values are presented as mean ± standard deviation.
D = diopter; K = corneal refractive power (keratometric diopter).