Literature DB >> 17889762

Effect of corneal thickness on the agreement between ultrasound and Orbscan II pachymetry.

Hassan Hashemi1, Maryam Roshani, Shiva Mehravaran, Hiva Parsafar, Kamran Yazdani.   

Abstract

PURPOSE: To study the agreement between ultrasound (US) and Orbscan II (Bausch & Lomb) measurements with different ranges of central thickness in normal corneas.
SETTING: Noor Ophthalmology Research Center, Tehran, Iran.
METHODS: The central corneal thickness was measured first by the Orbscan II and then with a US pachymeter in 177 right eyes. Data were compared in 3 thickness groups: less than 500 microm, 500 to 600 microm, and more than 600 microm. In all cases, uncorrected Orbscan II data were used. For each range, a correction factor was calculated through regression analysis, which was then used in a new set of analyses. To study the agreement between the 2 methods, the 95% limits of agreement (LoA) and intraclass correlations coefficients (ICC) were determined.
RESULTS: The mean interdevice difference was 2.7 +/- 16.9 (SD) in the less than 500 microm group (P = .365), 21.3 +/- 3.1 in the 500 to 600 microm group (P = .0001), and 27.2 +/- 20.9 in the more than 600 microm group (P = .0001). No clinically acceptable LoA were found in any group. The application of regression equations to the Orbscan II readings improved the agreement in the less than 500 microm group better than the other 2 groups. The ICC was 0.76, 0.61, and 0.43 in the less than 500 microm, 500 to 600 microm, and more than 600 microm groups, respectively.
CONCLUSIONS: The Orbscan corneal thickness readings had relatively lower validity than US measurements. The 2 devices should not be used interchangeably, especially when the validity of the measurement is vital. In patient screening for corneal surgery, it is advisable to recheck corneal thickness with a US pachymeter.

Entities:  

Mesh:

Year:  2007        PMID: 17889762     DOI: 10.1016/j.jcrs.2007.05.036

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  12 in total

1.  White-to-white corneal diameter, pupil diameter, central corneal thickness and thinnest corneal thickness values of emmetropic subjects.

Authors:  Juan A Sanchis-Gimeno; Daniel Sanchez-Zuriaga; Francisco Martinez-Soriano
Journal:  Surg Radiol Anat       Date:  2011-10-22       Impact factor: 1.246

2.  Intra-examiner repeatability and agreement of corneal pachymetry map measurement by time-domain and Fourier-domain optical coherence tomography.

Authors:  Jehn-Yu Huang; Melike Pekmezci; Stephanie Yaplee; Shan Lin
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2010-03-30       Impact factor: 3.117

3.  Lamina cribrosa thickness is not correlated with central corneal thickness or axial length in healthy eyes: central corneal thickness, axial length, and lamina cribrosa thickness.

Authors:  Eun Ji Lee; Tae-Woo Kim; Robert N Weinreb; Min Hee Suh; Hyunjoong Kim
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2012-09-20       Impact factor: 3.117

4.  Day to Day Clinically Relevant Corneal Elevation, Thickness, and Curvature Parameters Using the Orbscan II Scanning Slit Topographer and the Pentacam Scheimpflug Imaging Device.

Authors:  Hassan Hashemi; Shiva Mehravaran
Journal:  Middle East Afr J Ophthalmol       Date:  2010-01

5.  Measurement of Central Corneal Thickness Using Ultrasound Pachymetry and Orbscan II in Normal Eyes.

Authors:  Mohammad Mehdi Sadoughi; Bahram Einollahi; Neda Einollahi; Javad Rezaei; Danial Roshandel; Sepehr Feizi
Journal:  J Ophthalmic Vis Res       Date:  2015 Jan-Mar

6.  Central Corneal Thickness in Highly Myopic Eyes: Inter-device Agreement of Ultrasonic Pachymetry, Pentacam and Orbscan II Before and After Photorefractive Keratectomy.

Authors:  Nader Nassiri; Kourosh Sheibani; Sare Safi; Saman Nassiri; Alireza Ziaei; Alireza Ziaee; Farnaz Haji; Shiva Mehravaran; Nariman Nassiri
Journal:  J Ophthalmic Vis Res       Date:  2014-01

7.  Precision of a new ocular biometer in children and comparison with IOLMaster.

Authors:  Xinxin Yu; Hao Chen; Giacomo Savini; Qianqian Zheng; Benhao Song; Ruixue Tu; Jinhai Huang; Qinmei Wang
Journal:  Sci Rep       Date:  2018-01-22       Impact factor: 4.379

8.  Comparison of central corneal thickness with four noncontact devices: An agreement analysis of swept-source technology.

Authors:  Erhan Ozyol; Pelin Özyol
Journal:  Indian J Ophthalmol       Date:  2017-06       Impact factor: 1.848

9.  Comparison of central corneal thickness and anterior chamber depth measurements using three imaging technologies in normal eyes and after phakic intraocular lens implantation.

Authors:  Muriël Doors; Lars P J Cruysberg; Tos T J M Berendschot; John de Brabander; Frenne Verbakel; Carroll A B Webers; Rudy M M A Nuijts
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2009-04-30       Impact factor: 3.117

10.  Central Corneal Thickness Measurement Using Ultrasonic Pachymetry, Rotating Scheimpflug Camera, and Scanning-slit Topography Exclusively in Thin Non-keratoconic Corneas.

Authors:  Mehrdad Mohammadpour; Kazem Mohammad; Nasser Karimi
Journal:  J Ophthalmic Vis Res       Date:  2016 Jul-Sep
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.