Reproducibility of ocular biometry with a new noncontact optical low-coherence reflectometer in children.

Afsun Şahin1, Hüseyin Gürsoy, Hikmet Başmak, Nilgün Yildirim, Zuhat Usalp, Ertuğrul Çolak.   

Abstract

PURPOSE: To evaluate the reproducibility of Lenstar in the pediatric population.
METHODS: Three consecutive measurements per eye by 2 ophthalmologists with a total of 6 measurements per eye were performed using Lenstar. Axial length (AL), central corneal thickness (CCT), anterior chamber depth (ACD), lens thickness (LT), and corneal curvatures (K1 and K2) were measured for right and left eyes in 154 school-age children. Right and left eyes were analyzed separately. Each optical parameter obtained by 2 examiners was compared by paired samples t test. Intraobserver and interobserver correlation coefficients (r) were calculated for each variable. Significance was attributed when p<0.05.
RESULTS: Measurements could be performed in 152 of the children included. The mean measurements were not different statistically except for left AL (23.22±0.81 vs 23.21±0.82 mm; p=0.014), left CCT (548.5±32.3 vs 548.0±32.7 µm; p=0.013), and right ACD (3.19±0.25 vs 3.19±0.25 mm; p=0.033). Intraobserver r values were all above 0.957 except for left LT (r=0.786) for examiner 2. Interobserver r values were all above 0.979.
CONCLUSIONS: Although statistical significance was obtained in some variables, they were clinically negligible. The intraobserver and interobserver relation coefficients were very high for all optical parameters including the r value for left LT for examiner 2 (r=0.786). Lenstar is a highly reproducible, user independent, and tolerable instrument for ocular biometry in the pediatric population, with no anesthesia or discomfort.

Entities:  

Mesh:

Year:  2011        PMID: 20853265     DOI: 10.5301/ejo.2010.1298

Source DB:  PubMed          Journal:  Eur J Ophthalmol        ISSN: 1120-6721            Impact factor:   2.597


  4 in total

1.  Comparison of anterior segment parameters and axial lengths of myopic, emmetropic, and hyperopic children.

Authors:  Mehmethan Dogan; Ufuk Elgin; Emine Sen; Kemal Tekin; Pelin Yilmazbas
Journal:  Int Ophthalmol       Date:  2017-12-29       Impact factor: 2.031

2.  Intraobserver reliability of contact pachymetry in children.

Authors:  Katherine K Weise; Brett Kaminski; Michele Melia; Michael X Repka; Yasmin S Bradfield; Bradley V Davitt; David A Johnson; Raymond T Kraker; Ruth E Manny; Noelle S Matta; Susan Schloff
Journal:  J AAPOS       Date:  2013-04       Impact factor: 1.220

3.  Precision of a new ocular biometer in children and comparison with IOLMaster.

Authors:  Xinxin Yu; Hao Chen; Giacomo Savini; Qianqian Zheng; Benhao Song; Ruixue Tu; Jinhai Huang; Qinmei Wang
Journal:  Sci Rep       Date:  2018-01-22       Impact factor: 4.379

4.  Effect of orthokeratology on precision and agreement assessment of a new swept-source optical coherence tomography biometer.

Authors:  Bao Shu; Fangjun Bao; Giacomo Savini; Weicong Lu; Ruixue Tu; Haisi Chen; Benhao Song; Qinmei Wang; Jinhai Huang
Journal:  Eye Vis (Lond)       Date:  2020-03-02