Comparison of 3 biometry devices in cataract patients.

Susanne Goebels1, Max Pattmöller2, Timo Eppig2, Alan Cayless2, Berthold Seitz2, Achim Langenbucher2.   

Abstract

PURPOSE: Accurate biometry is an obligatory preoperative measurement for refractive surgery as well as cataract surgery. A new device based on partial coherence interferometry was compared with 2 currently used biometry devices.
SETTING: Department of Ophthalmology, Saarland University Medical Center, Homburg, Germany.
DESIGN: Prospective case series.
METHODS: Eyes of cataract patients were examined with a functional prototype of the new optical low-coherence reflectometry (OLCR) biometer OA-2000, the standard OLCR biometer Lenstar, and the partial coherence interferometry (PCI) biometer IOLMaster. The results were compared using a Wilcoxon-Mann-Whitney U test and Pearson correlation calculations.
RESULTS: A total of 138 eyes of 74 cataract patients were examined. Pearson correlation showed excellent correlation for axial length, anterior chamber depth and keratometry among the 3 devices tested. The highest correlation was found between standard OLCR biometer and PCI biometer for AL, R1, and R2 (r = 1.0, r = 0.936, r = 0.952, respectively; all P ≤ .001). For anterior chamber depth (ACD), the highest correlation was found between the standard OLCR biometer and the new OLCR biometer (r = 0.943; P ≤ .001). The mean values of AL/ACD/R1/R2 differed very little, but the differences were significant (all P ≤ .05) (new OLCR biometer 23.31/3.21/7.74/7.64 mm; standard OLCR biometer 23.30/3.13/7.80/7.60 mm; PCI biometer 23.37/3.00/7.78/7.6 mm).
CONCLUSIONS: Compared with other clinical instruments, the new OLCR biometer generated the most accurate results. Differences especially in measurement of axial length were statistically but not clinically significant. The new OLCR biometer yielded results that correlated very well with the values of the PCI biometer and standard OLCR biometer. FINANCIAL DISCLOSURE: None of the authors has a financial or proprietary interest in any material or method mentioned.
Copyright © 2015 ASCRS and ESCRS. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2015        PMID: 26703487     DOI: 10.1016/j.jcrs.2015.05.028

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  23 in total

1.  Comparison of the measurements of a novel optical biometry: Nidek AL-Scan with Sirius and a ultrasound biometry.

Authors:  Çağatay Çağlar; Sücattin İlker Kocamış; Emre Demir; Mustafa Durmuş
Journal:  Int Ophthalmol       Date:  2016-07-08       Impact factor: 2.031

2.  Comparison of biometric measurements obtained by the Verion Image-Guided System versus the auto-refracto-keratometer.

Authors:  Cecilio Velasco-Barona; Guadalupe Cervantes-Coste; Erick Mendoza-Schuster; Claudia Corredor-Ortega; Nadia L Casillas-Chavarín; Alejandro Silva-Moreno; Manuel Garza-León; Roberto Gonzalez-Salinas
Journal:  Int Ophthalmol       Date:  2017-04-25       Impact factor: 2.031

3.  Accuracy of intraocular lens power calculation using three optical biometry measurement devices: the OA-2000, Lenstar-LS900 and IOLMaster-500.

Authors:  Olga Reitblat; Adi Levy; Guy Kleinmann; Ehud I Assia
Journal:  Eye (Lond)       Date:  2018-03-12       Impact factor: 3.775

4.  Comparison of ocular parameters of two biometric measurement devices in highly myopic eyes.

Authors:  Xiao-Xiao Guo; Ran You; Shan-Shan Li; Xiu-Fen Yang; Lu Zhao; Fan Zhang; Yan-Ling Wang; Xi Chen
Journal:  Int J Ophthalmol       Date:  2019-10-18       Impact factor: 1.779

5.  Repeatability of a new swept-source optical coherence tomographer and agreement with other three optical biometers.

Authors:  Shi-Ming Cheng; Jia-Sheng Zhang; Xu Shao; Ze-Tong Wu; Tian-Tian Li; Po Wang; Jun-Hai Lin; A-Yong Yu
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2022-02-16       Impact factor: 3.117

6.  Comparison of a new optical biometer and a standard biometer in cataract patients.

Authors:  Pipat Kongsap
Journal:  Eye Vis (Lond)       Date:  2016-10-17

7.  Precision of a new ocular biometer in children and comparison with IOLMaster.

Authors:  Xinxin Yu; Hao Chen; Giacomo Savini; Qianqian Zheng; Benhao Song; Ruixue Tu; Jinhai Huang; Qinmei Wang
Journal:  Sci Rep       Date:  2018-01-22       Impact factor: 4.379

8.  Comparison of ocular biometric measurements between a new swept-source optical coherence tomography and a common optical low coherence reflectometry.

Authors:  Rongrong Gao; Hao Chen; Giacomo Savini; Yaxin Miao; Xiaorui Wang; Jing Yang; Weiqi Zhao; Qinmei Wang; Jinhai Huang
Journal:  Sci Rep       Date:  2017-05-30       Impact factor: 4.379

9.  Precision of a new ocular biometer in eyes with cataract using swept source optical coherence tomography combined with Placido-disk corneal topography.

Authors:  Wenwen Wang; Yaxin Miao; Giacomo Savini; Colm McAlinden; Hao Chen; Qingjie Hu; Qinmei Wang; Jinhai Huang
Journal:  Sci Rep       Date:  2017-10-23       Impact factor: 4.379

10.  Precision (repeatability and reproducibility) of ocular parameters obtained by the Tomey OA-2000 biometer compared to the IOLMaster in healthy eyes.

Authors:  Yanjun Hua; Wei Qiu; Qiuyi Xiao; Qiang Wu
Journal:  PLoS One       Date:  2018-02-27       Impact factor: 3.240

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.