Repeatability, reproducibility, and agreement of corneal power measurements obtained with a new corneal topographer.

Xinjie Mao1, Giacomo Savini, Zuopao Zhuo, Yifan Feng, Jian Zhang, Qinmei Wang, Hao Chen, Jinhai Huang.   

Abstract

PURPOSE: To study the repeatability and reproducibility of corneal power measurements obtained with a new corneal topographer (Keratograph 4) and assess their agreement with those obtained by a rotating Scheimpflug camera (Pentacam HR) and an automated keratometer (IOLMaster).
SETTING: Eye Hospital, Wenzhou Medical University, Wenzhou, China.
DESIGN: Observational cross-sectional study.
METHODS: One eye of normal subjects was measured 3 times using all devices. Another operator performed an additional 3 consecutive scans using the corneal topographer. One week later, the first operator repeated the corneal power measurements using the corneal topographer. Parameters were flat meridian power, steep meridian power, average power, J0, and J45. The repeatability and reproducibility of measurements were assessed by the within-subject standard deviation (Sw), repeatability (2.77 Sw), coefficient of variation (CoV), and intraclass correlation coefficient (ICC). Agreement between devices was assessed using 95% limits of agreement (LoA).
RESULTS: Intraoperator repeatability and interoperator and intersession reproducibility of all measured parameters showed a CoV of less than 0.3%, a 2.77 Sw of 0.32 diopter or less, and an ICC of more than 0.97. No significant differences in the parameters were noted between the corneal topographer and Scheimpflug camera. Statistically significant differences existed between the parameters analyzed by the corneal topographer and the automated keratometer, except J45. The mean differences between the corneal topographer and the other 2 devices were small, and the 95% LoA were narrow for all measurements.
CONCLUSION: The new corneal topographer had excellent reliability and high agreement with the other 2 devices in corneal power measurements in normal subjects. FINANCIAL DISCLOSURE: No author has a financial or proprietary interest in any material or method mentioned. Crown
Copyright © 2013. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2013        PMID: 23860010     DOI: 10.1016/j.jcrs.2013.04.029

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  13 in total

1.  Simulated Keratometry Repeatability in Subjects with and without Down Syndrome.

Authors:  Ayeswarya Ravikumar; Jason D Marsack; Julia S Benoit; Heather A Anderson
Journal:  Optom Vis Sci       Date:  2016-11       Impact factor: 1.973

2.  Repeatability of Ophtha Top topography and comparison with IOL-Master and LenstarLS900 in cataract patients.

Authors:  Sha-Sha Yu; Hui Song; Xin Tang
Journal:  Int J Ophthalmol       Date:  2017-11-18       Impact factor: 1.779

3.  Comparison of Total Corneal Astigmatism between IOLMaster and Pentacam.

Authors:  Xiaochun Li; Xiaoguang Cao; Yongzhen Bao
Journal:  Biomed Res Int       Date:  2022-07-08       Impact factor: 3.246

4.  Analysis of Age, Gender, and Refractive Error-Related Changes of the Anterior Corneal Surface Parameters Using Oculus Keratograph Topography.

Authors:  Nasrin Moghadas Sharif; Negareh Yazdani; Leila Shahkarami; Hadi Ostadi Moghaddam; Asieh Ehsaei
Journal:  J Curr Ophthalmol       Date:  2020-07-04

5.  Precision and agreement of corneal power measurements obtained using a new corneal topographer OphthaTOP.

Authors:  Jinhai Huang; Giacomo Savini; Hao Chen; Fangjun Bao; Yuanguang Li; Haisi Chen; Weicong Lu; Ye Yu; Qinmei Wang
Journal:  PLoS One       Date:  2015-01-05       Impact factor: 3.240

6.  Precision (Repeatability and Reproducibility) and Agreement of Corneal Power Measurements Obtained by Topcon KR-1W and iTrace.

Authors:  Yanjun Hua; Zequan Xu; Wei Qiu; Qiang Wu
Journal:  PLoS One       Date:  2016-01-11       Impact factor: 3.240

7.  Precision of a new ocular biometer in children and comparison with IOLMaster.

Authors:  Xinxin Yu; Hao Chen; Giacomo Savini; Qianqian Zheng; Benhao Song; Ruixue Tu; Jinhai Huang; Qinmei Wang
Journal:  Sci Rep       Date:  2018-01-22       Impact factor: 4.379

8.  Comparison of corneal power obtained from VERION image-guided surgery system and four other devices.

Authors:  Hung-Yuan Lin; Hsin-Yang Chen; Han Bor Fam; Ya-Jung Chuang; Ronald Yeoh; Pi-Jung Lin
Journal:  Clin Ophthalmol       Date:  2017-07-12

9.  Corneal thickness measurements with Scheimpflug and slit scanning imaging techniques in keratoconus.

Authors:  Mohammad Aghazadeh Amiri; Hassan Hashemi; Shahroukh Ramin; Abbasali Yekta; Azadeh Taheri; Payam Nabovati; Mehdi Khabazkhoob
Journal:  J Curr Ophthalmol       Date:  2016-10-08

10.  Comparison of Anterior Segment Biometric Measurements between Pentacam HR and IOLMaster in Normal and High Myopic Eyes.

Authors:  Jing Dong; Maolong Tang; Yaqin Zhang; Yading Jia; Haining Zhang; Zhijie Jia; Xiaogang Wang
Journal:  PLoS One       Date:  2015-11-17       Impact factor: 3.240

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.