Jinhai Huang1, Giacomo Savini2, Chengfang Wang3, Weicong Lu1, Rongrong Gao3, Yuanguang Li3, Qinmei Wang1, Yune Zhao1.
Abstract
Purpose. To assess the reliability and comparability of measuring central corneal thickness (CCT) and thinnest corneal thickness (TCT) using a new Scheimpflug-Placido analyzer (TMS-5, Japan) and ultrasound (US) pachymetry. Methods. Seventy-six healthy subjects were prospectively measured 3 times by 1 operator using the TMS-5, 3 additional consecutive scans were performed by a second operator, and ultrasound (US) pachymetry measurements were taken. The test-retest repeatability (TRT), coefficient of variation (CoV), and intraclass correlation coefficient (ICC) were calculated to evaluate intraoperator repeatability and interoperator reproducibility. Agreement among the devices was assessed using Bland-Altman plots and 95% limits of agreement (LoA). Results. The intraoperators TRT and CoV were <19 μm and 2.0%, respectively. The interoperators TRT and CoV were <12 μm and 1.0%, respectively, and ICC was >0.90. The mean CCT and TCT measurements using the TMS-5 were 15.97 μm (95% LoA from -26.42 to -5.52 μm) and 20.32 μm (95% LoA from -30.67 to -9.97 μm) smaller, respectively, than those using US pachymetry. Conclusions. The TMS-5 shows good repeatability and reproducibility for measuring CCT and TCT in normal subjects but only moderate agreement with US pachymetry results. Caution is warranted before using these techniques interchangeably.Entities:
Year: 2015 PMID: 25810919 PMCID: PMC4355121 DOI: 10.1155/2015/328798
Source DB: PubMed Journal: J Ophthalmol ISSN: 2090-004X Impact factor: 1.909
Intraoperator repeatability of central corneal thickness (CCT) and the thinnest corneal thickness (TCT) readings using the Scheimpflug-Placido topographic modeling system (TMS-5).
| Parameters | Mean ± SD ( | Sw ( | 2.77 Sw ( | CoV (%) | ICC (95% CI) |
|---|---|---|---|---|---|
| 1st observer | |||||
| CCT | 525.22 ± 30.41 | 6.51 | 18.05 | 1.24 | 0.955 (0.936 to 0.970) |
| TCT | 520.87 ± 30.49 | 6.60 | 18.27 | 1.27 | 0.955 (0.934 to 0.970) |
| 2nd observer | |||||
| CCT | 525.67 ± 29.62 | 6.32 | 17.50 | 1.20 | 0.956 (0.935 to 0.971) |
| TCT | 521.43 ± 29.64 | 6.39 | 17.70 | 1.23 | 0.955 (0.934 to 0.970) |
SD: standard deviation; Sw: within-subject standard deviation; CoV: coefficient of variation; ICC: intraclass correlation coefficient; CI: confidence interval.
Interoperator reproducibility of central corneal thickness (CCT) and the thinnest corneal thickness (TCT) readings by the Scheimpflug-Placido topographic modeling system (TMS-5).
| Parameters | Sw ( | 2.77 Sw ( | CoV (%) | ICC (95% CI) |
|---|---|---|---|---|
| CCT | 4.09 | 11.32 | 0.78 | 0.981 (0.971 to 0.988) |
| TCT | 4.01 | 11.11 | 0.77 | 0.982 (0.972 to 0.989) |
Sw: within-subject standard deviation; CoV: coefficient of variation; ICC: intraclass correlation coefficient; CI: confidence interval.
Comparison of central corneal thickness (CCT) and the thinnest corneal thickness (TCT) readings by the Scheimpflug-Placido topographic modeling system (TMS-5) and ultrasound pachymetry.
| Parameters | Mean difference ± SD |
| Lower 95% LoA (95% CI) | Upper 95% LoA (95% CI) |
|---|---|---|---|---|
| CCT ( | −15.97 ± 5.33 | <0.001 | −26.42 (−28.53 to −24.33) | −5.52 (−7.63 to −3.42) |
| TCT ( | −20.32 ± 5.28 | <0.001 | −30.67 (−32.75 to −28.58) | −9.97 (−12.06 to −7.89) |
SD: standard deviation; LoA: limits of agreement; CI: confidence interval.
Figure 1Bland-Altman plots of agreement in the central corneal thickness measurement between the Scheimpflug-Placido topographic modeling system (TMS-5) and ultrasound pachymetry. The blue solid line indicates the mean difference (bias). The dashed lines represent the 95% limits of agreement. The upper and lower green solid lines represent the 95% limits of agreement.
Figure 2Bland-Altman plots of agreement in the thinnest corneal thickness measurement between the Scheimpflug-Placido topographic modeling system (TMS-5) and ultrasound pachymetry. The blue solid line indicates the mean difference (bias). The dashed lines represent the 95% limits of agreement. The upper and lower green solid lines represent the 95% limits of agreement.