Literature DB >> 19289385

Accuracy of a new partial coherence interferometry analyser for biometric measurements.

M P Holzer1, M Mamusa, G U Auffarth.   

Abstract

AIMS: Precise biometry is an essential preoperative measurement for refractive surgery as well as cataract surgery. A new device based on partial coherence interferometry technology was tested and evaluated for accuracy of measurements.
METHODS: In a prospective study 200 eyes of 100 healthy phakic volunteers were examined with a functional prototype of the new ALLEGRO BioGraph (Wavelight AG)/LENSTAR LS 900 (Haag Streit AG) biometer and with the IOLMaster V.5 (Carl Zeiss Meditec AG). As recommended by the manufacturers, repeated measurements were performed with both devices and the results compared using Spearman correlation calculations (WinSTAT).
RESULTS: Spearman correlation showed high correlations for axial length and keratometry measurements between the two devices tested. Anterior chamber depth, however, had a lower correlation between the two biometry devices. In addition, the mean values of the anterior chamber depth differed (IOLMaster 3.48 (SD 0.42) mm versus BioGraph/LENSTAR 3.64 (SD 0.26) mm); however, this difference was not statistically different (p>0.05, t test).
CONCLUSION: The new biometer provided results that correlated very well with those of the IOLMaster. The ALLEGRO BioGraph/LENSTAR LS 900 is a precise device containing additional features that will be helpful tools for any cataract or refractive surgeon.

Entities:  

Mesh:

Year:  2009        PMID: 19289385     DOI: 10.1136/bjo.2008.152736

Source DB:  PubMed          Journal:  Br J Ophthalmol        ISSN: 0007-1161            Impact factor:   4.638


  45 in total

1.  [Effect of applanation tonometry on precision of biometry measurements].

Authors:  T M Rabsilber; S Bueltmann; M P Holzer; G U Auffarth
Journal:  Ophthalmologe       Date:  2011-01       Impact factor: 1.059

2.  Comparison of optical low-coherence reflectometry and applanation ultrasound biometry on intraocular lens power calculation.

Authors:  Mirjana Bjeloš Rončević; Mladen Bušić; Ivan Cima; Biljana Kuzmanović Elabjer; Damir Bosnar; Daliborka Miletić
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2010-09-18       Impact factor: 3.117

3.  Intraobserver and interobserver repeatability of ocular components measurement in cataract eyes using a new optical low coherence reflectometer.

Authors:  Mirjana Bjeloš Rončević; Mladen Bušić; Ivan Cima; Biljana Kuzmanović Elabjer; Damir Bosnar; Daliborka Miletić
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2010-10-28       Impact factor: 3.117

4.  Comparison of anterior segment measurements using rotating Scheimpflug imaging and partial coherence interferometry.

Authors:  Mohammad Soleimani; Hassan Hashemi; Shiva Mehravaran; Mehdi Khabazkhoob; Mohammad Hassan Emamian; Mohammad Shariati; Akbar Fotouhi
Journal:  Int J Ophthalmol       Date:  2013-08-18       Impact factor: 1.779

5.  Challenges and approaches in modern biometry and IOL calculation.

Authors:  Wolfgang Haigis
Journal:  Saudi J Ophthalmol       Date:  2012-01

6.  Agreement analysis of LENSTAR with other techniques of biometry.

Authors:  S Jasvinder; T F Khang; K K S Sarinder; V P Loo; V Subrayan
Journal:  Eye (Lond)       Date:  2011-03-11       Impact factor: 3.775

7.  [Comparison of IOL-Master 500 vs. Lenstar LS900 concerning the calculation of target refraction: A retrospective analysis].

Authors:  M Stattin; C Zehetner; N E Bechrakis; L Speicher
Journal:  Ophthalmologe       Date:  2015-05       Impact factor: 1.059

8.  Agreement of keratometric readings measured using rotating Scheimpflug imaging, auto-refractokeratometer, and biograph in eyes with keratoconus.

Authors:  Meltem Guzin Altınel; Hasim Uslu
Journal:  Int Ophthalmol       Date:  2021-02-05       Impact factor: 2.031

9.  Assessment of eye length changes in accommodation using dynamic extended-depth OCT.

Authors:  Yu-Cherng Chang; Keke Liu; Carolina de Freitas; Alex Pham; Florence Cabot; Siobhan Williams; Ethan Adre; Giovanni Gregori; Marco Ruggeri; Sonia H Yoo; Arthur Ho; Jean-Marie Parel; Fabrice Manns
Journal:  Biomed Opt Express       Date:  2017-04-26       Impact factor: 3.732

10.  Haigis and SRKT formulae accuracy for intentional myopic overcorrection.

Authors:  Rodrigo Faeda Dalto; Miriam A Ferreira; Wilian Queiroz; Roberto Pinto Coelho; Jayter Silva Paula; Andre Messias
Journal:  Int Ophthalmol       Date:  2017-06-21       Impact factor: 2.031

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.