Comparison of keratometric values of healthy and diseased eyes measured by Javal keratometer, EyeSys, and PAR.

D Varssano1, C J Rapuano, J I Luchs.   

Abstract

PURPOSE: To compare the simulated keratometric results of the PAR CTS and the EyeSys corneal imaging systems with Javal keratometer readings in keratoconus eyes, in eyes after corneal grafting, and in healthy eyes.
SETTING: Cornea Service, Wills Eye Hospital, Philadelphia, Pennsylvania, USA.
METHODS: This prospective study evaluated 69 patients divided into three groups (keratoconus; 6 to 12 months postkeratoplasty; normal corneas) treated at the cornea service. The eyes were measured with each of the three instruments. Keratometry results and number of attempts needed to obtain the results were collected.
RESULTS: Thirty keratoconus, 18 transplanted, and 21 normal corneas were examined. The PAR CTS Imaged all corneas and the EyeSys, 86% (P = .000627). The PAR CTS and Javal flat and steep K-readings in all three groups were the same. The EyeSys simulated keratometry results were lower than those of Javal keratometer in the flat K (P < .00001) and steep K (P < .00001) in the normal group and in the steep K in the transplanted cornea group (P = .00823). The EyeSys also measured less astigmatism than the Javal in the normal (P = .00124) and transplanted cornea groups (P = .00376).
CONCLUSION: The PAR CTS was better able than the EyeSys to provide keratometric values that agreed with those obtained with the Javal keratometer.

Entities:  

Mesh:

Year:  1997        PMID: 9159687     DOI: 10.1016/s0886-3350(97)80187-3

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  3 in total

1.  Keratometry with five different techniques: a study of device repeatability and inter-device agreement.

Authors:  Shiva Mehravaran; Soheila Asgari; Sara Bigdeli; Ashkan Shahnazi; Hassan Hashemi
Journal:  Int Ophthalmol       Date:  2014-02-23       Impact factor: 2.031

2.  Precision and agreement of corneal power measurements obtained using a new corneal topographer OphthaTOP.

Authors:  Jinhai Huang; Giacomo Savini; Hao Chen; Fangjun Bao; Yuanguang Li; Haisi Chen; Weicong Lu; Ye Yu; Qinmei Wang
Journal:  PLoS One       Date:  2015-01-05       Impact factor: 3.240

3.  A comprehensive assessment of the precision and agreement of anterior corneal power measurements obtained using 8 different devices.

Authors:  Qinmei Wang; Giacomo Savini; Kenneth J Hoffer; Zhen Xu; Yifan Feng; Daizong Wen; Yanjun Hua; Feng Yang; Chao Pan; Jinhai Huang
Journal:  PLoS One       Date:  2012-09-25       Impact factor: 3.240

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.