A novel Hartman Shack-based topography system: repeatability and agreement for corneal power with Scheimpflug+Placido topographer and rotating prism auto-keratorefractor.

Gaurav Prakash1, Dhruv Srivastava2, Sounak Choudhuri2.   

Abstract

The purpose of this study is to analyze the repeatability and agreement of corneal power using a new Hartman type topographer in comparison to Scheimpflug+Placido and autorefractor devices. In this cross sectional, observational study performed at the cornea services of a specialty hospital, 100 normal eyes (100 consecutive candidates) without any previous ocular surgery or morbidity except refractive error were evaluated. All candidates underwent three measurements each on a Full gradient, Hartman type topographer (FG) (iDesign, AMO), Scheimpflug+Placido topographer (SP) (Sirius, CSO) and rotating prism auto-keratorefractor (AR) (KR1, Nidek). The parameters assessed were flat keratometry (K1), steep keratometry (K2), steep axis (K2 axis), mean K, J 0 and J 45. Intra-device repeatability and inter-device agreement were evaluated. On repeatability analysis, the intra-device means were not significantly different (ANOVA, p > 0.05). Intraclass correlations (ICC) were >0.98 except for J 0 and J 45. In terms of intra-measurement standard deviation (Sw), the SP and FG groups fared better than AR group (p < 0.001, ANOVA). On Sw versus Average plots, no significantly predictive fit was seen (p > 0.05, R (2) < 0.1 for all the values). On inter-device agreement analysis, there was no difference in means (ANOVA, p > 0.05). ICC ranged from 0.92 to 0.99 (p < 0.001). Regression fits on Bland-Altman plots suggested no clinically significant effect of average values over difference in means. The repeatability of Hartman type topographer in normal eyes is comparable to SP combination device and better than AR. The agreement between the three devices is good. However, we recommend against interchanging these devices between follow-ups or pooling their data.

Entities:  

Keywords:  Auto keratorefractor; Full gradient topography; Hartman Shack topography; Repeatability; Scheimpflug topography

Mesh:

Year:  2015        PMID: 25820517     DOI: 10.1007/s10792-015-0065-7

Source DB:  PubMed          Journal:  Int Ophthalmol        ISSN: 0165-5701            Impact factor:   2.031


  20 in total

1.  Validation of a combined corneal topographer and aberrometer based on Shack-Hartmann wave-front sensing.

Authors:  Fan Zhou; Xin Hong; Donald T Miller; Larry N Thibos; Arthur Bradley
Journal:  J Opt Soc Am A Opt Image Sci Vis       Date:  2004-05       Impact factor: 2.129

Review 2.  Corneal imaging with slit-scanning and Scheimpflug imaging techniques.

Authors:  Cristina M Oliveira; Celina Ribeiro; Sandra Franco
Journal:  Clin Exp Optom       Date:  2010-08-16       Impact factor: 2.742

3.  Corneal topographer based on the Hartmann test.

Authors:  Yobani Mejía; Janneth C Galeano
Journal:  Optom Vis Sci       Date:  2009-04       Impact factor: 1.973

Review 4.  Measuring the cornea: the latest developments in corneal topography.

Authors:  Tracy Swartz; Lisa Marten; Ming Wang
Journal:  Curr Opin Ophthalmol       Date:  2007-07       Impact factor: 3.761

5.  Measurement error.

Authors:  J M Bland; D G Altman
Journal:  BMJ       Date:  1996-09-21

6.  Corneal topography reconstruction algorithm that avoids the skew ray ambiguity and the skew ray error.

Authors:  S A Klein
Journal:  Optom Vis Sci       Date:  1997-11       Impact factor: 1.973

7.  Comparison of keratometry measurements using the Pentacam HR, the Orbscan IIz, and the TMS-4 topographer.

Authors:  Zahra Tajbakhsh; Ramin Salouti; Mohammad H Nowroozzadeh; Mohammad Aghazadeh-Amiri; Seyedmehdi Tabatabaee; Mohammad Zamani
Journal:  Ophthalmic Physiol Opt       Date:  2012-11       Impact factor: 3.117

8.  Corneal thickness and elevation measurements using swept-source optical coherence tomography and slit scanning topography in normal and keratoconic eyes.

Authors:  Vishal Jhanji; Bingzhi Yang; Marco Yu; Cong Ye; Christopher K S Leung
Journal:  Clin Exp Ophthalmol       Date:  2013-04-25       Impact factor: 4.207

9.  Corneal ray tracing versus simulated keratometry for estimating corneal power changes after excimer laser surgery.

Authors:  Giacomo Savini; Antonio Calossi; Massimo Camellin; Francesco Carones; Marco Fantozzi; Kenneth J Hoffer
Journal:  J Cataract Refract Surg       Date:  2014-05-27       Impact factor: 3.351

10.  Comparison and repeatability of keratometric and corneal power measurements obtained by Orbscan II, Pentacam, and Galilei corneal tomography systems.

Authors:  Alexandra Z Crawford; Dipika V Patel; Charles N J McGhee
Journal:  Am J Ophthalmol       Date:  2013-03-28       Impact factor: 5.258

View more
  3 in total

1.  Intra-session repeatability of iridocorneal angle measurements provided by a Scheimpflug photography-based system in healthy eyes.

Authors:  Clara Ruiz-Belda; David P Piñero; Pedro Ruiz-Fortes; Roberto Soto-Negro; Myriam Moya; Rafael J Pérez-Cambrodí; Alberto Artola
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2015-07-15       Impact factor: 3.117

2.  Repeatability and Agreement of Central Corneal Thickness and Keratometry Measurements between Four Different Devices.

Authors:  Laszlo Kiraly; Jana Stange; Kathleen S Kunert; Saadettin Sel
Journal:  J Ophthalmol       Date:  2017-03-05       Impact factor: 1.909

3.  Comparison of Ocular Monochromatic Higher-order Aberrations in Normal Refractive Surgery Candidates of Arab and South Asian Origin.

Authors:  Gaurav Prakash; Dhruv Srivastava; Sounak Choudhuri; Ruthchel Bacero
Journal:  Middle East Afr J Ophthalmol       Date:  2016 Jan-Mar
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.