The performance of four different corneal topographers on normal human corneas and its impact on orthokeratology lens fitting.

Pauline Cho1, Andrew K C Lam, John Mountford, Larry Ng.   

Abstract

PURPOSE: To evaluate the performances of Humphrey Atlas 991, Orbscan II, Dicon CT200, Medmont E300 on young Chinese adults.
METHODS: Three sets of corneal topography measurements were obtained from each topographer from 22 subjects-two sets by the same examiner and one set by another examiner on the same day.
RESULTS: There were no significant within-examiner and between-examiner differences for any of the parameters tested for each topographer. However, only the repeatability and reproducibility (of apical radius[Ro], eccentricity, and elevation) of the Humphrey and Medmont were good. There was no statistically significant between-topographer difference in R(o), but significant differences in eccentricity and elevation values were found. The number of repeated readings that should be taken for a precision of 2 microm (elevation) were 12 for the Humphrey and 2 for the Medmont.
CONCLUSIONS: The performance of both the Humphrey and the Medmont was very good. R(o) and eccentricity values of different topographers cannot be used interchangeably, but the agreement in elevation values was good for these topographers. The number of repeated readings required for maximum precision varies with the topographer used, and they are not interchangeable.

Entities:  

Mesh:

Year:  2002        PMID: 11915858     DOI: 10.1097/00006324-200203000-00012

Source DB:  PubMed          Journal:  Optom Vis Sci        ISSN: 1040-5488            Impact factor:   1.973


  6 in total

1.  Corneal topography from spectral optical coherence tomography (sOCT).

Authors:  Sergio Ortiz; Damian Siedlecki; Pablo Pérez-Merino; Noelia Chia; Alberto de Castro; Maciej Szkulmowski; Maciej Wojtkowski; Susana Marcos
Journal:  Biomed Opt Express       Date:  2011-11-04       Impact factor: 3.732

2.  In vivo study of corneal responses to increased intraocular pressure loading.

Authors:  Ahmed Elsheikh; Charles W McMonnies; Charles Whitford; Gavin C Boneham
Journal:  Eye Vis (Lond)       Date:  2015-12-10

3.  Precision and agreement of corneal power measurements obtained using a new corneal topographer OphthaTOP.

Authors:  Jinhai Huang; Giacomo Savini; Hao Chen; Fangjun Bao; Yuanguang Li; Haisi Chen; Weicong Lu; Ye Yu; Qinmei Wang
Journal:  PLoS One       Date:  2015-01-05       Impact factor: 3.240

4.  Repeatability and Reproducibility of Anterior Segment Measurements in Normal Eyes Using Dual Scheimpflug Analyzer.

Authors:  Zeynep Altıparmak; Ramazan Yağcı; Emre Güler; Zeynel Arslanyılmaz; Metin Canbal; İbrahim F Hepşen
Journal:  Turk J Ophthalmol       Date:  2015-12-05

Review 5.  Orthokeratology: clinical utility and patient perspectives.

Authors:  Jessie Charm
Journal:  Clin Optom (Auckl)       Date:  2017-02-07

6.  A comprehensive assessment of the precision and agreement of anterior corneal power measurements obtained using 8 different devices.

Authors:  Qinmei Wang; Giacomo Savini; Kenneth J Hoffer; Zhen Xu; Yifan Feng; Daizong Wen; Yanjun Hua; Feng Yang; Chao Pan; Jinhai Huang
Journal:  PLoS One       Date:  2012-09-25       Impact factor: 3.240

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.