A comparison of the ARK-700A autokeratometer and Medmont E300 corneal topographer when measuring peripheral corneal curvature.

J M González-Méijome1, J Jorge, A Queiros, J B Almeida, M A Parafita.   

Abstract

PURPOSE: The purpose of this study was to evaluate if corneal topogometry measurements as taken with the ARK-700A autokeratometer are comparable with those measured by the Medmont E300 videokeratoscope at corresponding locations.
METHODS: Central and peripheral radius and eccentricity were measured in 122 right eyes of young normal subjects using autokeratometry and videokeratoscopy, obtained in a random order.
RESULTS: Curvature measurements obtained with the ARK700A and Medmont E300 correlated well for the central cornea. Larger differences were observed between peripheral autokeratometry readings and the empirically determined corresponding locations with the videokeratoscope. Correlations between the instruments are inconsistent, resulting in unacceptable confidence intervals. Corneal eccentricity was significantly different between the instruments for the vertical (t = 2.4; p = 0.018) and for the horizontal meridians. In the first case, the difference between the averaged values was not clinically significant, but in the horizontal meridian the AK significantly overestimated eccentricity values (t = -11.5; p < 0.001) with differences which were clinically significant.
CONCLUSIONS: Central corneal curvature data obtained by ARK700A and Medmont E300 can be interchanged but the same is not true of peripheral determinations of corneal shape. ARK700A probably measures peripheral corneal shape within an elliptical region between 5 and 7 mm in diameter, with the major axis in the vertical meridian.

Entities:  

Mesh:

Year:  2004        PMID: 15315653     DOI: 10.1111/j.1475-1313.2004.00203.x

Source DB:  PubMed          Journal:  Ophthalmic Physiol Opt        ISSN: 0275-5408            Impact factor:   3.117


  2 in total

1.  Which keratometer is most reliable for correcting astigmatism with toric intraocular lenses?

Authors:  Minwook Chang; Su-Yeon Kang; Hyo Myung Kim
Journal:  Korean J Ophthalmol       Date:  2012-01-14

2.  A comprehensive assessment of the precision and agreement of anterior corneal power measurements obtained using 8 different devices.

Authors:  Qinmei Wang; Giacomo Savini; Kenneth J Hoffer; Zhen Xu; Yifan Feng; Daizong Wen; Yanjun Hua; Feng Yang; Chao Pan; Jinhai Huang
Journal:  PLoS One       Date:  2012-09-25       Impact factor: 3.240

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.