Accuracy and precision of EyeSys and Orbscan systems on calibrated spherical test surfaces.

Javier González Pérez1, A Cerviño, M J Giraldez, M Parafita, E Yebra-Pimentel.   

Abstract

BACKGROUND: It is important to know how well the surface topography can be measured with current devices for corneal topographic analysis. There are several applications that need an accurate and precise method to measure corneal shape and variations, such as the effect of contact lens wear and the different refractive surgery techniques.
PURPOSE: The aim of this study is to compare the accuracy and reproducibility of the measurement of the central curvature on calibrated steel balls using the EyeSys videokeratoscope and the Orbscan corneal topography system.
METHODS: The videokeratoscope (EyeSys Corneal Analysis System 2000, version 3.1) and the Orbscan corneal topography system (Orbscan II version 3.0) were used by four trained investigators to measure a series of five uniform and calibrated test surfaces with known radius of curvature ranging from 6.13 to 9.00 mm. RESULT: No statistically significant difference was found between the videokeratoscope and Orbscan systems in relation to accuracy or precision. The 95% confidence limits showed a close agreement between both instruments. The mean bias was less than +/-0.05 mm for both devices. The precision of the instruments was found to be similar.
CONCLUSION: The EyeSys seems to measure more accurately, but the accuracy of the Orbscan was also acceptable, suggesting that both instruments are accurate and precise enough for research and clinical purposes. However, further studies of accuracy and repeatability of topographical measurements on human eyes generated by different topographers are necessary.

Entities:  

Mesh:

Year:  2004        PMID: 15260351     DOI: 10.1097/01.icl.0000111749.04644.92

Source DB:  PubMed          Journal:  Eye Contact Lens        ISSN: 1542-2321            Impact factor:   2.018


  6 in total

1.  Changes in corneal thickness and curvature after different excimer laser photorefractive procedures and their impact on intraocular pressure measurements.

Authors:  Helena Svedberg; Enping Chen; Helene Hamberg-Nyström
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2005-07-08       Impact factor: 3.117

2.  Repeatability of Ophtha Top topography and comparison with IOL-Master and LenstarLS900 in cataract patients.

Authors:  Sha-Sha Yu; Hui Song; Xin Tang
Journal:  Int J Ophthalmol       Date:  2017-11-18       Impact factor: 1.779

3.  Wound stability and surgically induced corneal astigmatism after transconjunctival single-plane sclerocorneal incision cataract surgery.

Authors:  Ken Hayashi; Soichiro Ogawa; Motoaki Yoshida; Koichi Yoshimura
Journal:  Jpn J Ophthalmol       Date:  2016-10-06       Impact factor: 2.447

4.  Precision and agreement of corneal power measurements obtained using a new corneal topographer OphthaTOP.

Authors:  Jinhai Huang; Giacomo Savini; Hao Chen; Fangjun Bao; Yuanguang Li; Haisi Chen; Weicong Lu; Ye Yu; Qinmei Wang
Journal:  PLoS One       Date:  2015-01-05       Impact factor: 3.240

5.  OPD scan III accuracy: Topographic and aberrometric indices after accelerated corneal cross-linking.

Authors:  Soheila Asgari; Hassan Hashemi
Journal:  J Curr Ophthalmol       Date:  2017-11-14

6.  A comprehensive assessment of the precision and agreement of anterior corneal power measurements obtained using 8 different devices.

Authors:  Qinmei Wang; Giacomo Savini; Kenneth J Hoffer; Zhen Xu; Yifan Feng; Daizong Wen; Yanjun Hua; Feng Yang; Chao Pan; Jinhai Huang
Journal:  PLoS One       Date:  2012-09-25       Impact factor: 3.240

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.