Corneal power measurement with Fourier-domain optical coherence tomography.

Maolong Tang1, Alex Chen, Yan Li, David Huang.   

Abstract

PURPOSE: To study the accuracy and repeatability of anterior, posterior, and net corneal power measured by Fourier-domain optical coherence tomography (OCT).
SETTING: Doheny Eye Institute, Los Angeles, California, USA.
DESIGN: Cross-sectional study.
METHODS: A Fourier-domain OCT system (RTVue) was used to scan normal eyes, eyes after myopic laser in situ keratomileusis (LASIK), and keratoconic eyes. After the corneal surfaces were delineated, the system calculated anterior and posterior corneal powers by curve fitting over the central 3.0 mm diameter area. Net corneal power was calculated using a thick-lens formula. The repeatability of the calculations was evaluated by the pooled standard deviation of 3 measurements from the same visit. The net corneal power values were compared with standard automated keratometry measurements (IOLMaster).
RESULTS: The repeatability of Fourier-domain OCT net corneal power was 0.19 diopters (D), 0.26 D, and 0.30 D in the normal, post-LASIK, and keratoconus groups, respectively. The Fourier-domain OCT net corneal power was significantly lower than keratometry by a mean of -1.21 D, -2.89 D, and -3.07 D, respectively (P<.001). The anterior-posterior curvature ratio was lower in post-LASIK and keratoconic eyes than in normal eyes (P<.001).
CONCLUSIONS: Corneal power measured by Fourier-domain OCT achieved good repeatability in all 3 groups. The repeatability was better than slower time-domain OCT systems. Because Fourier-domain OCT directly measures both anterior and posterior corneal surfaces, it may produce more consistent results than standard keratometry in post-LASIK and keratoconic eyes in which the anterior-posterior corneal curvature ratios are altered by surgery or disease.
Copyright © 2010 ASCRS and ESCRS. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2010        PMID: 21111315      PMCID: PMC3005697          DOI: 10.1016/j.jcrs.2010.07.018

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  24 in total

1.  Accuracy of Orbscan total optical power maps in detecting refractive change after myopic laser in situ keratomileusis.

Authors:  S Srivannaboon; D Z Reinstein; H F Sutton; S P Holland
Journal:  J Cataract Refract Surg       Date:  1999-12       Impact factor: 3.351

2.  Accuracy of Orbscan optical pachymetry in corneas with haze.

Authors:  Francesco Boscia; Maria Gabriella La Tegola; Giovanni Alessio; Carlo Sborgia
Journal:  J Cataract Refract Surg       Date:  2002-02       Impact factor: 3.351

3.  Reliability of pachymetric measurements using orbscan after excimer refractive surgery.

Authors:  Olivier Prisant; Nila Calderon; Philippe Chastang; Damien Gatinel; Thanh Hoang-Xuan
Journal:  Ophthalmology       Date:  2003-03       Impact factor: 12.079

4.  A direct method to measure the power of the central cornea after myopic laser in situ keratomileusis.

Authors:  Sergio Sónego-Krone; Gerson López-Moreno; Oscar V Beaujon-Balbi; Carlos G Arce; Paulo Schor; Mauro Campos
Journal:  Arch Ophthalmol       Date:  2004-02

5.  Reliability and repeatability of the Pentacam on corneal curvatures.

Authors:  Davie Chen; Andrew K C Lam
Journal:  Clin Exp Optom       Date:  2008-10-30       Impact factor: 2.742

6.  Comparison of corneal powers obtained from 4 different devices.

Authors:  Mariko Shirayama; Li Wang; Mitchell P Weikert; Douglas D Koch
Journal:  Am J Ophthalmol       Date:  2009-07-09       Impact factor: 5.258

7.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

8.  Reproducibility of optical biometry using partial coherence interferometry : intraobserver and interobserver reliability.

Authors:  A Vogel; H B Dick; F Krummenauer
Journal:  J Cataract Refract Surg       Date:  2001-12       Impact factor: 3.351

9.  A comparative analysis of five methods of determining corneal refractive power in eyes that have undergone myopic laser in situ keratomileusis.

Authors:  Abdelmonem M Hamed; Li Wang; Manjula Misra; Douglas D Koch
Journal:  Ophthalmology       Date:  2002-04       Impact factor: 12.079

10.  Clinical and theoretical results of intraocular lens power calculation for cataract surgery after photorefractive keratectomy for myopia.

Authors:  Monica T P Odenthal; Cathrien A Eggink; Gerrit Melles; Jan H Pameyer; Annette J M Geerards; W Houdijn Beekhuis
Journal:  Arch Ophthalmol       Date:  2002-04
View more
  31 in total

1.  Intraocular lens power calculation after myopic and hyperopic laser vision correction using optical coherence tomography.

Authors:  Maolong Tang; Li Wang; Douglas D Koch; Yan Li; David Huang
Journal:  Saudi J Ophthalmol       Date:  2012-01

2.  Wide-field whole eye OCT system with demonstration of quantitative retinal curvature estimation.

Authors:  Ryan P McNabb; James Polans; Brenton Keller; Moseph Jackson-Atogi; Charlene L James; Robin R Vann; Joseph A Izatt; Anthony N Kuo
Journal:  Biomed Opt Express       Date:  2018-12-21       Impact factor: 3.732

3.  Assessment of total corneal power after myopic corneal refractive surgery in Chinese eyes.

Authors:  Yanjun Hua; Chao Pan; Qinmei Wang
Journal:  Int Ophthalmol       Date:  2019-03-01       Impact factor: 2.031

4.  Golden angle based scanning for robust corneal topography with OCT.

Authors:  Joerg Wagner; David Goldblum; Philippe C Cattin
Journal:  Biomed Opt Express       Date:  2017-01-03       Impact factor: 3.732

5.  Intraocular lens power calculation after previous myopic laser vision correction based on corneal power measured by Fourier-domain optical coherence tomography.

Authors:  Maolong Tang; Li Wang; Douglas D Koch; Yan Li; David Huang
Journal:  J Cataract Refract Surg       Date:  2012-04       Impact factor: 3.351

6.  Segmentation guided registration of wide field-of-view retinal optical coherence tomography volumes.

Authors:  José Lezama; Dibyendu Mukherjee; Ryan P McNabb; Guillermo Sapiro; Anthony N Kuo; Sina Farsiu
Journal:  Biomed Opt Express       Date:  2016-11-01       Impact factor: 3.732

7.  Optical coherence tomography accurately measures corneal power change from laser refractive surgery.

Authors:  Ryan P McNabb; Sina Farsiu; Sandra S Stinnett; Joseph A Izatt; Anthony N Kuo
Journal:  Ophthalmology       Date:  2014-12-06       Impact factor: 12.079

8.  Quantitative single and multi-surface clinical corneal topography utilizing optical coherence tomography.

Authors:  Ryan P McNabb; Anthony N Kuo; Joseph A Izatt
Journal:  Opt Lett       Date:  2013-04-15       Impact factor: 3.776

9.  Optical coherence tomography-based corneal power measurement and intraocular lens power calculation following laser vision correction (an American Ophthalmological Society thesis).

Authors:  David Huang; Maolong Tang; Li Wang; Xinbo Zhang; Rebecca L Armour; Devin M Gattey; Lorinna H Lombardi; Douglas D Koch
Journal:  Trans Am Ophthalmol Soc       Date:  2013-09

10.  A comprehensive assessment of the precision and agreement of anterior corneal power measurements obtained using 8 different devices.

Authors:  Qinmei Wang; Giacomo Savini; Kenneth J Hoffer; Zhen Xu; Yifan Feng; Daizong Wen; Yanjun Hua; Feng Yang; Chao Pan; Jinhai Huang
Journal:  PLoS One       Date:  2012-09-25       Impact factor: 3.240

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.