Reproducibility of optical biometry using partial coherence interferometry : intraobserver and interobserver reliability.

A Vogel1, H B Dick, F Krummenauer.   

Abstract

PURPOSE: To evaluate the intraobserver and interobserver variability in axial length (AL), anterior chamber depth (ACD), and corneal radius measurements using an optical biometry instrument based on partial coherence interferometry (PCI).
SETTING: Johannes Gutenberg University, Mainz, Germany.
METHODS: In this observational case series and interobserver reliability trial, 30 healthy, emmetropic to moderately myopic eyes of 15 volunteers were evaluated. The AL, ACD, and corneal radius were measured 20 times in 10 eyes by 1 observer to evaluate the intraobserver variability. To evaluate the interobserver variability, the measurements were taken in 20 eyes by 5 different observers. Measurements were performed using the IOLMaster (Carl Zeiss) based on PCI. The data description is based on coefficients of variation and the statistical inference on reliability estimation based on analysis of variance. The main outcome measures were intraobserver and interobserver variability and reliability in AL, ACD, and corneal radii.
RESULTS: The intraobserver variability (SD) was +/-25.6 microm for AL, +/-33.4 microm for ACD, and +/-12.9 microm for corneal radius. The coefficients of variation were 0.1%, 0.9%, and 0.17%, respectively. The interobserver variability (SD) was +/-21.5 microm for AL, +/-29.8 microm for ACD, and +/-15.9 microm for corneal radius. The coefficients of variation were 0.09%, 0.82%, and 0.21%, respectively. The reliability was 99.9% for AL, 97.8% for ACD, and 99.8%/99.5% for corneal radius (r1/r2). The slightly reduced reliability in ACD measurement was caused by a reduced intraobserver reliability of 1 of the 5 observers (R = 0.87).
CONCLUSION: Partial coherence biometry using the IOLMaster for AL measurement was highly reliable, offering observer-independent measurement results.

Entities:  

Mesh:

Year:  2001        PMID: 11738911     DOI: 10.1016/s0886-3350(01)01214-7

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  40 in total

1.  IOLMaster biometry: refractive results of 100 consecutive cases.

Authors:  H Eleftheriadis
Journal:  Br J Ophthalmol       Date:  2003-08       Impact factor: 4.638

2.  Comparison of optical low-coherence reflectometry and applanation ultrasound biometry on intraocular lens power calculation.

Authors:  Mirjana Bjeloš Rončević; Mladen Bušić; Ivan Cima; Biljana Kuzmanović Elabjer; Damir Bosnar; Daliborka Miletić
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2010-09-18       Impact factor: 3.117

3.  Relationship between the retinal thickness of the macula and the difference in axial length.

Authors:  Tetsuo Ueda; Yoshiaki Nawa; Yoshiaki Hara
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2005-08-13       Impact factor: 3.117

4.  Comparison of anterior segment measurements using rotating Scheimpflug imaging and partial coherence interferometry.

Authors:  Mohammad Soleimani; Hassan Hashemi; Shiva Mehravaran; Mehdi Khabazkhoob; Mohammad Hassan Emamian; Mohammad Shariati; Akbar Fotouhi
Journal:  Int J Ophthalmol       Date:  2013-08-18       Impact factor: 1.779

5.  Agreement analysis of LENSTAR with other techniques of biometry.

Authors:  S Jasvinder; T F Khang; K K S Sarinder; V P Loo; V Subrayan
Journal:  Eye (Lond)       Date:  2011-03-11       Impact factor: 3.775

6.  Comparison of IOL-Master 700 and IOL-Master 500 biometers in ocular biological parameters of adolescents.

Authors:  Qiang Shi; Guang-Yan Wang; Yu-Hong Cheng; Cheng Pei
Journal:  Int J Ophthalmol       Date:  2021-07-18       Impact factor: 1.779

7.  Corneal power measurement with Fourier-domain optical coherence tomography.

Authors:  Maolong Tang; Alex Chen; Yan Li; David Huang
Journal:  J Cataract Refract Surg       Date:  2010-12       Impact factor: 3.351

8.  Evaluation of biometric methods for measuring the anterior chamber depth in the non-contact mode.

Authors:  B Meinhardt; O Stachs; J Stave; R Beck; R Guthoff
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2005-09-15       Impact factor: 3.117

9.  Comparison of anterior chamber depth measurements using the IOLMaster, scanning peripheral anterior chamber depth analyser, and anterior segment optical coherence tomography.

Authors:  Raghavan Lavanya; Livia Teo; David S Friedman; Han T Aung; Mani Baskaran; Hong Gao; Tamuno Alfred; Steve K Seah; Kenji Kashiwagi; Paul J Foster; Tin Aung
Journal:  Br J Ophthalmol       Date:  2007-02-27       Impact factor: 4.638

10.  Changes in macular thickness and depth of anterior chamber in patients after filtration surgery.

Authors:  Gulsara Karasheva; Winfried Goebel; Thomas Klink; Wolfgang Haigis; Franz Grehn
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2003-02-20       Impact factor: 3.117

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.