Comparison of anterior segment measurements with rotating Scheimpflug photography and partial coherence reflectometry.

Jinhai Huang1, Konrad Pesudovs, Daizong Wen, Shihao Chen, Thomas Wright, Xiaoyu Wang, Yini Li, Qinmei Wang.   

Abstract

PURPOSE: To compare central corneal thickness (CCT), anterior chamber depth (ACD), and keratometry (K) readings measured using optical low-coherence reflectometry (OLCR) biometry and high-resolution rotating Scheimpflug photography.
SETTING: Eye Hospital of Wenzhou Medical College, Wenzhou, China.
DESIGN: Comparative case series.
METHODS: The CCT, ACD endothelium to lens, ACD epithelium to lens, and K (mean; in flattest meridian; in steepest meridian) were measured 5 times using the LenStar/Biograph OLCR biometer and 3 times with the Pentacam Scheimpflug system in eyes of healthy volunteers. Concordance was evaluated using paired t tests, the Pearson correlation, and Bland-Altman analyses.
RESULTS: The CCT, ACD endothelium to lens, and ACD epithelium to lens measured with the Scheimpflug system were slightly, albeit significantly, higher than with the OLCR biometer (P<.05); the respective 95% limits of agreement (LoA) were -8.2 μm to 15.7 μm, -0.11 to 0.15 mm, and -0.13 to 0.17 mm. However, the Scheimpflug system gave significantly flatter readings for K in the flattest meridian (95% LoA, -0.54 to 0.32 diopters [D]), K in the steepest meridian (95% LoA, -0.63 to 0.45 D), and mean K (95% LoA, -0.53 to 0.33 D) (P<.001). The CCT, ACD, and K readings were all highly correlated between the 2 devices (r >0.95, P<.001).
CONCLUSIONS: The CCT and ACD measurements with the OLCR biometer and Scheimpflug system can be used interchangeably in healthy young subjects. However, for K measurements, these devices have wide LoA so may not be interchangeable under certain clinical circumstances. Crown
Copyright © 2011. Published by Elsevier Inc. All rights reserved.

Mesh:

Year:  2011        PMID: 21241919     DOI: 10.1016/j.jcrs.2010.08.044

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  15 in total

1.  Comparison of anterior segment measurements using rotating Scheimpflug imaging and partial coherence interferometry.

Authors:  Mohammad Soleimani; Hassan Hashemi; Shiva Mehravaran; Mehdi Khabazkhoob; Mohammad Hassan Emamian; Mohammad Shariati; Akbar Fotouhi
Journal:  Int J Ophthalmol       Date:  2013-08-18       Impact factor: 1.779

2.  Anterior chamber depth in normal subjects by rotating scheimpflug imaging.

Authors:  Matthew T Feng; Michael W Belin; Renato Ambrósio; Satinder P S Grewal; Wang Yan; Mohamed S Shaheen; Charles McGhee; Naoyuki Maeda; Tobias H Neuhann; H Burkhard Dick; Saleh A Alageel; Andreas Steinmueller
Journal:  Saudi J Ophthalmol       Date:  2011-04-30

3.  Agreement of keratometric readings measured using rotating Scheimpflug imaging, auto-refractokeratometer, and biograph in eyes with keratoconus.

Authors:  Meltem Guzin Altınel; Hasim Uslu
Journal:  Int Ophthalmol       Date:  2021-02-05       Impact factor: 2.031

Review 4.  Device interchangeability on anterior chamber depth and white-to-white measurements: a thorough literature review.

Authors:  Alberto Domínguez-Vicent; Cari Pérez-Vives; Teresa Ferrer-Blasco; Santiago García-Lázaro; Robert Montés-Micó
Journal:  Int J Ophthalmol       Date:  2016-07-18       Impact factor: 1.779

5.  Repeatability of Ophtha Top topography and comparison with IOL-Master and LenstarLS900 in cataract patients.

Authors:  Sha-Sha Yu; Hui Song; Xin Tang
Journal:  Int J Ophthalmol       Date:  2017-11-18       Impact factor: 1.779

6.  Comparison of corneal curvature and anterior chamber depth measurements using the manual keratometer, Lenstar LS 900 and the Pentacam.

Authors:  Omür Ö Uçakhan; Veysel Akbel; Zeynep Bıyıklı; Ayfer Kanpolat
Journal:  Middle East Afr J Ophthalmol       Date:  2013 Jul-Sep

7.  Central and midperipheral corneal thickness measured with Scheimpflug imaging and optical coherence tomography.

Authors:  Jinhai Huang; Xixia Ding; Giacomo Savini; Zhengxuan Jiang; Chao Pan; Yanjun Hua; Fan Wu; Yifan Feng; Ye Yu; Qinmei Wang
Journal:  PLoS One       Date:  2014-05-22       Impact factor: 3.240

8.  Clear-cornea cataract surgery: pupil size and shape changes, along with anterior chamber volume and depth changes. A Scheimpflug imaging study.

Authors:  Anastasios John Kanellopoulos; George Asimellis
Journal:  Clin Ophthalmol       Date:  2014-10-24

9.  Pupil Dilation with Intracameral Epinephrine Hydrochloride during Phacoemulsification and Intraocular Lens Implantation.

Authors:  A-Yong Yu; Hua Guo; Qin-Mei Wang; Fang-Jun Bao; Jing-Hai Huang
Journal:  J Ophthalmol       Date:  2016-01-20       Impact factor: 1.909

10.  A comprehensive assessment of the precision and agreement of anterior corneal power measurements obtained using 8 different devices.

Authors:  Qinmei Wang; Giacomo Savini; Kenneth J Hoffer; Zhen Xu; Yifan Feng; Daizong Wen; Yanjun Hua; Feng Yang; Chao Pan; Jinhai Huang
Journal:  PLoS One       Date:  2012-09-25       Impact factor: 3.240

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.