Reliability and repeatability of the Pentacam on corneal curvatures.

Davie Chen1, Andrew K C Lam.   

Abstract

BACKGROUND: The purpose of this study was to assess the intra-observer reliability and inter-session repeatability of the Pentacam system (Oculus, Inc.) on the simulated keratometry readings (Sim K) and peripheral corneal curvatures (PCC) at both the anterior and posterior corneal surfaces.
METHODS: One eye was randomly selected from each of the 39 normal subjects from the School of Optometry at The Hong Kong Polytechnic University. In the first session, three readings were taken for reliability analysis. Measurements were repeated in a second session scheduled one to two weeks later for repeatability analysis. Two approaches were used, comparing the first reading between the two sessions and comparing the average from three readings between the two sessions. Both the anterior and posterior Sim K and PCC were calculated from the tangential and axial maps. The inter-session repeatability was reported as the coefficient of repeatability (COR = +/- 1.96 x standard deviation of differences). Relative repeatability (RR) was a percentage of the ratio of COR to the mean.
RESULTS: All parameters showed good intra-observer reliability (Cronbach's alpha > or = 0.925; intra-class correlation coefficient, ICC > or = 0.804), except for the posterior power vector J45 (= 0.896; ICC = 0.742) and tangential PCC at the superior region (= 0.891; ICC = 0.732). The inter-session repeatability on anterior Sim K (COR < or = +/-0.21 D; RR < or = 0.48 per cent) and posterior Sim K (COR < or = +/-0.09 D; RR < or = 1.39 per cent) was good. Axial PCC showed better repeatability than the tangential PCC. The COR from using the first approach for all the posterior PCC was within 0.23 D, while the COR reduced to within 0.16 D from using the second approach.
CONCLUSIONS: The Pentacam showed good intra-observer reliability on Sim K and PCC measurements. Averaging three readings from one visit could improve repeatability between visits.

Entities:  

Mesh:

Year:  2008        PMID: 18983633     DOI: 10.1111/j.1444-0938.2008.00336.x

Source DB:  PubMed          Journal:  Clin Exp Optom        ISSN: 0816-4622            Impact factor:   2.742


  28 in total

1.  The effects of tolterodine on anterior segment and choroidal thickness in patients with overactive bladder syndrome.

Authors:  Hande Hüsniye Telek; Omer Gokhan Doluoglu; Ayse Burcu; Gulizar Demirok; Firdevs Ornek; Ali Ayyildiz
Journal:  Ther Adv Urol       Date:  2017-04-11

2.  [Scheimpflug photography of acute angle closure].

Authors:  J D Unterlauft; Y Yafai; P Wiedemann
Journal:  Ophthalmologe       Date:  2013-09       Impact factor: 1.059

3.  Repeatability and comparability of anterior segment biometry obtained by the Sirius and the Pentacam analyzers.

Authors:  Paola De la Parra-Colín; Manuel Garza-León; Tonatiuh Barrientos-Gutierrez
Journal:  Int Ophthalmol       Date:  2013-04-25       Impact factor: 2.031

4.  Scheimpflug imaged corneal changes on anterior and posterior surfaces after collagen cross-linking.

Authors:  Ziad Hassan; Laszlo Modis; Eszter Szalai; Andras Berta; Gabor Nemeth
Journal:  Int J Ophthalmol       Date:  2014-04-18       Impact factor: 1.779

5.  Corneal power measurement with Fourier-domain optical coherence tomography.

Authors:  Maolong Tang; Alex Chen; Yan Li; David Huang
Journal:  J Cataract Refract Surg       Date:  2010-12       Impact factor: 3.351

6.  Data quality and clinical decision-making: do we trust machines blindly?

Authors:  Konrad Pesudovs; Raymond A Applegate
Journal:  Clin Exp Optom       Date:  2009-05       Impact factor: 2.742

7.  Changes of anterior chamber architecture induced by laser peripheral iridotomy in acute angle closure crisis.

Authors:  J D Unterlauft; Y Yafai; P Wiedemann
Journal:  Int Ophthalmol       Date:  2014-08-02       Impact factor: 2.031

8.  A comprehensive assessment of the precision and agreement of anterior corneal power measurements obtained using 8 different devices.

Authors:  Qinmei Wang; Giacomo Savini; Kenneth J Hoffer; Zhen Xu; Yifan Feng; Daizong Wen; Yanjun Hua; Feng Yang; Chao Pan; Jinhai Huang
Journal:  PLoS One       Date:  2012-09-25       Impact factor: 3.240

9.  Corneal Elevation, Power, and Astigmatism to Assess Toric Orthokeratology Lenses in Moderate-to-High Astigmats.

Authors:  Erin S Tomiyama; Anna-Kaye Logan; Kathryn Richdale
Journal:  Eye Contact Lens       Date:  2021-02-01       Impact factor: 3.152

10.  Intra- and Intersession Repeatability of an Optical Quality and Intraocular Scattering Measurement System in Children.

Authors:  Mi Tian; Huamao Miao; Yang Shen; Jian Gao; Xiaofen Mo; Xingtao Zhou
Journal:  PLoS One       Date:  2015-11-04       Impact factor: 3.240

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.