Tommy C Y Chan1,2, Marco C Y Yu1,3, Vivian Chiu1, Gilda Lai1, Christopher K S Leung1,4,5, Poemen P M Chan6,7.
Abstract
To evaluate the repeatability and agreement of corneal and biometry measurements obtained with two swept-source optical coherence tomography (SSOCT) and a partial coherence interferometry-based device. This is a cross-sectional study. Forty-eight eyes of 48 patients had three consecutive measurements for ANTERION (Heidelberg Engineering, Germany), CASIAII (Tomey, Japan) and IOLMaster500 (Carl Zeiss Meditec, USA) on the same visit. Mean keratometry (Km), central corneal thickness (CCT), anterior chamber depth (ACD) and axial length (AL) were recorded. Corneal astigmatic measurements were converted into vector components-J0 and J45. Intra-device repeatability and agreements of measurements amongst the devices were evaluated using repeatability coefficients (RCs) and Bland-Altman plots, respectively. All devices demonstrated comparable repeatability for Km (p ≥ 0.138). ANTERION had the lowest RC for J0 amongst the devices (p ≤ 0.039). Systematic difference was found for the Km and J0 obtained with IOLMaster500 compared to either SSOCTs (p ≤ 0.010). The ACD and AL measured by IOLMaster500 showed a higher RC compared with either SSOCTs (p < 0.002). Systematic difference was found in CCT and ACD between the two SSOCTs (p < 0.001), and in AL between ANTERION and IOLMaster500 (p < 0.001), with a mean difference of 1.6 µm, 0.022 mm and 0.021 mm, respectively. Both SSOCTs demonstrated smaller test-retest variability for measuring ACD and AL compared with IOLMaster500. There were significant disagreement in keratometry and AL measurements between the SSOCTs and PCI-based device; their measurements should not be considered as interchangeable.Entities:
Year: 2021 PMID: 34290281 PMCID: PMC8295353 DOI: 10.1038/s41598-021-93999-8
Source DB: PubMed Journal: Sci Rep ISSN: 2045-2322 Impact factor: 4.379
Repeatability outcomes for biometric measurements obtained using the ANTERION, CASIA II and IOLMaster500.
| Parameters | Devices | Mean (± SD) | RC (95% CI) | p-value for pairwise comparison in RC | ||
|---|---|---|---|---|---|---|
| ANTERION vs CASIAII | ANTERION vs IOLMaster500 | CASIAII vs IOLMaster500 | ||||
| Km (D) | ANTERION | 43.878 ± 1.415 | 0.2345 (0.2010–0.2681) | 0.600 | 0.138 | 0.254 |
| CASIA II | 43.851 ± 1.440 | 0.2484 (0.2125–0.2843) | ||||
| IOLMaster500 | 43.902 ± 1.456 | 0.3148 (0.2666–0.3629) | ||||
| J0 (D) | ANTERION | 0.305 ± 0.457 | 0.2003 (0.1716–0.2289) | 0.002 | 0.039 | 0.361 |
| CASIA II | 0.376 ± 0.422 | 0.3217 (0.2752–0.3682) | ||||
| IOLMaster500 | 0.207 ± 0.469 | 0.5448 (0.4614–0.6282) | ||||
| J45 (D) | ANTERION | − 0.002 ± 0.180 | 0.2232 (0.1913–0.2551) | 0.980 | 0.245 | 0.072 |
| CASIA II | 0.029 ± 0.196 | 0.2115 (0.1809–0.2420) | ||||
| IOLMaster500 | − 0.024 ± 0.226 | 0.3336 (0.2825–0.3846) | ||||
| ACD (mm) | ANTERION | 2.899 ± 0.597 | 0.0311 (0.0264–0.0352) | 0.216 | < 0.001 | < 0.001 |
| CASIA II | 2.912 ± 0.586 | 0.0260 (0.0227–0.0298) | ||||
| IOLMaster500 | 2.932 ± 0.563 | 0.1947 (0.1652–0.2241) | ||||
| CCT (µm) | ANTERION | 543.17 ± 32.20 | 2.1759 (1.8648–2.4869) | 0.575 | – | – |
| CASIA II | 543.02 ± 33.45 | 2.3223 (1.9868–2.6579) | ||||
| AL (mm) | ANTERION | 23.683 ± 1.137 | 0.0154 (0.0130–0.0178) | – | 0.002 | – |
| IOLMaster500 | 23.881 ± 1.323 | 0.0260 (0.0221–0.0299) | ||||
Km mean keratometry, J0 J0 vector components of astigmatism, J45 J45 vector components of astigmatism, ACD anterior chamber depth, AL axial length, CCT central corneal thickness, RC repeatability coefficient.
10.9% of the 5 significant results out of the 14 tests are expected to be false discoveries due to random error as determined by the false discovery rate.
Systematic differences and proportional biases between different devices.
| Parameters | Systematic difference | Proportional bias | ||||
|---|---|---|---|---|---|---|
| Mean differences | 95% LoA | p | Scaling difference | R2 | p | |
| Km (D) | 0.026 | − 0.349 to 0.401 | 0.351 | 0.001 | 7.33 × 10–5 | 0.954 |
| J0 (D) | 0.045 | − 0.364 to 0.455 | 0.140 | − 0.041 | 0.008 | 0.557 |
| J45 (D) | 0.016 | − 0.256 to 0.289 | 0.419 | 0.058 | 0.007 | 0.581 |
| ACD (mm) | 0.022 | − 0.032 to 0.077 | < 0.001 | − 0.016 | 0.116 | 0.018 |
| CCT (µm) | − 1.625 | − 5.278 to 2.028 | < 0.001 | 0.012 | 0.045 | 0.149 |
| Km (D) | − 0.115 | − 0.588 to 0.357 | 0.002 | − 0.003 | 3.41 × 10–4 | 0.902 |
| J0 (D) | 0.153 | − 0.346 to 0.651 | < 0.001 | − 0.072 | 0.018 | 0.374 |
| J45 (D) | 0.044 | − 0.398 to 0.485 | 0.189 | − 0.274 | 0.058 | 0.104 |
| ACD (mm) | 0.002 | − 0.260 to 0.263 | 0.932 | 0.037 | 0.026 | 0.278 |
| AL (mm) | − 0.021 | − 0.055 to 0.013 | < 0.001 | 0.007 | 0.260 | < 0.001 |
| Km (D) | − 0.093 | − 0.562 to 0.376 | 0.010 | − 0.002 | 9.55 × 10–5 | 0.948 |
| J0 (D) | 0.195 | − 0.384 to 0.774 | < 0.001 | − 0.116 | 0.032 | 0.232 |
| J45 (D) | 0.067 | − 0.348 to 0.482 | 0.035 | − 0.201 | 0.039 | 0.185 |
| ACD (mm) | 0.024 | − 0.235 to 0.283 | 0.219 | 0.020 | 0.008 | 0.547 |
Km mean keratometry, J0 J0 vector components of astigmatism, J45 J45 vector components of astigmatism, ACD anterior chamber depth, CCT central corneal thickness, 95% LoA 95% limit of agreement.
9.8% of the 10 significant results out of the 28 tests are expected to be false discoveries due to random error as determined by the false discovery rate.
Figure 1Bland–Altman plots showing the pairwise agreement between ANTERION vs CASIAII (A), ANTERION vs IOLMaster500 (B), and CASIAII vs IOLMaster500 (C) for mean keratometry (Km).
Figure 2Bland–Altman plots showing the pairwise agreement between ANTERION vs CASIAII (A), ANTERION vs IOLMaster500 (B), and CASIAII vs IOLMaster500 (C) for J0 vector component of astigmatism.
Figure 3Bland–Altman plots showing the pairwise agreement between ANTERION vs CASIAII (A), ANTERION vs IOLMaster500 (B), and CASIAII vs IOLMaster500 (C) for J45 vector component of astigmatism.
Figure 4Bland–Altman plots showing the pairwise agreement between ANTERION vs CASIAII (A), ANTERION vs IOLMaster500 (B), and CASIAII vs IOLMaster500 (C) for anterior chamber depth (ACD).
Figure 5Bland–Altman plots showing the pairwise agreement between ANTERION vs CASIAII for central corneal thickness (CCT).
Figure 6Bland–Altman plots showing the pairwise agreement between ANTERION vs IOLMaster500 for axial length (AL).