| Literature DB >> 35766049 |
Ji Young Moon1, Soo Chang Cho1, Hyun Jin Kim1, Roo Min Jun2, Kyung Eun Han1.
Abstract
PURPOSE: To evaluate the level of agreement between ANTERION (Heidelberg Engineering, Heidelberg, Germany), OA-2000 (Tomey, Nagoya, Japan), and IOLMaster 500 (Carl Zeiss AG, Jena, Germany).Entities:
Keywords: Biometry; Partial coherence interferometer; Swept-source optical coherence tomography
Mesh:
Year: 2022 PMID: 35766049 PMCID: PMC9388892 DOI: 10.3341/kjo.2022.0017
Source DB: PubMed Journal: Korean J Ophthalmol ISSN: 1011-8942
Mean, standard deviation, and range of keratometry, vector components of astigmatism, anterior chamber depth, and axial length using four devices
| Variable | ANTERION | OA-2000 | IOLMaster 500 | AR/K | |
|---|---|---|---|---|---|
| Flat keratometry (D) | 43.58 ± 1.39 (39.88 to 46.28) | 43.83 ± 1.39 (39.99 to 46.30) | 43.93 ± 1.47 (40.23 to 47.14) | 43.87 ± 1.46 (40.25 to 46.75) | <0.001 |
| Steep keratometry (D) | 44.45 ± 1.22 (41.47 to 46.50) | 44.77 ± 1.29 (41.67 to 47.54) | 44.85 ± 1.30 (42.13 to 48.35) | 44.66 ± 1.30 (41.50 to 47.25) | <0.001 |
| 0.06 ± 0.39 (−0.80 to 1.57) | −0.09 ± 0.37 (−0.84 to 0.71) | 0.04 ± 0.42 (−1.34 to 1.38) | 0.02 ± 0.34 (−0.84 to 1.15) | 0.200 | |
| −0.07 ± 0.36 (−1.19 to 0.61) | −0.01 ± 0.43 (−1.43 to 1.59) | 0.06 ± 0.35 (−0.63 to 0.89) | −0.04 ± 0.34 (−1.09 to 0.76) | 0.307 | |
| Anterior chamber depth (mm) | 3.08 ± 0.38 (2.35 to 3.91) | 3.05 ± 0.37 (2.30 to 3.83) | 3.13 ± 0.39 (2.37 to 3.79) | - | 0.001 |
| Axial length (mm) | 23.64 ± 1.33 (21.50 to 27.63) | 23.64 ± 1.32 (21.51 to 27.62) | 23.61 ± 1.32 (21.49 to 27.61) | - | <0.001 |
Values are presented as mean ± standard deviation (range). The four devices are ANTERION (Heidelberg Engineering, Heidelberg, Germany), OA-2000 (Tomey, Nagoya, Japan), IOLMaster 500 (Carl Zeiss AG, Jena, Germany), and AR/K.
AR/K = autorefractor/keratometer; D = diopters; J = Jackson cross-cylinder at 0° and 90°; J = Jackson cross-cylinder at 45° and 135°.
Obtained from repeated measured analysis of variance test;
Vector components of astigmatism.
Difference, agreement, and correlation of flat and steep keratometry, J, and J using the four devices
| Variable | Mean difference ± SD | 95% CI | 95% LoA | Pearson correlation ( | |
|---|---|---|---|---|---|
| Flat keratometry (D) | |||||
| ANTERION vs. OA-2000 | −0.25 ± 0.22 | −0.34 to −0.17 | −0.68 to +0.18 | <0.001 | 0.987 (<0.001) |
| ANTERION vs. IOLMaster 500 | −0.35 ± 0.30 | −0.47 to −0.24 | −0.94 to +0.24 | <0.001 | 0.979 (<0.001) |
| OA-2000 vs. IOLMaster 500 | −0.10 ± 0.27 | −0.21 to 0 | −0.63 to +0.43 | 0.067 | 0.983 (<0.001) |
| AR/K vs. ANTERION | 0.29 ± 0.32 | +0.17 to +0.42 | −0.34 to +0.93 | <0.001 | 0.975 (<0.001) |
| AR/K vs. OA-2000 | 0.04 ± 0.32 | −0.08 to +0.16 | −0.59 to +0.67 | >0.999 | 0.976 (<0.001) |
| AR/K vs. IOLMaster 500 | −0.06 ± 0.32 | −0.18 to +0.06 | −0.69 to +0.57 | >0.999 | 0.976 (<0.001) |
| Steep keratometry (D) | |||||
| ANTERION vs. OA-2000 | −0.32 ± 0.24 | −0.41 to −0.23 | −0.78 to +0.15 | <0.001 | 0.984 (<0.001) |
| ANTERION vs. IOLMaster 500 | −0.40 ± 0.31 | −0.52 to −0.28 | −1.00 to +0.20 | <0.001 | 0.973 (<0.001) |
| OA-2000 vs. IOLMaster 500 | −0.08 ± 0.24 | −0.17 to +0.01 | −0.55 to +0.38 | 0.106 | 0.983 (<0.001) |
| AR/K vs. ANTERION | 0.21 ± 0.28 | +0.10 to +0.32 | −0.34 to +0.75 | <0.001 | 0.977 (<0.001) |
| AR/K vs. OA-2000 | −0.11 ± 0.31 | −0.23 to +0.01 | −0.72 to +0.50 | 0.093 | 0.971 (<0.001) |
| AR/K vs. IOLMaster 500 | −0.19 ± 0.34 | −0.32 to −0.06 | −0.86 to +0.48 | 0.001 | 0.965 (<0.001) |
| ANTERION vs. OA-2000 | 0.15 ± 0.60 | −0.07 to +0.38 | −1.02 to +1.32 | 0.420 | −0.241 (0.088) |
| ANTERION vs. IOLMaster 500 | 0.03 ± 0.64 | −0.22 to +0.27 | −1.23 to +1.28 | >0.999 | −0.263 (0.063) |
| OA-2000 vs. IOLMaster 500 | −0.13 ± 0.51 | −0.33 to +0.07 | −1.13 to +0.88 | 0.483 | 0.160 (0.261) |
| AR/K vs. ANTERION | −0.04 ± 0.59 | −0.27 to +0.18 | −1.19 to +1.11 | >0.999 | −0.289 (0.040) |
| AR/K vs. OA-2000 | 0.11 ± 0.48 | −0.07 to +0.30 | −0.83 to +1.05 | 0.578 | 0.110 (0.441) |
| AR/K vs. IOLMaster 500 | −0.01 ± 0.47 | −0.149 to +0.16 | −0.93 to +0.90 | >0.999 | 0.267 (0.058) |
| ANTERION vs. OA-2000 | −0.06 ± 0.55 | −0.28 to +0.15 | −1.14 to +1.01 | 0.420 | 0.042 (0.771) |
| ANTERION vs. IOLMaster 500 | −0.14 ± 0.50 | −0.33 to +0.05 | −1.12 to +0.85 | >0.999 | −0.009 (0.951) |
| OA-2000 vs. IOLMaster 500 | −0.07 ± 0.55 | −0.29 to +0.14 | −1.16 to +1.01 | 0.483 | 0.019 (0.894) |
| AR/K vs. ANTERION | 0.03 ± 0.52 | −0.17 to +0.23 | −0.98 to +1.05 | >0.999 | −0.111 (0.438) |
| AR/K vs. OA-2000 | −0.03 ± 0.64 | −0.28 to +0.22 | −1.29 to +1.23 | 0.578 | −0.380 (0.006) |
| AR/K vs. IOLMaster 500 | −0.11 ± 0.47 | −0.29 to +0.08 | −1.03 to +0.82 | >0.999 | 0.053 (0.714) |
The four devices are ANTERION (Heidelberg Engineering, Heidelberg, Germany), OA-2000 (Tomey, Nagoya, Japan), IOLMaster 500 (Carl Zeiss AG, Jena, Germany), and AR/K.
SD = standard deviation; CI = confidence interval; LoA = limits of agreement; D = diopters; AR/K = autorefractor/keratometer; J = Jackson cross-cylinder at 0° and 90°; J = Jackson cross-cylinder at 45° and 135°.
Obtained from repeated measured analysis of variance test with Bonferroni post hoc analysis;
Vector components of astigmatism.
Fig. 1Bland-Altman plots of (A–C) flat keratometry (K) and (D–F) steep K between ANTERION (Heidelberg Engineering, Heidelberg, Germany), OA-2000 (Tomey, Nagoya, Japan), and IOLMaster 500 (Carl Zeiss AG, Jena, Germany). The mean (middle continuous line) and the lower and upper limits of agreement (±1.96 standard deviation, top and bottom dotted lines) are depicted. D = diopters.
Fig. 2Bland-Altman plots of vector component of astigmatism (A–C) Jackson cross-cylinder at 0° and 90° (J) and (D–F) Jackson cross-cylinder at 45° and 135° (J) between ANTERION (Heidelberg Engineering, Heidelberg, Germany), OA-2000 (Tomey, Nagoya, Japan), and IOLMaster 500 (Carl Zeiss AG, Jena, Germany). The mean (middle continuous line) and the lower and upper limits of agreement (±1.96 standard deviation, top and bottom dotted lines) are depicted. D = diopters.
Difference, agreement, and correlation of anterior chamber depth and axial length using the three devices
| Variable | Mean difference ± SD | 95% CI | 95% LoA | Pearson correlation ( | |
|---|---|---|---|---|---|
| Anterior chamber depth (mm) | |||||
| ANTERION vs. OA-2000 | 0.03 ± 0.04 | +0.02 to +0.04 | −0.05 to +0.11 | <0.001 | 0.995 (<0.001) |
| ANTERION vs. IOLMaster 500 | −0.05 ± 0.14 | −0.10 to 0 | −0.33 to +0.22 | 0.030 | 0.934 (<0.001) |
| OA-2000 vs. IOLMaster 500 | −0.08 ± 0.15 | −0.13 to −0.03 | −0.37 to +0.20 | 0.001 | 0.927 (<0.001) |
| Axial length (mm) | |||||
| ANTERION vs. OA-2000 | −0.01 ± 0.04 | −0.02 to +0.01 | −0.09 to +0.08 | 0.917 | 1.000 (<0.001) |
| ANTERION vs. IOLMaster 500 | 0.02 ± 0.04 | 0.01 to +0.04 | −0.04 to +0.09 | <0.001 | 1.000 (<0.001) |
| OA-2000 vs. IOLMaster 500 | 0.03 ± 0.02 | 0.02 to +0.04 | −0.01 to +0.07 | <0.001 | 1.000 (<0.001) |
The three devices are ANTERION (Heidelberg Engineering, Heidelberg, Germany), OA-2000 (Tomey, Nagoya, Japan), and IOLMaster 500 (Carl Zeiss AG, Jena, Germany).
SD = standard deviation; CI = confidence interval; LoA = limits of agreement.
Obtained from repeated measured analysis of variance with Bonferroni post hoc analysis.
Fig. 3Bland-Altman plots of anterior chamber depth (ACD) between (A) ANTERION (Heidelberg Engineering, Heidelberg, Germany) and OA-2000 (Tomey, Nagoya, Japan), (B) ANTERION and IOLMaster 500 (Carl Zeiss AG, Jena, Germany), and (C) OA-2000 and IOLMaster 500. The mean (middle continuous line), the lower and upper limits of agreement (±1.96 standard deviation, top and bottom dotted lines) are depicted.
Fig. 4Bland-Altman plots of axial length (AL) between (A) ANTERION (Heidelberg Engineering, Heidelberg, Germany) and OA-2000 (Tomey, Nagoya, Japan), (B) ANTERION and IOLMaster 500 (Carl Zeiss AG, Jena, Germany), and (C) OA-2000 and IOLMaster 500. The mean (middle continuous line), the lower and upper limits of agreement (±1.96 standard deviation, top and bottom dotted lines) are depicted.
Mean, standard deviation, and range of predicted intraocular lens power using the three devices
| Formula | ANTERION | OA-2000 | IOLMaster 500 | |
|---|---|---|---|---|
| SRK/T | 20.80 ± 3.78 (10–27) | 20.44 ± 3.78 (10–26) | 20.46 ± 3.77 (10–26) | <0.001 |
| Haigis | 20.80 ± 3.76 (11–27) | 20.41 ± 3.75 (10–26) | 20.41 ± 3.70 (10–26) | <0.001 |
| Barrett Universal II | 20.71 ± 3.80 (11–27) | 20.30 ± 3.81 (10–26) | 20.33 ± 3.75 (10–26) | <0.001 |
| Kane | 21.12 ± 3.81 (11–27) | 20.71 ± 3.82 (10–27) | 20.76 ± 3.81 (10–27) | <0.001 |
Values are presented as mean ± standard deviation (range). The three devices are ANTERION (Heidelberg Engineering, Heidelberg, Germany), OA-2000 (Tomey, Nagoya, Japan), and IOLMaster 500 (Carl Zeiss AG, Jena, Germany).
Obtained from repeated measured analysis of variance test.
Difference, agreement, and correlation of predicted intraocular lens power using the three devices
| Formula | Mean difference ± SD | 95% CI | 95% LoA | Pearson correlation ( | |
|---|---|---|---|---|---|
| SRK/T | |||||
| ANTERION vs. OA-2000 | 0.36 ± 0.28 | +0.26 to +0.46 | −0.19 to +0.92 | <0.001 | 0.997 (<0.001) |
| ANTERION vs. IOLMaster 500 | 0.34 ± 0.35 | +0.22 to +0.47 | −0.35 to +1.04 | <0.001 | 0.996 (<0.001) |
| OA-2000 vs. IOLMaster 500 | −0.02 ± 0.32 | −0.13 to +0.09 | −0.64 to +0.60 | >0.999 | 0.997 (<0.001) |
| Haigis | |||||
| ANTERION vs. OA-2000 | 0.39 ± 0.34 | +0.28 to +0.51 | −0.27 to +1.05 | <0.001 | 0.996 (<0.001) |
| ANTERION vs. IOLMaster 500 | 0.39 ± 0.42 | +0.25 to +0.54 | −0.42 to +1.21 | <0.001 | 0.994 (<0.001) |
| OA-2000 vs. IOLMaster 500 | 0.00 ± 0.44 | −0.15 to +0.15 | −0.85 to +0.85 | >0.999 | 0.993 (<0.001) |
| Barrett Universal II | |||||
| ANTERION vs. OA-2000 | 0.40 ± 0.26 | +0.31 to +0.49 | −0.12 to +0.92 | <0.001 | 0.998 (<0.001) |
| ANTERION vs. IOLMaster 500 | 0.37 ± 0.43 | +0.22 to +0.52 | −0.48 to +1.22 | <0.001 | 0.993 (<0.001) |
| OA-2000 vs. IOLMaster 500 | −0.03 ± 0.35 | −0.15 to +0.09 | −0.72 to +0.66 | >0.999 | 0.996 (<0.001) |
| Kane | |||||
| ANTERION vs. OA-2000 | 0.40 ± 0.32 | +0.29 to +0.51 | −0.22 to +1.02 | <0.001 | 0.997 (<0.001) |
| ANTERION vs. IOLMaster 500 | 0.35 ± 0.40 | +0.21 to +0.49 | −0.44 to +1.14 | <0.001 | 0.994 (<0.001) |
| OA-2000 vs. IOLMaster 500 | −0.05 ± 0.39 | −0.18 to +0.09 | −0.81 to +0.72 | >0.999 | 0.995 (<0.001) |
The three devices are ANTERION (Heidelberg Engineering, Heidelberg, Germany), OA-2000 (Tomey, Nagoya, Japan), and IOLMaster 500 (Carl Zeiss AG, Jena, Germany).
SD = standard deviation; CI = confidence interval; LoA = limits of agreement.
Obtained from repeated measured analysis of variance with Bonferroni post hoc analysis.