Comparison of the Pentacam equivalent keratometry reading and IOL Master keratometry measurement in intraocular lens power calculations.

Nicholas Karunaratne1.   

Abstract

BACKGROUND: To compare the accuracy of the Pentacam Holladay equivalent keratometry readings with the IOL Master 500 keratometry in calculating intraocular lens power.
DESIGN: Non-randomized, prospective clinical study conducted in private practice. PARTICIPANTS: Forty-five consecutive normal patients undergoing cataract surgery.
METHODS: Forty-five consecutive patients had Pentacam equivalent keratometry readings at the 2-, 3 and 4.5-mm corneal zone and IOL Master keratometry measurements prior to cataract surgery. For each Pentacam equivalent keratometry reading zone and IOL Master measurement the difference between the observed and expected refractive error was calculated using the Holladay 2 and Sanders, Retzlaff and Kraff theoretic (SRKT) formulas. MAIN OUTCOME MEASURE: Mean keratometric value and mean absolute refractive error.
RESULTS: There was a statistically significantly difference between the mean keratometric values of the IOL Master, Pentacam equivalent keratometry reading 2-, 3- and 4.5-mm measurements (P < 0.0001, analysis of variance). There was no statistically significant difference between the mean absolute refraction error for the IOL Master and equivalent keratometry readings 2 mm, 3 mm and 4.5 mm zones for either the Holladay 2 formula (P = 0.14) or SRKT formula (P = 0.47). The lowest mean absolute refraction error for Holladay 2 equivalent keratometry reading was the 4.5 mm zone (mean 0.25 D ± 0.17 D). The lowest mean absolute refraction error for SRKT equivalent keratometry reading was the 4.5 mm zone (mean 0.25 D ± 0.19 D). Comparing the absolute refraction error of IOL Master and Pentacam equivalent keratometry reading, best agreement was with Holladay 2 and equivalent keratometry reading 4.5 mm, with mean of the difference of 0.02 D and 95% limits of agreement of -0.35 and 0.39 D.
CONCLUSIONS: The IOL Master keratometry and Pentacam equivalent keratometry reading were not equivalent when used only for corneal power measurements. However, the keratometry measurements of the IOL Master and Pentacam equivalent keratometry reading 4.5 mm may be similarly effective when used in intraocular lens power calculation formulas, following constant optimization.
© 2013 Royal Australian and New Zealand College of Ophthalmologists.

Entities:  

Keywords:  IOL Master keratometry; Pentacam; Scheimpflug; equivalent keratometry reading

Mesh:

Year:  2013        PMID: 23601493     DOI: 10.1111/ceo.12124

Source DB:  PubMed          Journal:  Clin Exp Ophthalmol        ISSN: 1442-6404            Impact factor:   4.207


  11 in total

1.  Precision of 5 different keratometry devices.

Authors:  Jonas Vejvad Nørskov Laursen; Peter Jeppesen; Thomas Olsen
Journal:  Int Ophthalmol       Date:  2015-03-29       Impact factor: 2.031

2.  Intraocular lens power calculation in patients with irregular astigmatism.

Authors:  Asaf Achiron; Omar Elhaddad; Duncan Leadbetter; Eliya Levinger; Oleksiy Voytsekhivskyy; Katy Smith; Venkata Avadhanam; Kieren Darcy; Derek Tole
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2022-07-01       Impact factor: 3.117

3.  Effects of posterior corneal astigmatism on the accuracy of AcrySof toric intraocular lens astigmatism correction.

Authors:  Bin Zhang; Jing-Xue Ma; Dan-Yan Liu; Cong-Rong Guo; Ying-Hua Du; Xiu-Jin Guo; Yue-Xian Cui
Journal:  Int J Ophthalmol       Date:  2016-09-18       Impact factor: 1.779

4.  Evaluation of biometry and corneal astigmatism in cataract surgery patients from Central China.

Authors:  Ji-Guo Yu; Jie Zhong; Zhong-Ming Mei; Fang Zhao; Na Tao; Yi Xiang
Journal:  BMC Ophthalmol       Date:  2017-04-26       Impact factor: 2.209

5.  Astigmatic Outcomes of Single, Non-Paired Intrastromal Limbal Relaxing Incisions During Femtosecond Laser-Assisted Cataract Surgery Based on a Custom Nomogram.

Authors:  Carter W Lim; Sohel Somani; Hannah H Chiu; Raj Maini; Eric S Tam
Journal:  Clin Ophthalmol       Date:  2020-04-22

6.  Calculation of the Real Corneal Refractive Power after Photorefractive Keratectomy Using Pentacam, When Only the Preoperative Refractive Error is Known.

Authors:  Maddalena De Bernardo; Maria Borrelli; Roberto Imparato; Nicola Rosa
Journal:  J Ophthalmol       Date:  2020-04-01       Impact factor: 1.909

7.  Evaluation and comparison of a novel Scheimpflug-based optical biometer with standard partial coherence interferometry for biometry and intraocular lens power calculation.

Authors:  Ziyang Wang; Wenli Yang; Dongjun Li; Wei Chen; Qi Zhao; Yifeng Li; Rui Cui; Lin Shen; Junfang Xian
Journal:  Exp Ther Med       Date:  2021-02-05       Impact factor: 2.447

8.  Comparison of two novel swept-source optical coherence tomography devices to a partial coherence interferometry-based biometer.

Authors:  Tommy C Y Chan; Marco C Y Yu; Vivian Chiu; Gilda Lai; Christopher K S Leung; Poemen P M Chan
Journal:  Sci Rep       Date:  2021-07-21       Impact factor: 4.379

9.  Comparison of Anterior Segment Biometric Measurements between Pentacam HR and IOLMaster in Normal and High Myopic Eyes.

Authors:  Jing Dong; Maolong Tang; Yaqin Zhang; Yading Jia; Haining Zhang; Zhijie Jia; Xiaogang Wang
Journal:  PLoS One       Date:  2015-11-17       Impact factor: 3.240

10.  Comparison of a new swept-source optical biometer with a partial coherence interferometry.

Authors:  Hyo Kyung Lee; Mee Kum Kim
Journal:  BMC Ophthalmol       Date:  2018-10-19       Impact factor: 2.209

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.