Xiaoting Ruan1, GuangYao Yang1, Zhaoxia Xia2, Jiaqing Zhang1, Xiaoxun Gu1, Yuan Tan1, Zhenzhen Liu1, Lixia Luo1.
Abstract
PURPOSE: To compare the difference and agreement in central corneal thickness (CCT), keratometry (K), anterior chamber depth (ACD), aqueous depth (AQD), and lens thickness (LT) measured with CASIA 2 and IOLMaster 700 in patients with cataract.Entities:
Keywords: CASIA 2; IOLMaster 700; agreement; anterior segment optical coherence tomography; anterior segment parameters
Year: 2022 PMID: 35223893 PMCID: PMC8874008 DOI: 10.3389/fmed.2022.777443
Source DB: PubMed Journal: Front Med (Lausanne) ISSN: 2296-858X
Baseline characteristics of the participants.
|
|
|
|---|---|
|
| |
| 61.5 ± 10.6 | |
|
| |
| Male, | 43 (53.1) |
| Female, | 38 (46.9) |
SD, Standard deviation.
Comparison of anterior segment parameter measurements with CASIA 2 and IOLMaster 700.
|
|
|
|
|
|
|
|---|---|---|---|---|---|
| CCT (μm) | |||||
| Mean ± SD | 530.96 ± 34.25 | 545.78 ± 34.40 | −14.81 ± 7.78 | 0.892, | |
| Range | 452–644 | 458–640 | |||
| Anterior Kf (D) | |||||
| Mean ± SD | 43.53 ± 1.59 | 43.45 ± 1.61 | 0.09 ± 0.20 | 0.991, | |
| Range | 39.90–47.10 | 39.85–47.00 | |||
| Anterior Ks (D) | |||||
| Mean ± SD | 44.30 ± 1.66 | 44.31 ± 1.67 | −0.02 ± 0.23 | 0.991, | |
| Range | 40.10–47.70 | 40.09–47.98 | |||
| Real Kf (D) | |||||
| Mean ± SD | 42.46 ± 1.57 | 43.45 ± 1.61 | −0.99 ± 0.25 | 0.827, | |
| Range | 38.90–46.10 | 39.85–47.00 | |||
| Real Ks (D) | |||||
| Mean ± SD | 43.26 ± 1.62 | 44.31 ± 1.67 | −1.06 ± 0.29 | 0.817, | |
| Range | 39.30–46.60 | 40.09–47.89 | |||
| ACD (mm) | |||||
| Mean ± SD | 3.36 ± 0.4 | 3.23 ± 0.39 | 0.13 ± 0.05 | 0.937, | |
| Range | 2.18–4.13 | 2.07–4.2 | |||
| AQD (mm) | |||||
| Mean ± SD | 2.83 ± 0.4 | 2.68 ± 0.39 | 0.15 ± 0.05 | 0.926, | |
| Range | 1.65–3.78 | 1.52–3.66 | |||
| LT (mm) | |||||
| Mean ± SD | 4.38 ± 0.54 | 4.36 ± 0.52 | 0.02 ± 0.04 | 0.997, | |
| Range | 3.23–5.67 | 3.24–5.67 |
K, keratometry; ACD, anterior chamber depth; AQD, aqueous depth; LT, lens thickness; CCT, central corneal thickness; D, diopter; mm, millimeter; μm, micron; SD, standard deviation; diff, difference; ICC, intraclass coefficient.
Figure 1Bland-Altman plots of cornea curvature measurement with CASIA2 and IOL-mater700. Kf, flat keratometric power; Ks, steep keratometric power. (A) Anterior Kf, (B) Anterior Ks, (C) Real Kf, and (D) Real Ks. Black dotted lines indicate the bias between both devices and red dotted lines indicate the 95% confidence interval for the difference.
Figure 2Bland-Altman plots of (A) ACD, (B) AQD, (C) CCT, and (D) LT measurements with CASIA2 and IOL-Master 700. ACD, anterior chamber depth; AQD, aqueous depth; CCT, central cornea thickness; LT, lens thickness.