Anmol A Pardeshi1, Abe E Song1, Naim Lazkani1, Xiaobin Xie2,3, Alex Huang3, Benjamin Y Xu1.
Abstract
Purpose: To assess the repeatability and agreement of ocular biometric parameters measured using the Tomey CASIA SS-1000 and Heidelberg ANTERION anterior segment optical coherence tomography (AS-OCT) devices.Entities:
Keywords: anterior segment OCT; ocular biometric measurements; primary angle closure disease; swept source OCT
Year: 2020 PMID: 32879770 PMCID: PMC7442878 DOI: 10.1167/tvst.9.9.14
Source DB: PubMed Journal: Transl Vis Sci Technol ISSN: 2164-2591 Impact factor: 3.283
Figure 1.Representative AS-OCT images from the CASIA SS-1000 (top) and ANTERION (bottom).
Intra-device Measurement Repeatability of the CASIA SS-1000.
| Parameter | CASIA Set 1 Mean | CASIA Set 2 Mean | Difference (% Mean) | P-Value | ICC | ICC 95% CI | LoA |
|---|---|---|---|---|---|---|---|
| PD (mm) | 4.74 ± 1.70 | 4.74 ± 1.68 | 0.04 | 0.97 | 0.99 | 0.99–1.00 | −0.16–0.16 |
| ACW (mm) | 11.95 ± 0.34 | 11.94 ± 0.30 | 0.09 | 0.60 | 0.93 | 0.86–0.97 | −0.23–0.25 |
| ACD (mm) | 2.88 ± 0.56 | 2.88 ± 0.56 | 0.03 | 0.39 | 0.97 | 0.94–0.99 | −0.26–0.26 |
| LV (mm) | 0.32 ± 0.50 | 0.31 ± 0.52 | 3.81 | 0.36 | 0.97 | 0.95–0.99 | −0.20–0.23 |
| AOD500 (mm) | 0.42 ± 0.26 | 0.41 ± 0.26 | 1.93 | 0.62 | 0.95 | 0.92–0.97 | −0.15–0.17 |
| TISA500 (mm2) | 0.14 ± 0.10 | 0.15 ± 0.09 | 4.83 | 0.06 | 0.94 | 0.92–0.97 | −0.06–0.05 |
| SSA500 (degrees) | 36.44 ± 16.89 | 35.88 ± 17.89 | 1.52 | 0.42 | 0.96 | 0.94–0.98 | −8.50–9.61 |
ICC = intraclass correlation coefficient, CI = confidence interval, LoA = limits of agreement.
PD = pupillary diameter, ACW = anterior chamber width, ACD = anterior chamber depth, LV = lens vault, AOD500 = angle opening distance at 500 µm, TISA500 = trabecular iris space area at 500 µm, SSA500 = scleral spur angle at 500 µm.
Values represented as mean ± standard deviation.
P-values calculated using Wilcoxon signed rank test.
Intra-device Measurement Repeatability of the ANTERION.
| Parameter | ANTERION Set 1 Mean | ANTERION Set 2 Mean | Difference (% Mean) | P-Value | ICC | ICC 95% CI | LoA |
|---|---|---|---|---|---|---|---|
| PD (mm) | 4.87 ± 1.72 | 4.85 ± 1.72 | 0.25 | 0.39 | 0.99 | 0.99–1.00 | −0.16–0.18 |
| ACW (mm) | 11.97 ± 0.33 | 11.97 ± 0.33 | 0.01 | 0.81 | 0.96 | 0.93–0.98 | −0.17–0.17 |
| ACD (mm) | 2.89 ± 0.70 | 2.89 ± 0.70 | 0.07 | 0.23 | 0.99 | 0.99–1.00 | −0.01–0.01 |
| LV (mm) | 0.23 ± 0.70 | 0.23 ± 0.70 | 0.43 | 0.83 | 0.99 | 0.92–0.98 | −0.08–0.08 |
| AOD500 (mm) | 0.45 ± 0.32 | 0.44 ± 0.31 | 1.12 | 0.75 | 0.97 | 0.95–0.98 | −0.13–0.14 |
| TISA500 (mm2) | 0.15 ± 0.11 | 0.15 ± 0.12 | 2.00 | 0.26 | 0.96 | 0.95–0.98 | −0.06–0.05 |
| SSA500 (degrees) | 37.59 ± 20.93 | 37.67 ± 21.50 | 0.66 | 0.69 | 0.97 | 0.96–0.98 | −9.71–9.21 |
ICC = intraclass correlation coefficient, CI = confidence interval, LoA = limits of agreement.
PD = pupillary diameter, ACW = anterior chamber width, ACD = anterior chamber depth, LV = lens vault, AOD500 = angle opening distance at 500 µm, TISA500 = trabecular iris space area at 500 µm, SSA500 = scleral spur angle at 500 µm.
Values represented as mean ± standard deviation.
P-values calculated using Wilcoxon signed rank test.
Inter-device Measurement Agreement Between the CASIA SS-1000 and ANTERION.
| Parameter | CASIA Mean | ANTERION Mean | Difference (% Mean) | P-Value | ICC | ICC 95% CI | LoA |
|---|---|---|---|---|---|---|---|
| PD (mm) | 4.75 ± 1.70 | 4.84 ± 1.75 | 1.77 | 0.15 | 0.98 | 0.98–0.99 | −0.55–0.38 |
| ACW (mm) | 11.95 ± 0.36 | 11.97 ± 0.34 | 0.12 | 0.84 | 0.9 | 0.82–0.95 | −0.31–0.28 |
| ACD (mm) | 2.83 ± 0.53 | 2.89 ± 0.70 | 2.06 | 0.06 | 0.89 | 0.79–0.95 | −0.62–0.50 |
| LV (mm) | 0.37 ± 0.48 | 0.39 ± 0.54 | 3.68 | 0.75 | 0.96 | 0.92–0.98 | −0.28–0.25 |
| AOD500 (mm) | 0.41 ± 0.26 | 0.45 ± 0.30 | 7.91 | 0.14 | 0.87 | 0.80–0.92 | −0.31–0.24 |
| TISA500 (mm2) | 0.14 ± 0.10 | 0.15 ± 0.11 | 5.52 | 0.27 | 0.85 | 0.78–0.91 | −0.12–0.10 |
| SSA500 (degrees) | 35.63 ± 17.47 | 38.35 ± 20.89 | 7.37 | 0.03 | 0.88 | 0.82–0.93 | −20.14–14.68 |
ICC = intraclass correlation coefficient, CI = confidence interval, LoA = limits of agreement.
PD = pupillary diameter, ACW = anterior chamber width, ACD = anterior chamber depth, LV = lens vault, AOD500 = angle opening distance at 500 µm, TISA500 = trabecular iris space area at 500 µm, SSA500 = scleral spur angle at 500 µm.
Values represented as mean ± standard deviation.
P-values calculated using Wilcoxon signed rank test.
Figure 2.Interdevice measurement agreement between the CASIA SS-1000 and ANTERION. Each subfigure includes linear regression (left) and Bland-Altman (right) plots. Dotted line in linear regression plot represents the equivalence line and solid line represents the regression line. Bland-Altman plot includes mean difference (solid line) and limits of agreement (dotted lines). Plots are shown for PD, ACW, LV, AOD500, TISA500, and SSA500.