Comparison of acquisition rate and agreement of axial length with two swept-source optical coherence tomographers and a partial coherence interferometer.

Shi-Ming Cheng1,2,3, Wen-Tao Yan1,2, Jia-Sheng Zhang1,2, Tian-Tian Li1,2, Xi Li1,2, A-Yong Yu4,5.   

Abstract

PURPOSE: To assess the acquisition rate and agreement of axial length (AL) measurements with the OA-2000, Anterion, and IOLMaster 500 in cataractous patients.
METHODS: In total, 298 eyes of 191 cataractous patients were enrolled and scanned with the three devices in random order. The success rate of AL measurements per device was calculated and a chi-square test was utilized to identify the differences in acquisition rate between the three devices. Logistic regression analysis was applied to evaluate the association of different cataract types and severity with the AL measurement acquisition rate. Bland-Altman plots were mapped to appraise the agreement of AL values.
RESULTS: AL measurements were successfully achieved in 288 eyes (96.64%) with the OA-2000, in 282 eyes (94.30%) with the Anteiron, and in 246 eyes (82.55%) with the IOLMaster 500. Significant differences in the acquisition rate were found between either of the SS-OCT devices and IOLMaster 500 by chi-square analysis (P < 0.001). No significant difference was noted between OA-2000 and Anterion. Increasing severity of posterior subcapsular cataract was associated with a higher failure rate with the IOLMaster 500. Bland-Altman analysis identified good agreement between the three biometers with narrow 95% limits of agreement.
CONCLUSIONS: The OA-2000 and Anterion showed similarly higher acquisition rate of AL measurements than IOLMaster 500 in cataractous patients. Good agreement for AL values was found between the three biometers in cataractous patients.
© 2022. The Author(s), under exclusive licence to Springer-Verlag GmbH Germany, part of Springer Nature.

Entities:  

Keywords:  Acquisition rate; Agreement; Axial length; Optical coherence tomographer; Swept-source

Mesh:

Year:  2022        PMID: 35488908     DOI: 10.1007/s00417-022-05681-y

Source DB:  PubMed          Journal:  Graefes Arch Clin Exp Ophthalmol        ISSN: 0721-832X            Impact factor:   3.535


  28 in total

1.  The repeatability and accuracy of axial length and anterior chamber depth measurements from the IOLMaster .

Authors:  A K Lam; R Chan; P C Pang
Journal:  Ophthalmic Physiol Opt       Date:  2001-11       Impact factor: 3.117

2.  Comparison of the Zeiss IOLMaster and applanation A-scan ultrasound: biometry for intraocular lens calculation.

Authors:  Loreto T Rose; Con N Moshegov
Journal:  Clin Exp Ophthalmol       Date:  2003-04       Impact factor: 4.207

3.  Precision of biometry, keratometry, and refractive measurements with a partial coherence interferometry-keratometry device.

Authors:  H John Shammas; Steven Chan
Journal:  J Cataract Refract Surg       Date:  2010-09       Impact factor: 3.351

4.  Sources of error in intraocular lens power calculation.

Authors:  T Olsen
Journal:  J Cataract Refract Surg       Date:  1992-03       Impact factor: 3.351

5.  Sources of error in intraocular lens power calculation.

Authors:  Sverker Norrby
Journal:  J Cataract Refract Surg       Date:  2008-03       Impact factor: 3.351

6.  Comparison of predictions made by the intraocular lens master and ultrasound biometry.

Authors:  Anand B Bhatt; Amy C Schefler; William J Feuer; Sonia H Yoo; Timothy G Murray
Journal:  Arch Ophthalmol       Date:  2008-07

7.  Reproducibility of optical biometry using partial coherence interferometry : intraobserver and interobserver reliability.

Authors:  A Vogel; H B Dick; F Krummenauer
Journal:  J Cataract Refract Surg       Date:  2001-12       Impact factor: 3.351

8.  Intraocular lens exchange due to incorrect lens power.

Authors:  George J C Jin; Alan S Crandall; Jason J Jones
Journal:  Ophthalmology       Date:  2006-11-21       Impact factor: 12.079

9.  Optical and ultrasound measurement of axial length and anterior chamber depth for intraocular lens power calculation.

Authors:  János Németh; Orsolya Fekete; Norbert Pesztenlehrer
Journal:  J Cataract Refract Surg       Date:  2003-01       Impact factor: 3.351

10.  Evaluation of a new IOLMaster algorithm to measure axial length.

Authors:  Warren Hill; Raymund Angeles; Todd Otani
Journal:  J Cataract Refract Surg       Date:  2008-06       Impact factor: 3.351

View more
  1 in total

1.  Risk factors for development of superior limbic keratoconjunctivitis in thyroid eye disease in Japanese.

Authors:  Yasuhiro Takahashi; Aric Vaidya; Hirohiko Kakizaki
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2022-09-09       Impact factor: 3.535

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.