Ahmet Akman1, Leyla Asena1, Sirel Gür Güngör1.
Abstract
PURPOSE: To compare the measurements and failure rates obtained with a new swept source optical coherence tomography (OCT)-based biometry to IOLMaster 500.Entities:
Keywords: Diagnostic tests/Investigation; Optics and Refraction
Mesh:
Year: 2015 PMID: 26674777 PMCID: PMC5013111 DOI: 10.1136/bjophthalmol-2015-307779
Source DB: PubMed Journal: Br J Ophthalmol ISSN: 0007-1161 Impact factor: 4.638
The comparison of values between IOLMaster 500 and IOLMaster 700
| n | Device | Mean | SD | Paired samples t test (p value) | Range (minimum–maximum) | ICC | Differences of the means | SD of difference of means | 95% CI of differences of means | Paired samples t test p value | |
|---|---|---|---|---|---|---|---|---|---|---|---|
| Anterior Chamber depth (mm) | 175 | IOLmaster 500 | 3.31 | 0.38 | <0.001 | 2.23–4.19 | 0.920 | −0.08 | 0.13 | −0.102 to −0.063 | <0.001 |
| IOLMaster 700 | 3.23 | 0.37 | 2.33–4.14 | ||||||||
| Axial length (mm) | 171 | IOLMaster 500 | 23.627 | 1.25 | 0.008 | 20.92–28.33 | 1.0 | −0.005 | 0.02 | −0.008 to −0.002 | 0.830 |
| IOLMaster 700 | 23.622 | 1.32 | 20.91–29.69 | ||||||||
| Keratometry (flat, D) | 188 | IOLMaster 500 | 43.46 | 1.77 | 0.001 | 48.35–37.17 | 0.992 | −0.05 | 0.22 | −0.079 to −0.014 | <0.001 |
| IOLMaster 700 | 43.41 | 1.73 | 37.25–48.21 | ||||||||
| Keratometry (steep, D) | 188 | IOLMaster 500 | 44.50 | 1.81 | <0.001 | 49.13–38.14 | 0.989 | −0.08 | 0.27 | −0.114 to −0.036 | <0.001 |
| IOLMaster 700 | 44.42 | 1.83 | 38.05–49.07 |
Difference of the means was computed by subtracting IOLMaster 500 values form IOLMaster 700 values.
ICC, intraclass correlation coefficient.
Figure 1Bland–Altman plot for anterior chamber depth (ACD) comparing IOLMaster 700 with IOLMaster 500. The 95% limits of agreement were −0.33 to 0.17 mm. Dotted lines, ±1.96 SD.
Figure 2Bland–Altman plot for axial length (AL) comparing IOLMaster 700 with IOLMaster 500. The 95% limits of agreement were −0.045 to 0.035 mm. Dotted lines, ±1.96 SD.
Figure 3Bland–Altman plot for K1 value comparing IOLMaster 700 with IOLMaster 500. The 95% limits of agreement were −0.48 to 0.38 D. Dotted lines, ±1.96 SD.
Figure 4Bland–Altman plot for K2 value comparing IOLMaster 700 with IOLMaster 500. The 95% limits of agreement were −0.61 to 0.45 D. Dotted lines, ±1.96 SD.
The differences in measurements between IOLMaster 500 and IOLMaster 700; n (%)
| Keratometry measurements (n=188) | |||||||
|---|---|---|---|---|---|---|---|
| Anterior chamber depth (n=175) | Axial length (n=188) | K1 | K2 | ||||
| Depth (mm) | n (%) | Length (mm) | n (%) | Keratometry measurements (D) | n (%) | Keratometry measurements (D) | n (%) |
| <0.1 mm | 90 (52) | No difference | 43 (24) | <0.5D | 181 (96) | <0.5D | 173 (92) |
| <0.2 mm | 154 (88) | <0.02 mm | 102 (56) | <1D | 188 (100) | <1D | 188 (100) |
| <0.3 mm | 175 (100) | <0.03 mm | 150 (82) | ||||
| <0.05 mm | 165 (91) | ||||||
| <0.10 mm | 177 (96) | ||||||
| <0.15 mm | 188 (100) | ||||||