Axial length measurements: Comparison of a new swept-source optical coherence tomography-based biometer and partial coherence interferometry in myopia.

Jong Yun Yang1, Hong Kyu Kim1, Sung Soo Kim2.   

Abstract

PURPOSE: To compare axial length (AL) measurements between a swept-source optical coherence tomography (SS-OCT) biometer (IOLMaster 700) and a partial coherence interferometry (PCI) biometer (IOLMaster, version 5.4) in myopic eyes.
SETTING: Severance Hospital, Yonsei University, Seoul, South Korea.
DESIGN: Prospective evaluation of a diagnostic test.
METHODS: This study enrolled patients with a diagnosis of myopia in their medical records. Two consecutive AL measurements were performed in each eye in random order using the 2 biometers. Subanalysis was performed according to lens status, fixation status, degree of myopia, and the presence of posterior staphyloma. The interdevice agreement was evaluated with Bland-Altman analyses and paired t tests.
RESULTS: This study comprised 219 eyes of 117 patients. During a fixation check using SS-OCT, 19.6% of eyes showed fixation loss. Overall, the SS-OCT biometer showed a longer AL than the PCI biometer. (P < .001) In a subanalysis that categorized eyes according to lens status, fixation status, degree of myopia, and the presence of posterior staphyloma, the SS-OCT biometer also showed longer ALs than the PCI biometer. Of eyes with good fixation, significant differences in AL measurements between devices were detected for those with posterior staphyloma (P < .001) but not for those without posterior staphyloma (P = .104).
CONCLUSIONS: The 2 devices showed differences in AL measurements in myopic eyes. Fixation status and the presence of posterior staphyloma were important factors in these differences. For myopic eyes with posterior staphyloma, the SS-OCT biometer is expected to produce more precise AL measurements because it allows evaluation of the fixation status.
Copyright © 2017 ASCRS and ESCRS. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2017        PMID: 28410713     DOI: 10.1016/j.jcrs.2016.12.023

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  10 in total

1.  Comparison of IOL-Master 700 and IOL-Master 500 biometers in ocular biological parameters of adolescents.

Authors:  Qiang Shi; Guang-Yan Wang; Yu-Hong Cheng; Cheng Pei
Journal:  Int J Ophthalmol       Date:  2021-07-18       Impact factor: 1.779

2.  Refractive error characteristics and influence on ocular parameters in patients with unilateral congenital ptosis.

Authors:  Yingli Liu; Tingting Chen; Jingwen Huang; Wentao Li; Yilin Chen; Lijun Huo
Journal:  BMC Ophthalmol       Date:  2022-07-02       Impact factor: 2.086

3.  Axial Length Change in Pseudophakic Eyes Measured by IOLMaster 700.

Authors:  Jiaqing Zhang; Zhenzhen Liu; Xiaozhang Qiu; Ling Jin; Lanhua Wang; Guangming Jin; Wei Wang; Xuhua Tan; Lixia Luo; Yizhi Liu
Journal:  Transl Vis Sci Technol       Date:  2021-05-03       Impact factor: 3.283

4.  Changes in the anterior segment after cycloplegia with a biometer using swept-source optical coherence tomography.

Authors:  Tomoaki Higashiyama; Maki Iwasa; Masahito Ohji
Journal:  PLoS One       Date:  2017-08-14       Impact factor: 3.240

5.  Comparison of a new biometer using swept-source optical coherence tomography and a conventional biometer using partial coherence interferometry.

Authors:  Tomoaki Higashiyama; Hazuki Mori; Fumi Nakajima; Masahito Ohji
Journal:  PLoS One       Date:  2018-04-24       Impact factor: 3.240

6.  Predictive accuracy of partial coherence interferometry and swept-source optical coherence tomography for intraocular lens power calculation.

Authors:  Woong-Joo Whang; Young-Sik Yoo; Min-Ji Kang; Choun-Ki Joo
Journal:  Sci Rep       Date:  2018-09-13       Impact factor: 4.379

7.  Comparison of Ocular Biometry Using New Swept-source Optical Coherence Tomography-based Optical Biometer with Other Devices.

Authors:  Young Joo Cho; Tae Hyung Lim; Kee Yong Choi; Beom Jin Cho
Journal:  Korean J Ophthalmol       Date:  2018-08

8.  Ocular biometry and refractive outcomes using two swept-source optical coherence tomography-based biometers with segmental or equivalent refractive indices.

Authors:  Miki Kamikawatoko Omoto; Hidemasa Torii; Sachiko Masui; Masahiko Ayaki; Kazuo Tsubota; Kazuno Negishi
Journal:  Sci Rep       Date:  2019-04-25       Impact factor: 4.379

9.  Comparison of a New Optical Biometer That Combines Scheimpflug Imaging With Partial Coherence Interferometry With That of an Optical Biometer Based on Swept-Source Optical Coherence Tomography and Placido-Disk Topography.

Authors:  Shihao Chen; Qiaoyue Zhang; Giacomo Savini; Shuangzhe Zhang; Xiaomin Huang; Jinjin Yu; Yirang Wang; Rui Ning; Jinhai Huang; Ruixue Tu
Journal:  Front Med (Lausanne)       Date:  2022-02-10

10.  Comparison of two novel swept-source optical coherence tomography devices to a partial coherence interferometry-based biometer.

Authors:  Tommy C Y Chan; Marco C Y Yu; Vivian Chiu; Gilda Lai; Christopher K S Leung; Poemen P M Chan
Journal:  Sci Rep       Date:  2021-07-21       Impact factor: 4.379

  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.