Agreement Between Swept-Source Optical Biometry and Scheimpflug-based Topography Measurements of Anterior Segment Parameters.

Pelin Özyol1, Erhan Özyol2.   

Abstract

PURPOSE: To estimate the agreement of anterior segment parameters between a swept-source optical biometry (IOLMaster 700; Carl Zeiss Meditec AG, Jena, Germany) and a Scheimpflug-based topography with high resolution (Pentacam HR; OCULUS, Wetzlar, Germany).
DESIGN: Interinstrument reliability analysis.
METHODS: A total of 62 eyes from 62 young adults were included in the study. Average keratometry (AveK) and simulated keratometry (SimK) along 2.0-mm-ring measurements provided by Pentacam HR, keratometry readings provided by IOLMaster 700, and central corneal thickness (CCT) and anterior chamber depth (ACD) values obtained from both devices were recorded. J0 and J45 vectoral components of astigmatism were obtained using power vector analysis. Mean keratometry (Km) values of IOLMaster 700 were compared for each type of Km value from Pentacam HR, while other parameters were compared between devices. To assess the agreement between measurements of the devices, Bland-Altman analysis was performed.
RESULTS: The Pentacam HR exhibited significantly lower Km and CCT measurements (P < .001, for all); however, no significant difference emerged in J0, J45, and ACD measurements (P = .057, P = .574, and P = .64, respectively). The mean difference between AveK, SimK 2.0 mm, and the IOLMaster 700 Km was -0.20 diopter (D) and -0.14 D, respectively, while the mean difference between J0, J45, CCT, and ACD measurements was 0.07 D, -0.016 D, -5.05 μm, and 0.004 mm, respectively.
CONCLUSION: In clinical practice, Pentacam HR and IOLMaster 700 can be used interchangeably to measure J0 and J45 vectoral components of astigmatism for SimK 2.0 mm and IOLMaster keratometry values, as well as ACD and CCT measurements. However, SimK 2.0 mm and AveK values can be not interchangeable, as the devices have clinical and statistical differences in measurements.
Copyright © 2016 Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2016        PMID: 27320057     DOI: 10.1016/j.ajo.2016.06.020

Source DB:  PubMed          Journal:  Am J Ophthalmol        ISSN: 0002-9394            Impact factor:   5.258


  16 in total

1.  Association between axial length and in vivo human crystalline lens biometry during accommodation: a swept-source optical coherence tomography study.

Authors:  Takuhei Shoji; Naoko Kato; Sho Ishikawa; Hisashi Ibuki; Norihiro Yamada; Itaru Kimura; Kei Shinoda
Journal:  Jpn J Ophthalmol       Date:  2019-11-23       Impact factor: 2.447

2.  Effects of corneal thickness distribution and apex position on postoperative refractive status after full-bed deep anterior lamellar keratoplasty.

Authors:  Bing-Hong Wang; Ye-Sheng Xu; Wen-Jia Xie; Yu-Feng Yao
Journal:  J Zhejiang Univ Sci B       Date:  2018 Nov.       Impact factor: 3.066

3.  Repeatability of total Keratometry and standard Keratometry by the IOLMaster 700 and comparison to total corneal astigmatism by Scheimpflug imaging.

Authors:  Giacomo Savini; Leonardo Taroni; Domenico Schiano-Lomoriello; Kenneth J Hoffer
Journal:  Eye (Lond)       Date:  2020-11-02       Impact factor: 4.456

4.  Corneal power evaluation after myopic corneal refractive surgery using artificial neural networks.

Authors:  Robert Koprowski; Michele Lanza; Carlo Irregolare
Journal:  Biomed Eng Online       Date:  2016-11-15       Impact factor: 2.819

5.  In vivo crystalline lens measurements with novel swept-source optical coherent tomography: an investigation on variability of measurement.

Authors:  Takuhei Shoji; Naoko Kato; Sho Ishikawa; Hisashi Ibuki; Norihiro Yamada; Itaru Kimura; Kei Shinoda
Journal:  BMJ Open Ophthalmol       Date:  2017-03-03

6.  Comparison of central corneal thickness with four noncontact devices: An agreement analysis of swept-source technology.

Authors:  Erhan Ozyol; Pelin Özyol
Journal:  Indian J Ophthalmol       Date:  2017-06       Impact factor: 1.848

7.  Evaluation and comparison of a novel Scheimpflug-based optical biometer with standard partial coherence interferometry for biometry and intraocular lens power calculation.

Authors:  Ziyang Wang; Wenli Yang; Dongjun Li; Wei Chen; Qi Zhao; Yifeng Li; Rui Cui; Lin Shen; Junfang Xian
Journal:  Exp Ther Med       Date:  2021-02-05       Impact factor: 2.447

8.  Evaluation of preoperative corneal astigmatism using swept-source optical biometry in Chinese cataract surgery candidates with high myopia: a prospective, comparative observational study.

Authors:  Yehui Tan; Liangping Liu; Jianbing Li; Yingyan Qin; Ao Sun; Mingxing Wu
Journal:  Ann Transl Med       Date:  2021-04

9.  Comparison of two novel swept-source optical coherence tomography devices to a partial coherence interferometry-based biometer.

Authors:  Tommy C Y Chan; Marco C Y Yu; Vivian Chiu; Gilda Lai; Christopher K S Leung; Poemen P M Chan
Journal:  Sci Rep       Date:  2021-07-21       Impact factor: 4.379

10.  Accuracy of OCT-derived net corneal astigmatism measurement.

Authors:  Clara Llorens-Quintana; Elias Pavlatos; Omkar Thaware; Seema Gupta; Daniel Gradin; Denzil Romfh; Yan Li; David Huang
Journal:  J Cataract Refract Surg       Date:  2022-03-01       Impact factor: 3.528

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.