Repeatability and agreement in optical biometry of a new swept-source optical coherence tomography-based biometer versus partial coherence interferometry and optical low-coherence reflectometry.

Kathleen S Kunert1, Monika Peter2, Marcus Blum2, Wolfgang Haigis2, Walter Sekundo2, Juliane Schütze2, Tobias Büehren2.   

Abstract

PURPOSE: To estimate the repeatability of biometric parameters obtained with a new swept-source biometer and to compare the agreement with that of partial coherence interferometry (PCI) and optical low-coherence reflectometry (OLCR).
SETTING: Department of Ophthalmology, Helios Hospital Erfurt, Erfurt, Julius-Maximilians University, Würzburg, and Philipps University, Marburg, Germany.
DESIGN: Prospective comparative multicenter clinical study.
METHODS: Biometry was taken with the use of 3 different biometers: the IOLMaster 700 swept-source biometer, the PCI-based IOLMaster 500, and the OCLR-based Lenstar LS 900. Axial length (AL), anterior chamber depth (ACD), and spherical equivalent (SE) were compared between swept-source and PCI biometry and central corneal thickness (CCT) and lens thickness (LT) between swept-source and OLCR biometry. The repeatability of swept-source biometry was evaluated on the basis of 3 measurements captured for each patient.
RESULTS: One hundred twenty cataract eyes were included in the study. The mean difference between swept-source and PCI biometry for AL, ACD, and SE measurements was 4 μm ± 25 (SD), 17 ± 122 μm, and -0.001 ± 0.19 diopter (D), respectively. The mean difference between swept-source and OLCR biometry for LT and CCT measurements was 21 ± 122 μm and 0.15 ± 4.51 μm, respectively. Differences between swept-source biometry and the other devices distributed around zero without statistical significance. The standard deviation of repeatability for AL, ACD, LT, CCT, and SE was 8.8 μm, 9.8 μm, 2.3 μm, 19.5 μm, and 0.1 D, respectively.
CONCLUSIONS: Swept-source biometry showed high repeatability performance for all biometric parameters. The agreement of AL, ACD, and SE between swept-source and PCI biometry as well as that of LT and CCT between swept-source and OLCR biometry was excellent. It remains to be validated whether high repeatability shown by swept-source biometry will result in better postoperative refractive outcomes. FINANCIAL DISCLOSURE: Drs. Blum and Sekundo are members of the Scientific Advisory Board of Carl Zeiss Meditec AG. Drs. Peter and Bühren are employees of Carl Zeiss Meditec AG.
Copyright © 2016 The Authors. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2016        PMID: 26948781     DOI: 10.1016/j.jcrs.2015.07.039

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  35 in total

1.  Mechanism of accommodation assessed by change in precisely registered ocular images associated with concurrent change in auto-refraction.

Authors:  Andrzej Grzybowski; Ronald A Schachar; Magdalena Gaca-Wysocka; Ira H Schachar; Farhad Kamangar; Barbara K Pierscionek
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2017-11-16       Impact factor: 3.117

2.  Ocular anatomic changes for different accommodative demands using swept-source optical coherence tomography: a pilot study.

Authors:  José J Esteve-Taboada; Teresa Ferrer-Blasco; Miguel A Aloy; José E Adsuara; Pablo Cerdá-Durán; Petar Mimica; Robert Montés-Micó
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2017-09-26       Impact factor: 3.117

3.  Evaluation of the repeatability of a swept-source ocular biometer for measuring ocular biometric parameters.

Authors:  Teresa Ferrer-Blasco; Alberto Domínguez-Vicent; José J Esteve-Taboada; Miguel A Aloy; Jose E Adsuara; Robert Montés-Micó
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2016-11-30       Impact factor: 3.117

4.  Agreement and clinical comparison between a new swept-source optical coherence tomography-based optical biometer and an optical low-coherence reflectometry biometer.

Authors:  P Arriola-Villalobos; J Almendral-Gómez; N Garzón; J Ruiz-Medrano; C Fernández-Pérez; J M Martínez-de-la-Casa; D Díaz-Valle
Journal:  Eye (Lond)       Date:  2016-11-11       Impact factor: 3.775

5.  Evaluating relaxed ciliary muscle tone in presbyopic eyes.

Authors:  Erhan Özyol; Pelin Özyol
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2017-02-25       Impact factor: 3.117

6.  Algorithmic intraocular lens power calculation formula selection by keratometry, anterior chamber depth and axial length.

Authors:  Jung Wan Kim; Youngsub Eom; Eun Gyu Yoon; Young Choi; Jong Suk Song; Ji Won Jeong; Seh Kwang Park; Hyo Myung Kim
Journal:  Acta Ophthalmol       Date:  2021-08-11       Impact factor: 3.988

7.  Changes in ocular morphology after cataract surgery in open angle glaucoma patients.

Authors:  Giedre Pakuliene; Loreta Kuzmiene; Brent Siesky; Alon Harris; Ingrida Januleviciene
Journal:  Sci Rep       Date:  2021-06-09       Impact factor: 4.379

8.  Determination of Personalized IOL-Constants for the Haigis Formula under Consideration of Measurement Precision.

Authors:  Simon Schröder; Christina Leydolt; Rupert Menapace; Timo Eppig; Achim Langenbucher
Journal:  PLoS One       Date:  2016-07-08       Impact factor: 3.240

9.  Comparison of two novel swept-source optical coherence tomography devices to a partial coherence interferometry-based biometer.

Authors:  Tommy C Y Chan; Marco C Y Yu; Vivian Chiu; Gilda Lai; Christopher K S Leung; Poemen P M Chan
Journal:  Sci Rep       Date:  2021-07-21       Impact factor: 4.379

10.  Comparison of Repeatability and Agreement between Swept-Source Optical Biometry and Dual-Scheimpflug Topography.

Authors:  Soyeon Jung; Hee Seung Chin; Na Rae Kim; Kang Won Lee; Ji Won Jung
Journal:  J Ophthalmol       Date:  2017-12-10       Impact factor: 1.909

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.