Poemen Pui-Man Chan1,2, Gilda Lai1, Vivian Chiu1, Anita Chong1, Marco Yu1,3, Christopher Kai-Shun Leung4,5.
Abstract
This study compared the test-retest variabilities and measurement agreement of anterior chamber angle (ACA) dimensions measured by two anterior segment swept-source optical coherence tomography (SS-OCT)-the ANTERION (Heidelberg Engineering, Heidelberg, Germany) and CASIAII (Tomey, Nagoya, Japan). Thirty-eight subjects, 18 patients with primary angle closure and 20 healthy participants with open angles, were included. The mean age was 54.7 ± 15.8 years (range: 26-75 years). One eye of each subject was randomly selected for anterior segment imaging by ANTERION and CASIAII, using the same scan pattern (6 evenly spaced radial scans across the anterior segment for three times) in the same visit. The between- and within-instrument agreement and repeatability coefficients of angle open distance (AOD500), trabecular-iris space area (TISA500), lens vault (LV), scleral spur-scleral spur distance (SSD), anterior chamber depth (ACD), and pupil diameter (PD) were measured. The anterior and posterior boundaries of the cornea, iris, and lens were automatically segmented by the SS-OCT instruments; the scleral spur was manually located by a single masked observer. There were significant differences between ANTERION and CASIAII measurements; the SSD, PD, and ACD were smaller whereas AOD500 and TISA500 were greater in ANTERION compared with CASIAII (P < 0.001). Anterior segment measurements obtained from the two SS-OCT instruments showed strong associations (R2 ranged between 0.866 and 0.998) although the between-instrument agreement was poor; the spans of 95% limits of between-instrument agreement were ≥ 1.5-folds than the within-instrument agreement for either instrument. Whereas both SS-OCT instruments showed low test-retest measurement variabilities, the repeatability coefficients of AOD500, TISA500, ACD, and PD were slightly smaller for CASIAII than ANTERION (P ≤ 0.012).Entities:
Mesh:
Year: 2020 PMID: 33127944 PMCID: PMC7603499 DOI: 10.1038/s41598-020-74813-3
Source DB: PubMed Journal: Sci Rep ISSN: 2045-2322 Impact factor: 4.379
Specifications and scan protocols of anterior chamber angle imaging for ANTERION and CASIAII.
| ANTERION (Metrics App) | CASIAII | |
|---|---|---|
| Light source wavelength (nm) | 1300 | 1310 |
| Axial resolution (µm) | < 10 | < 10 |
| Transverse resolution (µm) | < 30 | < 30 |
| Scan speed (A-scans per second) | 50,000 | 50,000 |
| Scan depth (mm) | 14 ± 0.5 | 13 |
| Maximum scan width (mm) | 16.5 | 16 |
| Scan pattern | 6 evenly-spaced radial B-scans | 6 evenly-spaced radial B-scansa |
| Number of A-scans per B-scan | 768 | 800 |
aCustom-designed for the study.
Demographics and anterior segment optical coherence tomography measurements (mean ± SD) of the normal and primary angle closure disease (PACD) groups.
| Normal group | PACD group | P* | |
|---|---|---|---|
| Age (years) | 46.35 ± 17.46 | 64.00 ± 5.52 | < 0.001 |
| Intraocular pressure (mmHg) | 14.95 ± 2.80 | 16.26 ± 4.30 | 0.256 |
| Axial Length (mm) | 24.93 ± 1.12 | 23.10 ± 0.78 | < 0.001 |
| Dark-room gonioscopy grading (Shaffer) | 3.55 ± 0.69 | 0.64 ± 0.84 | < 0.001 |
| Anterior chamber depth (mm) | 2.91 ± 0.44 | 2.00 ± 0.33 | < 0.001 |
| Pupil diameter (mm) | 5.60 ± 1.10 | 4.59 ± 0.93 | 0.003 |
| Scleral spur-scleral spur distance (mm) | 11.79 ± 0.34 | 11.58 ± 0.28 | 0.004 |
| Lens vault (mm) | 0.176 ± 0.332 | 0.979 ± 0.282 | < 0.001 |
| Angle opening distance 500 (mm) | 0.452 ± 0.227 | 0.109 ± 0.086 | < 0.001 |
| Trabecular iris space area 500 (mm2) | 0.149 ± 0.072 | 0.044 ± 0.025 | < 0.001 |
| Anterior chamber depth (mm) | 2.95 ± 0.44 | 2.04 ± 0.33 | < 0.001 |
| Pupil diameter (mm) | 5.73 ± 1.08 | 4.77 ± 1.10 | 0.010 |
| Scleral spur-scleral spur distance (mm) | 11.82 ± 0.33 | 11.60 ± 0.27 | 0.034 |
| Lens vault (mm) | 0.178 ± 0.329 | 0.986 ± 0.281 | < 0.001 |
| Angle opening distance 500 (mm) | 0.426 ± 0.220 | 0.093 ± 0.087 | < 0.001 |
| Trabecular iris space area 500 (mm2) | 0.145 ± 0.070 | 0.034 ± 0.030 | < 0.001 |
*Independent t-test. OCT measurements were derived from the mean of 6 evenly-spaced radial scan measurements.
Comparisons of anterior segment parameters (mean ± SD) between ANTERION and CASIAII (n = 38 eyes; 20 healthy eyes with open-angles and 18 eyes with PACD).
| Clock-hours | Anterior chamber depth, ACD (mm) | Pupil diameter, PD (mm) | ||||
|---|---|---|---|---|---|---|
| ANTERION | CASIAII | ANTERION | CASIAII | |||
| Mean | 2.48 ± 0.60 | 2.51 ± 0.60 | < 0.001 | 5.11 ± 1.12 | 5.36 ± 1.17 | < 0.001 |
*Bonferroni correction for multiple comparisons was applied. The P value considered to be significant was 0.0083 for SSD, and 0.0042 for AOD500/TISA500.
Figure 1Bland–Altman plot: comparison of different anterior segment parameters. (A) Between-instrument agreement (ANTERION vs. CASIAII). (B) Within-instrument agreement (ANTERION). (C) Within-instrument agreement (CASIAII).
Figure 2Regression analysis of the anterior segment parameters measured from ANTERION and CASIAII. ACD, anterior chamber depth; PD, pupil diameter; SSD, scleral spur-scleral spur distance; LV, lens vault; angle open distance 500 (AOD500); trabecular-iris space area 500 (TISA500).
Figure 3An eye with open angle (A) and an eye with primary angle closure (B) imaged by the ANTERION and CASIAII. The scleral spur locations were manually marked (middle panel); the anterior and posterior boundaries of the cornea and the lens were automatically segmented by the respective instruments (right panel). ACD, anterior chamber depth; PD, pupil diameter; SSD, scleral spurs distance; LV, lens vault; AOD, angle opening distance at 500 µm; TISA, trabecular iris space area at 500 µm. Only the horizontal scan (0°–180°) are shown.
Comparison of repeatability coefficients of anterior segment parameters measured by ANTERION and CASIA II.
| ANTERION | CASIA II | P | |
|---|---|---|---|
| Repeatability coefficient (95% confidence interval) | Repeatability coefficient (95% confidence interval) | ||
| Anterior chamber depth (mm) | 0.017 (0.014 to 0.022) | 0.016 (0.013 to 0.020) | < 0.001 |
| Pupil diameter (mm) | 0.398 (0.326 to 0.513) | 0.371 (0.304 to 0.479) | 0.012 |
| Scleral spur-scleral spur distance (mm) | 0.109 (0.089 to 0.140) | 0.094 (0.077 to 0.121) | 0.067 |
| Lens vault (mm) | 0.050 (0.041 to 0.065) | 0.061 (0.050 to 0.079) | 0.471 |
| Angle opening distance 500 (mm) | 0.042 (0.035 to 0.054) | 0.035 (0.028 to 0.045) | < 0.001 |
| Trabecular iris space area 500 (mm2) | 0.016 (0.013 to 0.020) | 0.014 (0.012 to 0.018) | 0.001 |