Accuracy and reliability of IOL master and A-scan immersion biometry in silicone oil-filled eyes.

P Kunavisarut1, P Poopattanakul, C Intarated, K Pathanapitoon.   

Abstract

PURPOSE: To compare the accuracy and reliability of intraocular lens (IOL) master and A-scan immersion biometry in silicone oil (SO)-filled eyes.
METHODS: A prospective, consecutive, nonrandomized study was performed in 34 SO-filled eyes of 34 patients, who underwent a pars plana vitrectomy, with SO removal and cataract surgery, as well as IOL implantation. Both IOL master and immersion A-scan were performed to measure the axial length (AXL) before SO removal. Three months after removal of the SO, AXL measurements using IOL master and refraction was performed. Accuracy of the two techniques was determined by a mean postoperative AXL using an IOL master and reliability was determined by mean actual postoperative refractive error.
RESULTS: Preoperative mean AXL was 23.91±0.24 mm (range 21.33-28.61 mm) and 23.71±0.59 mm (range 19.27-36.18 mm) by IOL master and A-scan immersion, respectively. Postoperative mean AXL by IOL master was 23.90±0.23 mm (range 21.58-27.94 mm), which showed a statistically significant difference from the preoperative mean AXL by A-scan immersion (P=0.005). The AXL measurement by IOL master also was more accurate than A-scan immersion by Pearson's correlation (0.966 vs 0.410). For reliability of the two techniques, the predictive postoperative refractive error in A-scan immersion (mean 1.79±1.04 D, range -14.62 to 16.41 D) was greater than that in IOL master (mean 0.60±0.23 D, range -2.74 to 2.33 D), with a statistically significant difference (P=0.049).
CONCLUSION: IOL master had more accuracy and less deviation in predictive postoperative refractive error than A-scan immersion in SO-filled eyes.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22878446      PMCID: PMC3470048          DOI: 10.1038/eye.2012.163

Source DB:  PubMed          Journal:  Eye (Lond)        ISSN: 0950-222X            Impact factor:   3.775


  24 in total

1.  Evaluation of the practicality of optical biometry and applanation ultrasound in 253 eyes.

Authors:  Mana Tehrani; Frank Krummenauer; Eric Blom; H Burkhard Dick
Journal:  J Cataract Refract Surg       Date:  2003-04       Impact factor: 3.351

2.  Refractive changes after scleral buckling surgery.

Authors:  W E Smiddy; D N Loupe; R G Michels; C Enger; B M Glaser; S deBustros
Journal:  Arch Ophthalmol       Date:  1989-10

3.  Ultrasonic examination of the silicone-filled eye: theoretical and practical considerations.

Authors:  J K Shugar; E de Juan; B W McCuen; J Tiedeman; M R Landers; R Machemer
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  1986       Impact factor: 3.117

4.  Ultrasound velocities for axial eye length measurement.

Authors:  K J Hoffer
Journal:  J Cataract Refract Surg       Date:  1994-09       Impact factor: 3.351

5.  Change in postoperative refractive error when vitrectomy is added to intraocular lens implantation.

Authors:  M Shioya; N Ogino; U Shinjo
Journal:  J Cataract Refract Surg       Date:  1997-10       Impact factor: 3.351

6.  Biometry of the silicone oil-filled eye: II.

Authors:  D C Murray; O M Durrani; P Good; M T Benson; G R Kirkby
Journal:  Eye (Lond)       Date:  2002-11       Impact factor: 3.775

7.  Intraoperative biometry for intraocular lens (IOL) power calculation at silicone oil removal.

Authors:  S M el-Baha; A el-Samadoni; H F Idris; K M Rashad
Journal:  Eur J Ophthalmol       Date:  2003 Aug-Sep       Impact factor: 2.597

8.  Predicted and actual refraction after intraocular lens implantation in eyes with silicone oil.

Authors:  A Grinbaum; G Treister; J Moisseiev
Journal:  J Cataract Refract Surg       Date:  1996 Jul-Aug       Impact factor: 3.351

9.  Partial coherence interferometry: a novel approach to biometry in cataract surgery.

Authors:  W Drexler; O Findl; R Menapace; G Rainer; C Vass; C K Hitzenberger; A F Fercher
Journal:  Am J Ophthalmol       Date:  1998-10       Impact factor: 5.258

10.  Partial coherence laser interferometry vs conventional ultrasound biometry in intraocular lens power calculations.

Authors:  M S Rajan; I Keilhorn; J A Bell
Journal:  Eye (Lond)       Date:  2002-09       Impact factor: 3.775

View more
  16 in total

1.  Accurate biometry in silicone oil-filled eyes.

Authors:  R J Symes
Journal:  Eye (Lond)       Date:  2013-04-12       Impact factor: 3.775

2.  Reply to Richard Symes.

Authors:  P Kunavisarut
Journal:  Eye (Lond)       Date:  2013-04-12       Impact factor: 3.775

3.  Changes in axial length after vitrectomy for rhegmatogenous retinal detachment combined with choroidal detachment.

Authors:  Ji-Peng Li; Jun Xu; Meng Zhao
Journal:  Int J Ophthalmol       Date:  2022-08-18       Impact factor: 1.645

4.  Secondary multifocal intraocular lens implantation: A novel management strategy for white cataracts.

Authors:  Jiawen Ling; Yiyong Qian; Peirong Lu
Journal:  SAGE Open Med Case Rep       Date:  2018-01-04

5.  Precision of a new ocular biometer in eyes with cataract using swept source optical coherence tomography combined with Placido-disk corneal topography.

Authors:  Wenwen Wang; Yaxin Miao; Giacomo Savini; Colm McAlinden; Hao Chen; Qingjie Hu; Qinmei Wang; Jinhai Huang
Journal:  Sci Rep       Date:  2017-10-23       Impact factor: 4.379

6.  Refractive outcomes of a single-step and a two-step approach for silicone oil removal and cataract surgery.

Authors:  V G Madanagopalan; Pradeep Susvar; M Arthi
Journal:  Indian J Ophthalmol       Date:  2019-05       Impact factor: 1.848

7.  Commentary: Refractive outcomes of a single step and a two-step approach for silicone oil removal and cataract surgery.

Authors:  Gopal Lingam
Journal:  Indian J Ophthalmol       Date:  2019-07       Impact factor: 1.848

8.  Comparison of two novel swept-source optical coherence tomography devices to a partial coherence interferometry-based biometer.

Authors:  Tommy C Y Chan; Marco C Y Yu; Vivian Chiu; Gilda Lai; Christopher K S Leung; Poemen P M Chan
Journal:  Sci Rep       Date:  2021-07-21       Impact factor: 4.379

9.  Evaluation and comparison of the new swept source OCT-based IOLMaster 700 with the IOLMaster 500.

Authors:  Ahmet Akman; Leyla Asena; Sirel Gür Güngör
Journal:  Br J Ophthalmol       Date:  2015-12-16       Impact factor: 4.638

10.  Accuracy of axial length measurements obtained by optical biometry and acoustic biometry in rhegmatogenous retinal detachment: a prospective study.

Authors:  Pear Pongsachareonnont; Sagol Tangjanyatam
Journal:  Clin Ophthalmol       Date:  2018-05-23
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.