Abstract
BACKGROUND: The aim was to compare the anterior chamber depth (ACD) measurements taken with Orbscan II, ultrasound biomicroscopy (UBM) and the Artemis-2 VHF (very-high-frequency) ultrasound scanner in normal subjects.Entities:
Mesh:
Year: 2014 PMID: 24564379 PMCID: PMC3936803 DOI: 10.1186/1471-2415-14-20
Source DB: PubMed Journal: BMC Ophthalmol ISSN: 1471-2415 Impact factor: 2.209
The intra-observer coefficient of variance (CV) (%) in anterior chamber depth (ACD) measurements and intra-session repeatability for Orbscan II, Artemis-2 and UBM evaluated by intra-class correlation coefficient (ICC)
| CV mean ± SD (%) | 0.015 ± 0.014 | 0.08 ± 0.09 | 0.07 ± 0.06 |
| Significance of differences between CVs (p value) | 0.12 | 0.10 | 0.70 |
| ICC (95% CI) | 0.990 (0.984–0.994) | 0.990 (0.984–0.994) | 0.990 (0.984–0.994) |
CI, confidence interval; SD, standard deviation; UBM, ultrasound biomicroscopy.
The mean differences ± standard deviation (SD), limits of agreement and pair-wise comparisons of p-values for anterior chamber depth (ACD) measurements using the three instruments
| Orbscan II | 0.17 ± 0.31 mm | 0.80 to - 0.43 mm | |
| Orbscan II | 0.27 ± 0.34 mm | 0.94 to - 0.39 mm | |
| UBM | 0.10 ± 0.18 mm | 0.45 to - 0.26 mm |
UBM, ultrasound biomicroscopy; VHFUS, very high-frequency ultrasound scanner. *Pairwise comparison.
Figure 1Bland–Altman plot of agreement showing the mean differences and limits of agreement of anterior chamber depth (ACD) measurements between UBM and Orbscan II.
Figure 2Bland–Altman plot of agreement showing the mean differences and limits of agreement of anterior chamber depth (ACD) measurements between UBM and Artemis-2 VHFUS.
Figure 3Bland–Altman plot of agreement showing the mean differences and limits of agreement of anterior chamber depth (ACD) measurements between Orbscan II and Artemis-2 VHFUS.
Overview of recent studies comparing the repeatability of different instruments for measuring anterior chamber depth (ACD)
| Lee | Orbscan II* | ACD | 44.25 | -5.50 to + 3.00 D | No |
| Hashemi | Orbscan I* | ACD | 30.2 ± 8.5 | -15.00 to - 1.00 D | No |
| Chaidaroon and Jengialern (2005)
[ | Orbscan I* | ACD | 27.90 ± 8.80 | -8.49 ± 4.00 D | No |
SD, standard deviation; UBM, ultrasound biomicroscopy; USP, ultrasound probe.
*Optical non-contact.
**Ultrasound immersion.
†Ultrasound contact.
‡In all studies the difference between instruments was statistically significant.