Literature DB >> 15177602

Comparative estimation of anterior chamber depth by ultrasonography, Orbscan II, and IOLMaster.

Aravind R Reddy1, Milind V Pande, Paul Finn, Hazem El-Gogary.   

Abstract

PURPOSE: To assess the degree of agreement of anterior chamber depth (ACD) measurements by 2 optical devices (Orbscan II and IOLMaster) and compare them with contact ultrasound A-scan.
SETTING: Department of Ophthalmology at Hull Royal Infirmary, England. methods: In this prospective study of 81 eyes of 41 consecutive patients, ACD estimation was done by 3 methods-scanning slit topography (Orbscan II), partial coherence interferometry (IOLMaster), and contact ultrasound A-scan in that order. The same observer performed all measurements. Repeated measures Analysis of Variance (ANOVA) was used to analyze difference between ACD measurements by the 3 methods. Difference in measurements between methods was assessed using the paired t test.
RESULTS: The mean ACD (+/-SD) by the 3 methods was 3.32 mm (0.60), 3.33 mm (0.61), and 2.87 mm (0.55), respectively. There was a statistically significant difference between measurements recorded by the 3 methods (P<.01). Mean contact A-scan measurements were 0.40 mm and 0.43 mm lower than by Orbscan II and IOLMaster respectively (P<.01).
CONCLUSION: Applanation ultrasound gives consistently lower measurements for ACD compared to Orbscan II and IOLMaster. Although a high degree of agreement between Orbscan II and IOLMaster was noted, further studies are needed to assess the interchangeability of measurements in clinical practice.

Entities:  

Mesh:

Year:  2004        PMID: 15177602     DOI: 10.1016/j.jcrs.2003.11.053

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  20 in total

1.  Orbscan II and double-K method for IOL calculation after refractive surgery.

Authors:  Sérgio Kwitko; Diane R Marinho; Samuel Rymer; Norton Severo; Carlos G Arce
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2012-03-29       Impact factor: 3.117

2.  Comparison of optical low-coherence reflectometry and applanation ultrasound biometry on intraocular lens power calculation.

Authors:  Mirjana Bjeloš Rončević; Mladen Bušić; Ivan Cima; Biljana Kuzmanović Elabjer; Damir Bosnar; Daliborka Miletić
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2010-09-18       Impact factor: 3.117

3.  Evaluation of the agreement between Pentacam and ultrasound biomicroscopy measurements of anterior chamber depth in Chinese patients with primary angle-closure glaucoma.

Authors:  Juan Liang; Wei Liu; Xiaoli Xing; Hua Liu; Shaozhen Zhao; Jian Ji
Journal:  Jpn J Ophthalmol       Date:  2010-08-11       Impact factor: 2.447

4.  Anterior chamber width measurement by high-speed optical coherence tomography.

Authors:  Jason A Goldsmith; Yan Li; Maria Regina Chalita; Volker Westphal; Chetan A Patil; Andrew M Rollins; Joseph A Izatt; David Huang
Journal:  Ophthalmology       Date:  2005-02       Impact factor: 12.079

Review 5.  Recent advances in ophthalmic anterior segment imaging: a new era for ophthalmic diagnosis?

Authors:  Aristides Konstantopoulos; Parwez Hossain; David F Anderson
Journal:  Br J Ophthalmol       Date:  2007-04       Impact factor: 4.638

6.  [Scheimpflug and topography systems in ophthalmologic diagnostics].

Authors:  G U Auffarth; A F M Borkenstein; A Ehmer; A Mannsfeld; T M Rabsilber; M P Holzer
Journal:  Ophthalmologe       Date:  2008-09       Impact factor: 1.059

7.  Comparison of techniques for measuring anterior chamber depth: Orbscan imaging, Smith's technique, and van Herick's method.

Authors:  Frank Eperjesi; Claire Holden
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2010-09-10       Impact factor: 3.117

8.  Anterior chamber measurements by pentacam and AS-OCT in eyes with normal open angles.

Authors:  Jeong-Ho Yi; Samin Hong; Gong Je Seong; Sung Yong Kang; Kyoung Tak Ma; Chan Yun Kim
Journal:  Korean J Ophthalmol       Date:  2008-12

9.  Optical biometry intraocular lens power calculation using different formulas in patients with different axial lengths.

Authors:  Jia-Kang Wang; Shu-Wen Chang
Journal:  Int J Ophthalmol       Date:  2013-04-18       Impact factor: 1.779

10.  Anterior segment biometry using ultrasound biomicroscopy and the Artemis-2 very high frequency ultrasound scanner.

Authors:  Haya M Al-Farhan; Reem N Almutairi
Journal:  Clin Ophthalmol       Date:  2013-01-16